Biomechanical analysis of intact versus ruptured Poly Implant Prothese breast implants

被引:2
|
作者
Ramiao, N. G. [1 ]
Martins, P. S. [1 ]
Barroso, M. L. [2 ]
Santos, D. C. [2 ]
Fernandes, A. A. [1 ]
机构
[1] Univ Porto, Fac Engn, LAETA, INEGI, Campus FEUP,Rua Dr Roberto Frias,400, P-4200465 Porto, Portugal
[2] Gaia Hosp Ctr, Dept Plast Surg, Vila Nova De Gaia, Portugal
关键词
breast implants; mechanical properties; Poly Implant Prothese; implant failure; SILICONE GEL; PIP BREAST; MECHANICAL ANALYSIS; FAILURE; DEGRADATION; PROSTHESES; STRENGTH;
D O I
10.1098/rsfs.2018.0086
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Despite there being many studies that have evaluated breast implant rupture, there is no consensus about causes and incidence. Most studies lack a multifactorial analysis of what causes breast implants to rupture. To fill this gap, an experimental protocol was developed to compare ruptured and intact Poly Implant Prothese (PIP) breast implants from the same woman. These conditions guarantee that the physical/biological variables are the same for each pair of ruptured and intact implants. A total of 1008 samples from 22 PIP explants (11 intact and 11 ruptured) and three control PIP implants were analysed. The mechanical properties (tensile strength) of the ruptured and intact implants were compared according to brand, lot, implantation time and demographic conditions. In general, statistically significant differences were found between the intact and ruptured PIP implants. Ruptured implants were thinner (0.73+0.10 mm versus 0.91+0.11 mm) and weaker (7.42+2.65 MPa versus 9.59+2.37 MPa) than intact implants. Intact and ruptured implants have shown distinct mechanical behaviours and variations in thickness. Our understanding is that these differences may be associated with the typical manufacturing process of breast implant shells. These results stress the importance of thorough control of the shell thickness. Given its relevance, shell thickness should be used as a quality control measure for homologation purposes. Thus, the homogeneity of the shell should be considered as a relevant parameter during the manufacturing process. This will translate into an improved quality of life for patients and will potentiate safer and longer lasting products.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] An unusual case of a ruptured Poly Implant Prothese breast implant associated with xanthoma
    Dinets, Andrii
    Unukovych, Dmytro
    Khrapach, Vasyl
    Zakhartseva, Olga
    Sulik, Volodymyr
    Kaminskyi, Edgar
    Khrapach, Olexandr
    Meloyan, Ani
    Zakhartseva, Lubov
    Mishalov, Volodymyr
    CASE REPORTS IN PLASTIC SURGERY AND HAND SURGERY, 2016, 3 (01): : 11 - 15
  • [2] Complications of Poly Implant Prothese breast implants: the current discussion
    Kolios, Leila
    Hirche, Christoph
    Spiethoff, Alexander
    Daigeler, Adrien
    Lehnhardt, Marcus
    EXPERT REVIEW OF MEDICAL DEVICES, 2013, 10 (02) : 167 - 170
  • [3] The Clinical Implications of Poly Implant Prothese Breast Implants: An Overview
    Wazir, Umar
    Kasem, Abdul
    Mokbel, Kefah
    ARCHIVES OF PLASTIC SURGERY-APS, 2015, 42 (01): : 4 - 10
  • [4] Poly Implant Prothese (PIP) breast implants: Our experience
    Chummun, Shaheel
    McLean, Neil R.
    SURGEON-JOURNAL OF THE ROYAL COLLEGES OF SURGEONS OF EDINBURGH AND IRELAND, 2013, 11 (05): : 241 - 245
  • [5] Cutaneous pseudolymphoma caused by Poly Implant Prothese breast implants
    Olisova, O. Y.
    Grabovskaya, O. V.
    Teplyuk, N. P.
    Tertychnyy, A. S.
    Varshavsky, V. A.
    Grekova, E. V.
    JOURNAL OF THE EUROPEAN ACADEMY OF DERMATOLOGY AND VENEREOLOGY, 2018, 32 (09) : E358 - E360
  • [6] Rupture of Poly Implant Prothese Silicone Breast Implants: An Implant Retrieval Study
    Swarts, Eric
    Kop, Alan M.
    Nilasaroya, Anastasia
    Keogh, Catherine V.
    Cooper, Timothy
    PLASTIC AND RECONSTRUCTIVE SURGERY, 2013, 131 (04) : 480E - 489E
  • [7] Complication of Ruptured Poly Implant Prothese® Breast Implants Combined with AQUAfilling® Gel Injection: A Case Report and Literature Review
    Kim, Jaewoo
    Chang, Hak
    Park, Ji-Ung
    AESTHETIC PLASTIC SURGERY, 2019, 43 (01) : 46 - 52
  • [8] Poly Implant Prothese Asymmetrical Anatomical Breast Implants: A Product Recall Study
    De Lorenzi, Francesca
    Gazzola, Riccardo
    Sangalli, Claudia
    Villa, Oriana
    Marchetti, Miriam
    Monturano, Massimo
    Petit, Jean Yves
    Garusi, Cristina
    Martella, Stefano
    Manconi, Andrea
    Barbieri, Benedetta
    Gottardi, Alessandra
    Iera, Marco
    Silva, Marco Aurelio da Costa
    Veronesi, Paolo
    Rietjens, Mario
    PLASTIC AND RECONSTRUCTIVE SURGERY, 2015, 135 (01) : 25 - 33
  • [9] Biodurability of Poly Implant Prothese (PIP) breast implants: A prospective analysis of 1028 prostheses in 514 patients
    Thomson, Richard M.
    Healy, Samuel E.
    Basheer, Mohammed Haj
    Morris, Keith
    Whitaker, Iain S.
    JOURNAL OF PLASTIC RECONSTRUCTIVE AND AESTHETIC SURGERY, 2018, 71 (06): : 953 - 955
  • [10] Sentinel lymph node biopsy in a patient with ruptured poly implant prothese (PIP) implants: A case report
    Tafazal, H.
    Basu, N. N.
    Jewkes, A.
    INTERNATIONAL JOURNAL OF SURGERY CASE REPORTS, 2014, 5 (12): : 992 - 994