The accuracy of transcutaneous bilirubinometer measurements to identify the hyperbilirubinemia in outpatient newborn population

被引:18
作者
Ercan, Serif [1 ]
Ozgun, Gunay [2 ]
机构
[1] Luleburgaz State Hosp, Dept Clin Biochem, Kirklareli, Turkey
[2] Luleburgaz State Hosp, Dept Pediat, Kirklareli, Turkey
关键词
Hyperbilirubinemia; Neonatal jaundice; Total serum bilirubin; Transcutaneous bilirubin; DIRECT SPECTROPHOTOMETRIC METHOD; SERUM BILIRUBIN; TERM;
D O I
10.1016/j.clinbiochem.2018.03.018
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: To assess the ability of transcutaneous bilirubinometer for prediction total serum bilirubin levels in newborn infants after hospital discharge. Methods: Newborn infants requiring total serum bilirubin (TSB) level measurement during an outpatient follow-up visit were included into the study. Transcutaneous bilirubin (TcB) measurement was carried out using JH20-1C (Ningbo David, China) transcutaneous jaundice detector and total serum bilirubin was simultaneously determined by direct spectrophotometry. The agreement between paired TSB and TcB measurements were assessed by Bland-Altman plot and Deming regression analysis. Predictive indices were also identified in different TcB cut-off values for TSB levels of 222, 256 and 291 mu mol/L. Results: A total of 271 paired TcB and TSB measurements were obtained from 218 newborn infants. 40.2% had an age of above 7 days at measurement time. The mean difference (95% CI) between TcB and TSB values was -1.7 (- 5.4 to 2.1) mu mol/L. For TSB levels of at least 256 and 291 mu mol/L, a TcB cut-off of 222 mu mol/L shows sensitivity of 90.6% and 100%, respectively. It was also determined that 39.4% of TSB measurements could be avoided when using TcB cut-off value of 222 mu mol/L. Conclusion: The measurement of TcB with JH20-1C seems to be a reliable screening method for hyperbilirubinemia in the outpatient population when used a TcB cut-off of 222 mu mol/L. The use of transcutaneous bilirubinometer could reduce the number of invasive blood sampling for the determination serum bilirubin.
引用
收藏
页码:69 / 74
页数:6
相关论文
共 19 条
[1]   Transcutaneous bilirubinometry is not influenced by term or skin color in neonates [J].
Afanetti, Mickael ;
Trolli, Sergio Eleni Dit ;
Yousef, Nadya ;
Jrad, Ikram ;
Mokhtari, Mostafa .
EARLY HUMAN DEVELOPMENT, 2014, 90 (08) :417-420
[2]   Evaluation of a point-of-care direct spectrophotometric method for measurement of total serum bilirubin in term and near-term neonates [J].
Barko, H. A. ;
Jackson, G. L. ;
D Engle, W. .
JOURNAL OF PERINATOLOGY, 2006, 26 (02) :100-105
[3]   Limitations and Opportunities of Transcutaneous Bilirubin Measurements [J].
Bosschaart, Nienke ;
Kok, Joke H. ;
Newsum, Astrid M. ;
Ouweneel, Dagmar M. ;
Mentink, Rosaline ;
van Leeuwen, Ton G. ;
Aalders, Maurice C. G. .
PEDIATRICS, 2012, 129 (04) :689-694
[4]  
Canadian Paediatric Society, 2016, GUID DET MAN PREV HY
[5]  
DOUMAS BT, 1985, CLIN CHEM, V31, P1779
[6]   Hyperbilirubinemia and Transcutaneous Bilirubinometry [J].
El-Beshbishi, Samar N. ;
Shattuck, Karen E. ;
Mohammad, Amin A. ;
Petersen, John R. .
CLINICAL CHEMISTRY, 2009, 55 (07) :1280-1287
[7]   Neonatal total bilirubin measurements: Still room for harmonization [J].
Greene, Dina N. ;
Liang, Joy ;
Holmes, Daniel T. ;
Resch, Ann ;
Lorey, Thomas S. .
CLINICAL BIOCHEMISTRY, 2014, 47 (12) :1112-1115
[8]   Bilirubin measurement for neonates:: Comparison of 9 frequently used methods [J].
Grohmann, K ;
Roser, M ;
Rolinski, B ;
Kadow, I ;
Müller, C ;
Goerlach-Graw, A ;
Nauck, M ;
Küster, H .
PEDIATRICS, 2006, 117 (04) :1174-1183
[9]  
Kazmierczak SC, 2002, CLIN CHEM, V48, P1096
[10]   Correlation of Transcutaneous and Serum Bilirubin Measurements in the Outpatient Setting [J].
Kumra, Tina ;
Weaver, Sallie J. ;
Prather, Kristal ;
Garnepudi, Lavanya ;
Bartlett, Edward L., Jr. ;
Crocetti, Michael .
CLINICAL PEDIATRICS, 2018, 57 (02) :231-234