Performance measurement of PSF modeling reconstruction (True X) on Siemens Biograph TruePoint TrueV PET/CT

被引:38
作者
Lee, Young Sub [1 ,2 ,3 ]
Kim, Jin Su [1 ]
Kim, Kyeong Min [1 ]
Kang, Joo Hyun [1 ]
Lim, Sang Moo [1 ]
Kim, Hee-Joung [2 ,3 ]
机构
[1] Korea Inst Radiol & Med Sci, Mol Imaging Res Ctr, Seoul 139706, South Korea
[2] Yonsei Univ, Dept Radiol Sci, Wonju, South Korea
[3] Yonsei Univ, Res Inst Hlth Sci, Wonju, South Korea
关键词
Performance measurement; PET; SMALL-ANIMAL PET; IMAGE-RECONSTRUCTION; HIGH-RESOLUTION; SCANNER; NOISE;
D O I
10.1007/s12149-014-0815-z
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The Siemens Biograph TruePoint TrueV (B-TPTV) positron emission tomography (PET) scanner performs 3D PET reconstruction using a system matrix with point spread function (PSF) modeling (called the True X reconstruction). PET resolution was dramatically improved with the True X method. In this study, we assessed the spatial resolution and image quality on a B-TPTV PET scanner. In addition, we assessed the feasibility of animal imaging with a B-TPTV PET and compared it with a microPET R4 scanner. Spatial resolution was measured at center and at 8 cm offset from the center in transverse plane with warm background activity. True X, ordered subset expectation maximization (OSEM) without PSF modeling, and filtered back-projection (FBP) reconstruction methods were used. Percent contrast (% contrast) and percent background variability (% BV) were assessed according to NEMA NU2-2007. The recovery coefficient (RC), non-uniformity, spill-over ratio (SOR), and PET imaging of the Micro Deluxe Phantom were assessed to compare image quality of B-TPTV PET with that of the microPET R4. When True X reconstruction was used, spatial resolution was < 3.65 mm with warm background activity. % contrast and % BV with True X reconstruction were higher than those with the OSEM reconstruction algorithm without PSF modeling. In addition, the RC with True X reconstruction was higher than that with the FBP method and the OSEM without PSF modeling method on the microPET R4. The non-uniformity with True X reconstruction was higher than that with FBP and OSEM without PSF modeling on microPET R4. SOR with True X reconstruction was better than that with FBP or OSEM without PSF modeling on the microPET R4. This study assessed the performance of the True X reconstruction. Spatial resolution with True X reconstruction was improved by 45 % and its % contrast was significantly improved compared to those with the conventional OSEM without PSF modeling reconstruction algorithm. The noise level was higher than that with the other reconstruction algorithm. Therefore, True X reconstruction should be used with caution when quantifying PET data.
引用
收藏
页码:340 / 348
页数:9
相关论文
共 17 条
[11]  
National Electrical Manufacturers Association (NEMA), 2007, NEMA STAND PUBL
[12]   Fully 3-D PET reconstruction with system matrix derived from point source measurements [J].
Panin, Vladimir Y. ;
Kehren, Frank ;
Michel, Christian ;
Casey, Michael .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2006, 25 (07) :907-921
[13]   Fully-3D PET Image Reconstruction Using Scanner-Independent, Adaptive Projection Data and Highly Rotation-Symmetric Voxel Assemblies [J].
Scheins, J. J. ;
Herzog, H. ;
Shah, N. J. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2011, 30 (03) :879-892
[14]   Detector response models for statistical iterative image reconstruction in high resolution PET [J].
Selivanov, VV ;
Picard, Y ;
Cadorette, J ;
Rodrigue, S ;
Lecomte, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (03) :1168-1175
[15]   NOISE AND EDGE ARTIFACTS IN MAXIMUM-LIKELIHOOD RECONSTRUCTIONS FOR EMISSION TOMOGRAPHY [J].
SNYDER, DL ;
MILLER, MI ;
THOMAS, LJ ;
POLITTE, DG .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1987, 6 (03) :228-238
[16]   Noise and signal properties in PSF-based fully 3D PET image reconstruction: an experimental evaluation [J].
Tong, S. ;
Alessio, A. M. ;
Kinahan, P. E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (05) :1453-1473
[17]   Advancement in PET quantification using 3D-OP-OSEM point spread function reconstruction with the HRRT [J].
Varrone, Andrea ;
Sjoholm, Nils ;
Eriksson, Lars ;
Gulyas, Balazs ;
Halldin, Christer ;
Farde, Lars .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 (10) :1639-1650