Integrated microfluidic immunoassay for the rapid determination of clenbuterol

被引:72
作者
Kong, Jing [1 ,2 ]
Jiang, Lei [1 ]
Su, Xiaoou [3 ]
Qin, Jianhua [1 ]
Du, Yuguang [1 ]
Lin, Bingcheng [1 ]
机构
[1] Chinese Acad Sci, Dalian Inst Chem Phys, Dalian 116023, Peoples R China
[2] Chinese Acad Sci, Grad Sch, Beijing 100049, Peoples R China
[3] Chinese Acad Agr Sci, Inst Qual Stand & Testing Technol Agriprod, Beijing 100081, Peoples R China
关键词
PERFORMANCE LIQUID-CHROMATOGRAPHY; TANDEM MASS-SPECTROMETRY; CAPILLARY-ELECTROPHORESIS; PHARMACEUTICAL FORMULATIONS; CHEMILUMINESCENCE DETECTION; HUMAN URINE; HUMAN SERUM; GAS; RESIDUES; TISSUES;
D O I
10.1039/b818430e
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An integrated microfluidic immunoassay system was established for high throughput analysis of clenbuterol. This system consisted of an integrated microchip and a linear confocal laser induced fluorescence (LIF) scanner. The microchip was composed of three layers: a fluidic channel layer, a PDMS membrane layer and a pneumatic control layer. The multi-layer chip was integrated with 36 pneumatic micro-valves and multiple micro-pumps to realize the flexible reagent delivery, facilitating the automatic assays with less consumption of samples and reduced analysis time. The homemade LIF scanner was able to simultaneously detect multi-channels and provide the potential capability of high throughput assays. The performance of the system was demonstrated by the determination of clenbuterol, one of the most widely used beta-agonists. Under the optimal conditions, the linear range and the limit of detection of clenbuterol were 0 similar to 5.0 ng mL(-1) and 0.088 ng mL(-1), respectively. The recovery rates determined with pig urine samples of 1.0 ng mL(-1) and 2.0 ng mL(-1) were 98.74% and 102.51% (n = 3), respectively. The total detection time was less than 30 min. The system had the potential application for rapid detection of multiple beta-agonists in clinical, pharmaceutical and chemical analyses.
引用
收藏
页码:1541 / 1547
页数:7
相关论文
共 36 条
[1]   Simultaneous high-throughput determination of clenbuterol, ambroxol and bromhexine in pharmaceutical formulations by HPLC with potentiometric detection [J].
Bazylak, G ;
Nagels, LJ .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2003, 32 (4-5) :887-903
[2]   Microfabricated bioprocessor for integrated nanoliter-scale Sanger DNA sequencing [J].
Blazej, Robert G. ;
Kumaresan, Palani ;
Mathies, Richard A. .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2006, 103 (19) :7240-7245
[3]   Screening and confirmation analysis of anabolic agents in human urine by gas chromatography hybrid mass spectrometry (high resolution time of flight) [J].
Buiarelli, F ;
Cartoni, GP ;
Amendola, L ;
Botrè, F .
ANALYTICA CHIMICA ACTA, 2001, 447 (1-2) :75-88
[4]   Total nucleic acid analysis integrated on microfluidic devices [J].
Chen, Lin ;
Manz, Andreas ;
Day, Philip J. R. .
LAB ON A CHIP, 2007, 7 (11) :1413-1423
[5]   ANABOLIC EFFECTS OF CLENBUTEROL ON SKELETAL-MUSCLE ARE MEDIATED BY BETA(2)-ADRENOCEPTOR ACTIVATION [J].
CHOO, JJ ;
HORAN, MA ;
LITTLE, RA ;
ROTHWELL, NJ .
AMERICAN JOURNAL OF PHYSIOLOGY, 1992, 263 (01) :E50-E56
[6]   DETERMINATION OF CLENBUTEROL IN BOVINE-TISSUES AND URINE BY ENZYME-IMMUNOASSAY [J].
DEGAND, G ;
BERNESDUYCKAERTS, A ;
MAGHUINROGISTER, G .
JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 1992, 40 (01) :70-75
[7]   THE ABUSE OF DOPING AGENTS IN COMPETING BODY-BUILDERS IN FLANDERS (1988-1993) [J].
DELBEKE, FT ;
DESMET, N ;
DEBACKERE, M .
INTERNATIONAL JOURNAL OF SPORTS MEDICINE, 1995, 16 (01) :66-70
[8]   Development of a diphasic dialysis method for the extraction/purification of residues of ethinylestradiol in hair of cattle, and determination by gas chromatography-tandem mass spectrometry [J].
Durant, AA ;
Fente, CA ;
Franco, CM ;
Vázquez, BI ;
Mayo, S ;
Cepeda, A .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 766 (02) :251-256
[9]   OBSERVATIONS ON THE EFFECTS OF LONG-TERM WITHDRAWAL ON CARCASS COMPOSITION AND RESIDUE CONCENTRATIONS IN CLENBUTEROL-MEDICATED CATTLE [J].
ELLIOTT, CT ;
CROOKS, SRH ;
MCEVOY, JGD ;
MCCAUGHEY, WJ ;
HEWITT, SA ;
PATTERSON, D ;
KILPATRICK, D .
VETERINARY RESEARCH COMMUNICATIONS, 1993, 17 (06) :459-468
[10]  
Hahnau S, 1996, FOOD ADDIT CONTAM, V13, P259, DOI 10.1080/02652039609374408