摘要
We have developed, fabricated and tested a lidless polymeric microchip which integrates a microreactor with an electrospray ionization (ESI) tip for mass spectrometric (MS) analysis. The microchip contains a microreactor spot, a micropillar-filled channel and an electrospray tip, all made of an SU-8 epoxy polymer. The fabrication process of the microchips required just two lithographic steps, making it simple, cost-effective, rapid, and well-suited for mass production. The lidless structure of the chip provides easy sampling of reagents and samples. On-chip trypsin protein digestion of bovine heart cytochrome c (CytC) and bovine serum albumin (BSA) was performed on the microreactor in a solvent drop without the need for immobilization of the enzyme to the microreactor. The digestion was accomplished in the microreactor within 8 min, after which the digestion products were measured directly by an integrated on-chip ESI combined to MS. The whole procedure was rapid as the total time for on-chip digestion and measurement was less than 10 min. An average sequence coverage obtained with CytC was 90% and with BSA it was 82%. The microchip can also be used as a stand-alone electrospray ionization tip for mass spectrometric analysis. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:414 / 420
页数:7
相关论文
共 39 条
  • [1] A silicon beam-based microcantilever nanoelectrosprayer
    Arscott, Steve
    Legrand, Bernard
    Buchaillot, Lionel
    Ashcroft, Alison E.
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2007, 125 (01): : 72 - 78
  • [2] Integrated microfabricated systems including a purification module and an on-chip nano electrospray ionization interface for biological analysis
    Carlier, J
    Arscott, S
    Thomy, V
    Camart, JC
    Cren-Olivé, C
    Le Gac, S
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2005, 1071 (1-2) : 213 - 222
  • [3] Evaluation of various immobilized enzymatic microreactors coupled on-line with liquid chromatography and mass spectrometry detection for quantitative analysis of cytochrome c
    Cingoz, A.
    Hugon-Chapuis, F.
    Pichon, V.
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2008, 1209 (1-2) : 95 - 103
  • [4] Microfabricated porous glass channels for electrokinetic separation devices
    Costa, RCDA
    Mogensen, KB
    Kutter, JP
    [J]. LAB ON A CHIP, 2005, 5 (11) : 1310 - 1314
  • [5] On the 3-dimensional effects in etched chips for high performance liquid chromatography-separations
    De Smet, J.
    Gzil, P.
    Baron, G. V.
    Desmet, G.
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2007, 1154 (1-2) : 189 - 197
  • [6] On-microchip multiphase chemistry - a review of microreactor design principles and reagent contacting modes
    Doku, GN
    Verboom, W
    Reinhoudt, DN
    van den Berg, A
    [J]. TETRAHEDRON, 2005, 61 (11) : 2733 - 2742
  • [7] Porous silicon as the carrier matrix in microstructured enzyme reactors yielding high enzyme activities
    Drott, J
    Lindstrom, K
    Rosengren, L
    Laurell, T
    [J]. JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 1997, 7 (01) : 14 - 23
  • [8] High-speed biomarker identification utilizing porous silicon nanovial arrays and MALDI-TOF mass spectrometry
    Finnskog, D
    Jaras, K
    Ressine, A
    Malm, J
    Marko-Varga, G
    Lilja, H
    Laurell, T
    [J]. ELECTROPHORESIS, 2006, 27 (5-6) : 1093 - 1103
  • [9] Microfluidic glass chips with an integrated nanospray emitter for coupling to a mass spectrometer
    Hoffmann, Peter
    Haeusig, Ulrich
    Schulze, Philipp
    Belder, Detlev
    [J]. ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 2007, 46 (26) : 4913 - 4916
  • [10] Enhanced protein digestion through the confinement of nanozeolite-assembled microchip reactors
    Ji, Ji
    Zhang, Yahong
    Zhou, Xiaoqin
    Kong, Jilie
    Tang, Yi
    Liu, Baohong
    [J]. ANALYTICAL CHEMISTRY, 2008, 80 (07) : 2457 - 2463