Rapid assessment of in vivo cholinergic transmission by amperometric detection of changes in extracellular choline levels

被引:87
作者
Parikh, V
Pomerleau, F
Huettl, P
Gerhardt, GA
Sarter, M
Bruno, JP [1 ]
机构
[1] Ohio State Univ, Dept Psychol, Columbus, OH 43210 USA
[2] Univ Kentucky, Med Ctr, Dept Anat & Neurobiol, Ctr Sensor Technol, Lexington, KY 40536 USA
关键词
acetylcholine; frontoparietal cortex; microelectrode arrays; rat;
D O I
10.1111/j.1460-9568.2004.03614.x
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Conventional microdialysis methods for measuring acetylcholine (ACh) efflux do not provide sufficient temporal resolution to relate cholinergic transmission to individual stimuli or behavioral responses, or sufficient spatial resolution to investigate heterogeneities in such regulation within a brain region. In an effort to overcome these constraints, we investigated a ceramic-based microelectrode array designed to measure amperometrically rapid changes in extracellular choline as a marker for cholinergic transmission in the frontoparietal cortex of anesthetized rats. These microelectrodes exhibited detection limits of 300 nm for choline and selectivity (> 100 : 1) of choline over interferents such as ascorbic acid. Intracortical pressure ejections of choline (20 mm, 66-400 nL) and ACh (10 and 100 mm, 200 nL) dose-dependently increased choline-related signals that were cleared to background levels within 10 s. ACh, but not choline-induced signals, were significantly attenuated by co-ejection of the acetylcholinesterase inhibitor neostigmine (Neo; 100 mm). Pressure ejections of drugs known to increase cortical ACh efflux, potassium (KCl; 70 mm, 66, 200 nL) and scopolamine (Scop; 10 mm, 200 nL), also markedly increased extracellular choline signals, which again were inhibited by Neo. Scop-induced choline signals were also found to be tetrodotoxin-sensitive. Collectively, these findings suggest that drug-induced increases in current measured with these microelectrode arrays reflect the oxidation of choline that is neuronally derived from the release and subsequent hydrolysis of ACh. Choline signals assessed using enzyme-selective microelectrode arrays may represent a rapid, sensitive and spatially discrete measure of cholinergic transmission.
引用
收藏
页码:1545 / 1554
页数:10
相关论文
共 52 条
[1]   Maternal dietary choline availability alters mitosis, apoptosis and the localization of TOAD-64 protein in the developing fetal rat septum [J].
Albright, CD ;
Friedrich, CB ;
Brown, EC ;
Mar, MH ;
Zeisel, SH .
DEVELOPMENTAL BRAIN RESEARCH, 1999, 115 (02) :123-129
[2]   Choline is a selective agonist of α7 nicotnic acetylcholine receptors in the rat brain neurons [J].
Alkondon, M ;
Pereira, EFR ;
Cortes, WS ;
Maelicke, A ;
Albuquerque, EX .
EUROPEAN JOURNAL OF NEUROSCIENCE, 1997, 9 (12) :2734-2742
[3]   Differential cortical acetylcholine release in rats performing a sustained attention task versus behavioral control tasks that do not explicitly tax attention [J].
Arnold, HM ;
Burk, JA ;
Hodgson, EM ;
Sarter, M ;
Bruno, JP .
NEUROSCIENCE, 2002, 114 (02) :451-460
[4]   DOPAMINERGIC INNERVATION OF THE CEREBRAL-CORTEX - UNEXPECTED DIFFERENCES BETWEEN RODENTS AND PRIMATES [J].
BERGER, B ;
GASPAR, P ;
VERNEY, C .
TRENDS IN NEUROSCIENCES, 1991, 14 (01) :21-27
[5]  
Bruno JP, 1999, REV NEUROSCIENCE, V10, P25
[6]  
Bunin MA, 1998, J NEUROSCI, V18, P4854
[7]   Improved ceramic-based multisite microelectrode for rapid measurements of L-glutamate in the CNS [J].
Burmeister, JJ ;
Pomerleau, F ;
Palmer, M ;
Day, BK ;
Huettl, P ;
Gerhardt, GA .
JOURNAL OF NEUROSCIENCE METHODS, 2002, 119 (02) :163-171
[8]   Self referencing ceramic based multisite microelectrodes for the detection and elimination of interferences from the measurement of L-glutamate and other analytes [J].
Burmeister, JJ ;
Gerhardt, GA .
ANALYTICAL CHEMISTRY, 2001, 73 (05) :1037-1042
[9]   Ceramic-based multisite microelectrode array for rapid choline measures in brain tissue [J].
Burmeister, JJ ;
Palmer, M ;
Gerhardt, GA .
ANALYTICA CHIMICA ACTA, 2003, 481 (01) :65-74
[10]   Ceramic-based multisite microelectrodes for electrochemical recordings [J].
Burmeister, JJ ;
Moxon, K ;
Gerhardt, GA .
ANALYTICAL CHEMISTRY, 2000, 72 (01) :187-192