A 1.5 T transverse magnetic field in radiotherapy of rectal cancer: Impact on the dose distribution

被引:22
|
作者
Uilkema, Sander [1 ]
van der Heide, Uulke [1 ]
Sonke, Jan-Jakob [1 ]
Moreau, Michel [2 ]
van Triest, Baukelien [1 ]
Nijkamp, Jasper [1 ]
机构
[1] NKI AVL, Dept Radiotherapy, NL-1066 CX Amsterdam, Netherlands
[2] Elekta, RTP Res Grp, Maryland Hts, MO 63043 USA
关键词
electron return effect (ERE); 1.5 T transverse magnetic field; rectal cancer; MR-guided; radiotherapy; IMRT; MRI SCANNER; MONTE-CARLO; ACCELERATOR; SYSTEMS; OPTIMIZATION; IRRADIATION; GPUMCD; VOLUME; IMRT;
D O I
10.1118/1.4936097
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: MRI guidance during radiotherapy has the potential to enable more accurate dose delivery, optimizing the balance between local control and treatment related toxicity. However, the presence of a permanent magnetic field influences the dose delivery, especially around air cavities. Here, electrons are able to return to the surface through which they entered the air cavity (electron return effect, ERE) locally resulting in dose hot-and cold-spots. Where RT of rectal cancer patients might benefit from MRI guidance for margin reduction, air cavities in and around the target volume are frequently present. The purpose of this research is to evaluate the impact of the presence of a 1.5 T transverse magnetic field on dose delivery in patients with rectal cancer. Methods: Ten patients treated with 5x5 Gy RT having large changes in pelvic air content were selected out of a cohort of 33 patients. On the planning CT, a 1.5 T, 6 MV, 7-field intensity modulated radiotherapy (IMRT) plan was created. This plan was subsequently recalculated on daily CT scans. For each daily CT, the CTV V-95% and V-107% and bowel area V-5Gy, V-10Gy, V-15Gy, V-20Gy, and V-25Gy were calculated to evaluate the changes in dose distribution from fraction to fraction. For comparison, the authors repeated this procedure for the 0 T situation. To study the effect of changing air cavities separate from other anatomical changes, the authors also generated artificial air cavities in the CTV of one patient (2 and 5 cm diameter), in the high dose gradient region (2 cm), and in the low dose area (2 cm). Treatment plans were optimized without and with each simulated air cavity. For appearing and disappearing air cavities, the CTV V-95% and V-107% were evaluated. The authors also evaluated the ERE separate from attenuation changes locally around appearing gas pockets. Results: For the ten patients, at 1.5 T, the V-95% was influenced by both appearing and disappearing air, and dropped to < 98% in 2 out of 50 fractions due a disappearing air cavity of 150 cm(3). V-95% differences between 0 and 1.5 T were all within 2%. The V-107% was below 1% in 46 out of 50 fractions, and increased to 3% in the remaining fractions due to appearing air of around 120 cm(3). For comparison, V-107% was < 1% at 0 T for all fractions. In the bowel area, the V-15Gy varied strongest from fraction to fraction, but differences between 1.5 and 0 T were minimal with an average difference of 2.3 cm(3) (SD = 18.7 cm(3), p = 0.38). For the simulated air cavities, the ERE resulted in cold-spots maximally 5% lower than prescribed and hot-spots maximally 6% higher than prescribed. Conclusions: The presence of a 1.5 T magnetic field has an impact on the dose distribution when the air content changes of within a few percent in these selected rectal cancer patients. The authors consider this influence of the transverse magnetic field on the dose distribution in IMRT for rectal cancer patients clinically acceptable. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:7182 / 7189
页数:8
相关论文
共 50 条
  • [1] Assessing the Impact of a 1.5 T Transverse Magnetic Field in Radiotherapy for Esophageal Cancer Patients
    Li, Yukun
    Li, Baosheng
    Zhu, Jian
    Yin, Yong
    Li, Zhenjiang
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2024, 23
  • [2] Technical Note: Dose effects of 1.5 T transverse magnetic field on tissue interfaces in MRI-guided radiotherapy
    Chen, Xinfeng
    Prior, Phil
    Chen, Guang-Pei
    Schultz, Christopher J.
    Li, X. Allen
    MEDICAL PHYSICS, 2016, 43 (08) : 4797 - 4802
  • [3] Investigation of the dose perturbation effect for therapeutic beams with the presence of a 1.5 T transverse magnetic field in magnetic resonance imaging-guided radiotherapy
    Shao, Wencheng
    Tang, Xiaobin
    Bai, Yanling
    Shu, Diyun
    Geng, Changran
    Gong, Chunhui
    Guan, Fada
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2018, 14 (01) : 184 - 195
  • [4] Dosimetric feasibility of intensity modulated proton therapy in a transverse magnetic field of 1.5 T
    Hartman, J.
    Kontaxis, C.
    Bol, G. H.
    Frank, S. J.
    Lagendijk, J. J. W.
    van Vulpen, M.
    Raaymakers, B. W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (15): : 5955 - 5969
  • [5] Impact of Magnetic Field on Dose Distribution in MR-Guided Radiotherapy of Head and Neck Cancer
    Xia, Wenlong
    Zhang, Ke
    Li, Minghui
    Tian, Yuan
    Men, Kuo
    Wang, Jingbo
    Yi, Junlin
    Li, Yexiong
    Dai, Jianrong
    FRONTIERS IN ONCOLOGY, 2020, 10
  • [6] Effect of uniform magnetic field on dose distribution in the breast radiotherapy
    Esmaeeli, A. D.
    Pouladian, M.
    Monfared, A. S.
    Mahdavi, S. R.
    Moslemi, D.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2014, 12 (02): : 151 - 160
  • [7] Monte Carlo simulations of patient dose perturbations in rotational-type radiotherapy due to a transverse magnetic field: A tomotherapy investigation
    Yang, Y. M.
    Geurts, M.
    Smilowitz, J. B.
    Sterpin, E.
    Bednarz, B. P.
    MEDICAL PHYSICS, 2015, 42 (02) : 715 - 725
  • [8] Effect of 0.35 T and 1.5 T magnetic fields on superficial dose in MR-guided radiotherapy of laryngeal cancer
    Conrad, Mireille
    Dal Bello, Riccardo
    van Timmeren, Janita E.
    Andratschke, Nicolaus
    Wilke, Lotte
    Guckenberger, Matthias
    Tanadini-Lang, Stephanie
    Balermpas, Panagiotis
    CLINICAL AND TRANSLATIONAL RADIATION ONCOLOGY, 2023, 40
  • [9] Comparison of dose distributions between transverse magnetic fields of 0.35 T and 1.5 T for radiotherapy in lung tumor using Monte Carlo calculation
    Kubota, Takahiro
    Araki, Fujio
    Ohno, Takeshi
    MEDICAL DOSIMETRY, 2020, 45 (02) : 179 - 185
  • [10] Improvement of dose distribution in breast radiotherapy using a reversible transverse magnetic field Linac-MR unit
    Esmaeeli, A. D.
    Mahdavi, S. R.
    Pouladian, M.
    Monfared, A. S.
    Bagheri, S.
    MEDICAL PHYSICS, 2014, 41 (01)