Performance measurement of PSF modeling reconstruction (True X) on Siemens Biograph TruePoint TrueV PET/CT

被引:38
作者
Lee, Young Sub [1 ,2 ,3 ]
Kim, Jin Su [1 ]
Kim, Kyeong Min [1 ]
Kang, Joo Hyun [1 ]
Lim, Sang Moo [1 ]
Kim, Hee-Joung [2 ,3 ]
机构
[1] Korea Inst Radiol & Med Sci, Mol Imaging Res Ctr, Seoul 139706, South Korea
[2] Yonsei Univ, Dept Radiol Sci, Wonju, South Korea
[3] Yonsei Univ, Res Inst Hlth Sci, Wonju, South Korea
关键词
Performance measurement; PET; SMALL-ANIMAL PET; IMAGE-RECONSTRUCTION; HIGH-RESOLUTION; SCANNER; NOISE;
D O I
10.1007/s12149-014-0815-z
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The Siemens Biograph TruePoint TrueV (B-TPTV) positron emission tomography (PET) scanner performs 3D PET reconstruction using a system matrix with point spread function (PSF) modeling (called the True X reconstruction). PET resolution was dramatically improved with the True X method. In this study, we assessed the spatial resolution and image quality on a B-TPTV PET scanner. In addition, we assessed the feasibility of animal imaging with a B-TPTV PET and compared it with a microPET R4 scanner. Spatial resolution was measured at center and at 8 cm offset from the center in transverse plane with warm background activity. True X, ordered subset expectation maximization (OSEM) without PSF modeling, and filtered back-projection (FBP) reconstruction methods were used. Percent contrast (% contrast) and percent background variability (% BV) were assessed according to NEMA NU2-2007. The recovery coefficient (RC), non-uniformity, spill-over ratio (SOR), and PET imaging of the Micro Deluxe Phantom were assessed to compare image quality of B-TPTV PET with that of the microPET R4. When True X reconstruction was used, spatial resolution was < 3.65 mm with warm background activity. % contrast and % BV with True X reconstruction were higher than those with the OSEM reconstruction algorithm without PSF modeling. In addition, the RC with True X reconstruction was higher than that with the FBP method and the OSEM without PSF modeling method on the microPET R4. The non-uniformity with True X reconstruction was higher than that with FBP and OSEM without PSF modeling on microPET R4. SOR with True X reconstruction was better than that with FBP or OSEM without PSF modeling on the microPET R4. This study assessed the performance of the True X reconstruction. Spatial resolution with True X reconstruction was improved by 45 % and its % contrast was significantly improved compared to those with the conventional OSEM without PSF modeling reconstruction algorithm. The noise level was higher than that with the other reconstruction algorithm. Therefore, True X reconstruction should be used with caution when quantifying PET data.
引用
收藏
页码:340 / 348
页数:9
相关论文
共 17 条
[1]  
[Anonymous], 2008, NEMA STAND PUBL
[2]  
Blinder SAL, 2010, IEEE NUCL SCI CONF R, P3333, DOI 10.1109/NSSMIC.2010.5874422
[3]   Performance evaluation of the ECAT HRRT: an LSO-LYSO double layer high resolution, high sensitivity scanner [J].
de Jong, Hugo W. A. M. ;
van Velden, Floris H. P. ;
Kloet, Reina W. ;
Buijs, Fred L. ;
Boellaard, Ronald ;
Lammertsma, Adriaan A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (05) :1505-1526
[4]   Image-Quality Assessment for Several Positron Emitters Using the NEMA NU 4-2008 Standards in the Siemens Inveon Small-Animal PET Scanner [J].
Disselhorst, Jonathan A. ;
Brom, Maarten ;
Laverman, Peter ;
Slump, Cornelius H. ;
Boerman, Otto C. ;
Oyen, Wim J. G. ;
Gotthardt, Martin ;
Visser, Eric P. .
JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (04) :610-617
[5]  
Ito M, 2011, BIOMED ENG LETT, V1, P70
[6]   Performance Characteristics of a New LSO PET/CT Scanner With Extended Axial Field-of-View and PSF Reconstruction [J].
Jakoby, Bjoern W. ;
Bercier, Yanic ;
Watson, Charles C. ;
Bendriem, Bernard ;
Townsend, David W. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) :633-639
[7]   Performance evaluation of the microPET R4 PET scanner for rodents [J].
Knoess, C ;
Siegel, S ;
Smith, A ;
Newport, D ;
Richerzhagen, R ;
Winkeler, A ;
Jacobs, A ;
Goble, RN ;
Graf, R ;
Wienhard, K ;
Heiss, WD .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (05) :737-747
[8]   An Assessment of the Impact of Incorporating Time-of-Flight Information into Clinical PET/CT Imaging [J].
Lois, Cristina ;
Jakoby, Bjoern W. ;
Long, Misty J. ;
Hubner, Karl F. ;
Barker, David W. ;
Casey, Michael E. ;
Conti, Maurizio ;
Panin, Vladimir Y. ;
Kadrmas, Dan J. ;
Townsend, David W. .
JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (02) :237-245
[9]   Performance characterization of the Inveon preclinical small-animal PET/SPECT/CT system for multimodality imaging [J].
Magota, Keiichi ;
Kubo, Naoki ;
Kuge, Yuji ;
Nishijima, Ken-ichi ;
Zhao, Songji ;
Tamaki, Nagara .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 (04) :742-752
[10]   Role of small animal PET for molecular imaging in pre-clinical studies [J].
Nanni, Cristina ;
Rubello, Domenico ;
Fanti, Stefano .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 (11) :1819-1822