Ray tracing flux calculation for the small and wide angle x-ray scattering diffraction station at the SESAME synchrotron radiation facility

被引:3
作者
Salah, Wa'el [1 ,2 ]
del Rio, M. Sanchez [3 ]
Hoorani, H. [1 ]
机构
[1] SESAME, Allan 19252, Jordan
[2] Hashemite Univ, Dept Phys, Zarqa 13115, Jordan
[3] European Synchrotron Radiat Facil, F-38043 Grenoble, France
关键词
SOFTWARE; CRYSTALS; SHADOW; DESIGN; OPTICS; BENT;
D O I
10.1063/1.3213620
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The calculation for the optics of the synchrotron radiation small and wide angle x-ray scattering beamline, currently under construction at SESAME is described. This beamline is based on a cylindrically bent germanium (111) single crystal with an asymmetric cut of 10.5, followed by a 1.2 m long rhodium coated plane mirror bent into a cylindrical form. The focusing properties of bent asymmetrically cut crystals have not yet been studied in depth. The present paper is devoted to study of a particular application of a bent asymmetrically cut crystal using ray tracing simulations with the SHADOW code. These simulations show that photon fluxes of order of 1.09 x 10(11) photons/s will be available at the experimental focus at 8.79 keV. The focused beam dimensions will be 2.2 mm horizontal full width at half maximum (FWHM) by 0.12 mm vertical (FWHM). (C) 2009 American Institute of Physics. [doi:10.1063/1.3213620]