Low cost capillary flow meter

被引:17
作者
Topolnicki, Juliusz [1 ]
Kudasik, Mateusz [1 ]
Skoczylas, Norbert [1 ]
Sobczyk, Jacek [1 ]
机构
[1] Polish Acad Sci, Strata Mech Res Inst, PL-30059 Krakow, Poland
关键词
Flow meter; Capillary; Differential pressure; Flow meter calibration; Capillary calibration;
D O I
10.1016/j.sna.2009.03.023
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This article introduces a project and some specifications of the capillary flow meter. The functioning of the presented flow meter is based on the measurement of pressure drop caused by the gas flowing through the capillary. The article includes in detail a description of the calibration process of the above-mentioned measuring device and presents its basic specifications and capabilities. The measurement results are barometric and temperature compensated. The upper range boundary of the presented device amounts to 45 mu mol/s and the measuring error accounts for 0.3% of the full measuring range of the meter. The costs of material used for constructing the prototype of the capillary flow meter did not exceed US$ 30. (C) 2009 Elsevier BY. All rights reserved.
引用
收藏
页码:146 / 150
页数:5
相关论文
共 6 条
[1]  
ANKLIN M, 2006, FLOW MEASUREMENT INS, V17
[2]  
BELFORTE G, 1997, MEASUREMENT, V20
[3]  
BERG RF, 2004, REV SCI INSTRUMENTS, V75
[4]  
DAVIS L, 2002, PRINCIPLES ENV ENG S
[5]  
ROMER E, 1970, MIERNICTWO PRZEMYSLO
[6]   THE HISTORY OF POISEUILLE LAW [J].
SUTERA, SP ;
SKALAK, R .
ANNUAL REVIEW OF FLUID MECHANICS, 1993, 25 :1-19