Comparative diagnostic test characteristics of oscillometric and Doppler ultrasonographic methods in the detection of systolic hypertension in dogs

被引:0
作者
Stepien, RL
Rapoport, GS
Henik, RA
Wenholz, L
Thomas, CB
机构
[1] Univ Wisconsin, Sch Vet Med, Dept Med Sci, Madison, WI 53706 USA
[2] Univ Wisconsin, Sch Vet Med, Dept Pathobiol Biosci, Madison, WI 53706 USA
关键词
blood pressure; diagnosis; likelihood ratio; receiver operator characteristic curve; sensitivity; specificity;
D O I
10.1892/0891-6640(2003)017<0065:CDTCOO>2.3.CO;2
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Comparison of test characteristics allows a clinician to choose the optimal diagnostic test method for an individual patient. This study assessed the comparative test characteristics of noninvasive (NI) blood pressure measurement methods (oscillometric and Doppler) and used this information to develop optimal cutoff values for diagnosis of systolic hypertension in dogs by these NI methods. Simultaneous NI (oscillometric or Doppler methods) and invasive (arterial puncture [AP]) systolic blood pressure (SBP) measurements were obtained prospectively from normal dogs and dogs suspected of having systemic hypertension based on clinical signs. Oscillometric SBP readings were obtained from the distal hind limb (Osc-L, n = 54) or the proximal tail (T, n = 27). Doppler BP measurements were obtained using a forelimb cuff (n = 57). AP-SBP was categorized as hypertensive if greater than or equal to160 mmHg, and sensitivity (Se), specificity (Sp), and likelihood ratios (LR) were calculated for diagnostic cutoff values ranging from 130 to 220 mmHg. Receiver operator characteristic (ROC) curves were analyzed to determine optimal cutoff values for diagnosis of AP-SBP greater than or equal to160 mmHg. Optimal NI SBP cutoff values considered to reflect AP values greater than or equal to160 mmHg were: Osc-L = 160 mmHg (Se: 65%, Sp: 85%, LR = 4.33: 1), Osc-T = 150 mmHg (Se: 84%, Sp: 75%, LR = 3.36: 1), and Doppler = 160 mmHg (Se: 71%, Sp: 86%, LR: 5: 1).
引用
收藏
页码:65 / 72
页数:8
相关论文
共 31 条
[1]  
ANDERSON LJ, 1968, RES VET SCI, V9, P304
[2]   Comparison of direct and indirect (oscillometric) measurements of arterial blood pressure in conscious dogs [J].
Bodey, AR ;
Michell, AR ;
Bovee, KC ;
Buranakurl, C ;
Garg, T .
RESEARCH IN VETERINARY SCIENCE, 1996, 61 (01) :17-21
[3]   Longitudinal studies of reproducibility and variability of indirect (oscillometric) blood pressure measurements in dogs: evidence for tracking [J].
Bodey, AR ;
Michell, AR .
RESEARCH IN VETERINARY SCIENCE, 1997, 63 (01) :15-21
[4]   Epidemiological study of blood pressure in domestic dogs [J].
Bodey, AR ;
Michell, AR .
JOURNAL OF SMALL ANIMAL PRACTICE, 1996, 37 (03) :116-125
[5]  
BOVEE KC, 1996, J VET INTERN MED, V10, P161
[6]  
BROWN S, 2001, 19 ANN VET MED FOR, P845
[7]  
BROWN SA, 2000, KIRKS CURRENT VET TH, V13, P835
[8]  
CHALIFOUX A, 1985, CAN J COMP MED, V49, P419
[9]  
Cowgill L. G., 1983, CURRENT VET THERAPY, P1025
[10]  
CROWE DT, 1995, CURRENT VET THERAPY, V12, P113