Cascaded systems analysis of photon counting detectors

被引:46
作者
Xu, J. [1 ]
Zbijewski, W. [1 ]
Gang, G. [1 ]
Stayman, J. W. [1 ]
Taguchi, K. [2 ]
Lundqvist, M. [3 ]
Fredenberg, E. [3 ]
Carrino, J. A. [2 ]
Siewerdsen, J. H. [1 ,2 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA
[2] Johns Hopkins Univ, Russell H Morgan Dept Radiol, Baltimore, MD 21205 USA
[3] Philips Healthcare, S-17141 Solna, Sweden
关键词
photon counting detector; cascaded systems analysis; modulation transfer function; noise-power spectrum; detective quantum efficiency; image quality; charge sharing model; MODULATION TRANSFER-FUNCTION; SILICON STRIP DETECTOR; FLAT-PANEL IMAGERS; NOISE-POWER SPECTRUM; X-RAY-DETECTORS; DIGITAL MAMMOGRAPHY; QUANTUM EFFICIENCY; ACTIVE-MATRIX; PERFORMANCE EVALUATION; DIAGNOSTIC-RADIOLOGY;
D O I
10.1118/1.4894733
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Photon counting detectors (PCDs) are an emerging technology with applications in spectral and low-dose radiographic and tomographic imaging. This paper develops an analytical model of PCD imaging performance, including the system gain, modulation transfer function (MTF), noise-power spectrum (NPS), and detective quantum efficiency (DQE). Methods: A cascaded systems analysis model describing the propagation of quanta through the imaging chain was developed. The model was validated in comparison to the physical performance of a silicon-strip PCD implemented on an experimental imaging bench. The signal response, MTF, and NPS were measured and compared to theory as a function of exposure conditions (70 kVp, 1-7 mA), detector threshold, and readout mode (i.e., the option for coincidence detection). The model sheds new light on the dependence of spatial resolution, charge sharing, and additive noise effects on threshold selection and was used to investigate the factors governing PCD performance, including the fundamental advantages and limitations of PCDs in comparison to energy-integrating detectors (EIDs) in the linear regime for which pulse pileup can be ignored. Results: The detector exhibited highly linear mean signal response across the system operating range and agreed well with theoretical prediction, as did the system MTF and NPS. The DQE analyzed as a function of kilovolt (peak), exposure, detector threshold, and readout mode revealed important considerations for system optimization. The model also demonstrated the important implications of false counts from both additive electronic noise and charge sharing and highlighted the system design and operational parameters that most affect detector performance in the presence of such factors: for example, increasing the detector threshold from 0 to 100 (arbitrary units of pulse height threshold roughly equivalent to 0.5 and 6 keV energy threshold, respectively), increased the f50 (spatial-frequency at which the MTF falls to a value of 0.50) by similar to 30% with corresponding improvement in DQE. The range in exposure and additive noise for which PCDs yield intrinsically higher DQE was quantified, showing performance advantages under conditions of very low-dose, high additive noise, and high fidelity rejection of coincident photons. Conclusions: The model for PCD signal and noise performance agreed with measurements of detector signal, MTF, and NPS and provided a useful basis for understanding complex dependencies in PCD imaging performance and the potential advantages (and disadvantages) in comparison to EIDs as well as an important guide to task-based optimization in developing new PCD imaging systems. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:15
相关论文
共 56 条
  • [1] Optimization of phosphor-based detector design for oblique x-ray incidence in digital breast tomosynthesis
    Acciavatti, Raymond J.
    Maidment, Andrew D. A.
    [J]. MEDICAL PHYSICS, 2011, 38 (11) : 6188 - 6202
  • [2] An Analytical Model of NPS and DQE Comparing Photon Counting and Energy Integrating Detectors
    Acciavatti, Raymond J.
    Maidment, Andrew D. A.
    [J]. MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [3] Quantitative material characterization from multi-energy photon counting CT
    Alessio, Adam M.
    MacDonald, Lawrence R.
    [J]. MEDICAL PHYSICS, 2013, 40 (03)
  • [4] Optimization of K-edge imaging for vulnerable plaques using gold nanoparticles and energy resolved photon counting detectors: a simulation study
    Alivov, Yahya
    Baturin, Pavlo
    Le, Huy Q.
    Ducote, Justin
    Molloi, Sabee
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (01) : 135 - 152
  • [5] Strategies to improve the signal and noise performance of active matrix, flat-panel imagers for diagnostic x-ray applications
    Antonuk, LE
    Jee, KW
    El-Mohri, Y
    Maolinbay, M
    Nassif, S
    Rong, X
    Zhao, Q
    Siewerdsen, JH
    Street, RA
    Shah, KS
    [J]. MEDICAL PHYSICS, 2000, 27 (02) : 289 - 306
  • [6] Initial performance evaluation of an indirect-detection, active matrix flat-panel imager (AMFPI) prototype for megavoltage imaging
    Antonuk, LE
    El-Mohri, Y
    Huang, WD
    Jee, KW
    Siewerdsen, JH
    Maolinbay, M
    Scarpine, VE
    Sandler, H
    Yorkston, J
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (02): : 437 - 454
  • [7] Oblique incidence effects in direct x-ray detectors: A first-order approximation using a physics-based analytical model
    Badano, Aldo
    Freed, Melanie
    Fang, Yuan
    [J]. MEDICAL PHYSICS, 2011, 38 (04) : 2095 - 2098
  • [8] High resolution X-ray imaging using a silicon strip detector
    Beuville, E
    Cahn, R
    Cederström, B
    Danielsson, M
    Hall, A
    Hasegawa, B
    Luo, L
    Lundqvist, M
    Nygren, D
    Oltman, E
    Walton, J
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (06) : 3059 - 3063
  • [9] Accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV
    Boone, JM
    Seibert, JA
    [J]. MEDICAL PHYSICS, 1997, 24 (11) : 1661 - 1670
  • [10] Detective quantum efficiency dependence on x-ray energy weighting in mammography
    Cahn, RN
    Cederström, B
    Danielsson, M
    Hall, A
    Lundqvist, M
    Nygren, D
    [J]. MEDICAL PHYSICS, 1999, 26 (12) : 2680 - 2683