被引:68
作者
Trajanoski, Z
Brunner, GA
Gfrerer, RJ
Wach, P
Pieber, TR
机构
[1] LUDWIG BOLTZMANN INST TECH LEBENSHILFEN, GRAZ, AUSTRIA
[2] KARL FRANZENS UNIV GRAZ, DEPT INTERNAL MED, GRAZ, AUSTRIA
关键词
D O I
10.2337/diacare.19.12.1412
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE - To evaluate the accuracy of home blood glucose meters during hypoglycemia. METHODS - Six blood glucose meter-One Touch II (LifeScan, Milpitas, CA), Companion II (Medisense, Cambridge, U.K.), Reflolux (Boehringer Mannheim, Mannheim, Germany), Accutrend (Boehringer Mannheim), Elite (Bayer, Munich, Germany), and HemoCue (HemoCue, Angelholm, Sweden)-were compared with a reference method (Beekman Glucose Analyzer 2). Glucose concentrations from arterialized venous blood samples were measured using all glucose meters (whole blood) and the reference method (plasma) during hypoglycemic-hyperinsulinemic clamps in 15 subjects. RESULTS - In total, 663 Mood glucose monitor readings and 119 reference values ranging from 2,28 to 3.89 mmol/l were analyzed. The correlation coefficients and the percentage of measurements within 20% and outside 40% of the reference values for each glucose meter were as follows: One Touch II: 0.91, 99.2% and 0%; Companion II: 0.81, 88.2%, and 2.5%; Reflolux: 0.78, 85.0% and 0.9%; Accutrend: 0.88, 46.0% and 6.6%; Elite: 0.78, 75.6% and 4.2%; and HemoCue: 0.93, 96.6% and 0% (P < 0.001). CONCLUSIONS - There were substantial differences between the blood glucose meters during hypoglycemia, and none of the devices met the latest criteria recommended by the American Diabetes Association.
引用
收藏
页码:1412 / 1415
页数:4
相关论文
共 16 条
[1]  
*AM DIAB ASS, 1987, DIABETES CARE, V10, P93
[2]  
ASHWORTH L, 1992, CLIN CHEM, V38, P1479
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]  
BOLLI GB, 1995, INT TXB DIABETES MEL, P1131
[5]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[6]   COMPARISON OF 5 GLUCOSE METERS FOR SELF-MONITORING OF BLOOD-GLUCOSE BY DIABETIC-PATIENTS [J].
GIFFORDJORGENSEN, RA ;
BORCHERT, J ;
HASSANEIN, R ;
TILZER, L ;
EAKS, GA ;
MOORE, WV .
DIABETES CARE, 1986, 9 (01) :70-76
[7]   ARTERIAL, ARTERIALIZED VENOUS, VENOUS AND CAPILLARY BLOOD-GLUCOSE MEASUREMENTS IN NORMAL MAN DURING HYPERINSULINEMIC EUGLYCEMIA AND HYPOGLYCEMIA [J].
LIU, D ;
MOBERG, E ;
KOLLIND, M ;
LINS, PE ;
ADAMSON, U ;
MACDONALD, IA .
DIABETOLOGIA, 1992, 35 (03) :287-290
[8]   HOW ACCURATE ARE HOME BLOOD-GLUCOSE METERS WITH SPECIAL RESPECT TO THE LOW GLYCEMIC RANGE [J].
MOBERG, E ;
LUNDBLAD, S ;
LINS, PE ;
ADAMSON, U .
DIABETES RESEARCH AND CLINICAL PRACTICE, 1993, 19 (03) :239-243
[9]  
MOBERG E, 1994, THESIS KAROLINSKA I
[10]  
MOLITCH ME, 1994, DIABETES CARE, V17, P81