Development of a specific troponin I detection system with enhanced immune sensitivity using a single monoclonal antibody

被引:9
作者
Bozdogan, Anil [1 ,2 ]
El-Kased, Reham F. [3 ]
Jungbluth, Vanessa [2 ]
Knoll, Wolfgang [1 ,2 ]
Dostalek, Jakub [2 ]
Kasry, Amal [1 ,4 ]
机构
[1] CEST Competence Ctr Elect Surface Technol, A-2700 Wiener Neustadt, Austria
[2] AIT Austrian Inst Technol GmbH, Biosensor Technol, Konrad Lorenz Str 24, A-3430 Tulln, Austria
[3] British Univ Egypt BUE, Fac Pharm, Dept Microbiol & Immunol, Suez Desert Rd, Cairo 11837, Egypt
[4] British Univ Egypt BUE, Nanotechnol Res Ctr NTRC, Suez Desert Rd, Cairo 11837, Egypt
来源
ROYAL SOCIETY OPEN SCIENCE | 2020年 / 7卷 / 10期
关键词
troponin I; surface plasmon; single monoclonal antibody; biosensor; SURFACE-PLASMONS; BIOSENSOR; BIOMARKER; SELECTION; FILMS;
D O I
10.1098/rsos.200871
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Using an immunoassay in combination with surface plasmon fluorescence spectroscopy (SPFS), we report the rapid detection of troponin I, a valuable biomarker for diagnosis of myocardial infarction. We discuss the implementation of (i) direct, (ii) sandwich, and (iii) competitive assay formats, based on surface plasmon resonance and SPFS. To elucidate the results, we relate the experiments to orientation-dependent interaction of troponin I epitopes with respective immunoglobulin G antibodies. A limit of detection (LoD) of 19 pM, with 45 min readout time, was achieved using single monoclonal antibody that is specific for one epitope. The borderline between normal people and patients is 20 pM to 83 pM cTnI concentration, and upon the outbreak of acute myocardial infraction it can raise to 2 nM and levels at 20 nM for 6-8 days, therefore the achieved LoD covers most of the clinically relevant range. In addition, this system allows for the detection of troponin I using a single specific monoclonal antibody, which is highly beneficial in case of detection in real samples, where the protein has a complex form leading to hidden epitopes, thus paving the way towards a system that can improve early-stage screening of heart attacks.
引用
收藏
页数:10
相关论文
共 40 条
  • [1] [Anonymous], 2007, J AM COLL CARDIOL, DOI [DOI 10.1016/J.JACC.2007.09.011, DOI 10.1016/j.jacc.2007.09.011]
  • [2] [Anonymous], 2020, About cardiovascular disease
  • [3] [Anonymous], 2018, Cardiovascular disease
  • [4] Assessment of the Multiple-Biomarker Approach for Diagnosis of Myocardial Infarction in Patients Presenting with Symptoms Suggestive of Acute Coronary Syndrome
    Apple, Fred S.
    Smith, Stephen W.
    Pearce, Lesly A.
    Murakami, MaryAnn M.
    [J]. CLINICAL CHEMISTRY, 2009, 55 (01) : 93 - 100
  • [5] Rapid and Sensitive Detection of Troponin I in Human Whole Blood Samples by Using Silver Nanoparticle Films and Microwave Heating
    Aslan, Kadir
    Grell, Tsehai A. J.
    [J]. CLINICAL CHEMISTRY, 2011, 57 (05) : 746 - 752
  • [6] Troponin: the biomarker of choice for the detection of cardiac injury
    Babuin, L
    Jaffe, AS
    [J]. CANADIAN MEDICAL ASSOCIATION JOURNAL, 2005, 173 (10) : 1191 - 1202
  • [7] Plasmon-Enhanced Fluorescence Biosensors: a Review
    Bauch, Martin
    Toma, Koji
    Toma, Mana
    Zhang, Qingwen
    Dostalek, Jakub
    [J]. PLASMONICS, 2014, 9 (04) : 781 - 799
  • [8] BLAST, GLOB AL
  • [9] Rapid and Sensitive Detection of Cardiac Troponin I for Point-of-Care Tests Based on Red Fluorescent Microspheres
    Cai, Yanxue
    Kang, Keren
    Li, Qianru
    Wang, Yu
    He, Xiaowei
    [J]. MOLECULES, 2018, 23 (05):
  • [10] Multichannel SPR biosensor for detection of endocrine-disrupting compounds
    Dostalek, J.
    Pribyl, J.
    Homola, J.
    Skladal, P.
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2007, 389 (06) : 1841 - 1847