Transapical Transcatheter Aortic Valve Implantation Using the JenaValve: A One-Year Follow-up

被引:7
作者
Reuthebuch, Oliver [1 ]
Koechlin, Luca [1 ]
Kaufmann, Beat A. [2 ]
Kessel-Schaefer, Arnheid [2 ]
Gahl, Brigitta [3 ]
Eckstein, Friedrich S. [1 ]
机构
[1] Univ Basel Hosp, Dept Cardiac Surg, CH-4031 Basel, Switzerland
[2] Univ Basel Hosp, Dept Cardiol, CH-4031 Basel, Switzerland
[3] Univ Basel Hosp, Dept Cardiovasc Surg, CH-4031 Basel, Switzerland
关键词
aortic valve and root; heart valve; transapical; TAVI; outcomes; SINGLE-CENTER EXPERIENCE; REGURGITATION; STENOSIS;
D O I
10.1055/s-0035-1552980
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective Since the first transcatheter aortic valve implantation (TAVI) in 2002, TAVI technique has gained an increasing popularity especially in high-risk patients. In this study, we present the first echocardiographic midterm outcome with the second-generation transapical JenaValve TAVI system (JenaValve Technology GmbH, Munich, Germany) in patients with aortic stenosis (AS). Methods Between November 2011 and November 2012, a total of 28 patients received transapical TAVI using the JenaValve. Primary endpoint was a combined efficacy endpoint after 1year, which included all-cause mortality after more than 30 days, failure of current therapy for AS requiring hospitalization for symptoms of valve-related cardiac decompensation or prosthetic heart valve dysfunction. Moreover, we analyzed secondary endpoints after 3 and 12 months including cardiovascular mortality; major stroke; and life-threatening, disabling, or major bleeding. Mean echocardiographic follow-up was 471.35102.72 days. Results Mean age was 80.43 +/- 6.03 years and EuroSCORE II was 8.80 +/- 7.21%. Successful implantation was accomplished in 100% (n=28). Median transvalvular aortic mean pressure gradient was 44.5mm Hg (interquartile range [IQR]: 34.5; 55.5) preoperatively, 12mm Hg (IQR: 9; 16) postoperatively, and 11mm Hg (IQR: 8; 16) after 1 year. After 12 months, no paravalvular leakage was seen in 52.38% of the patients and grade 1 paravalvular leakage was seen in 47.62% of the patients. There was no grade 2 or 3 leakage detected. Stroke, valve thrombosis or dislocation, myocardial infarction, or bleeding was also not observed. However, criteria for the combined efficacy endpoint after 1year were met in five patients (17.86%). Thirty-day mortality was 14.29% (n=4) and all-cause mortality after 1year was 21.43% (n=6). Conclusion The JenaValve transapical TAVI system is a safe and feasible procedure with low peri- and postoperative complications and convincing midterm performance of the prosthesis.
引用
收藏
页码:493 / 500
页数:8
相关论文
共 50 条
  • [31] One-year safety and efficacy profile of transcatheter aortic valve-in-valve implantation with the portico system
    Casenghi, Matteo
    Gorla, Riccardo
    Popolo Rubbio, Antonio
    De Marco, Federico
    Brambilla, Nedy
    Agnifili, Mauro
    Testa, Luca
    Bedogni, Francesco
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2021, 98 (01) : E145 - E152
  • [32] Transapical transcatheter aortic valve implantation for predominant aortic regurgitation with a self-expandable valve
    Liu, Huan
    Liu, Shun
    Lu, Yuntao
    Yang, Ye
    Wang, Wenshuo
    Zhu, Liming
    Wei, Lai
    Wang, Chunsheng
    JOURNAL OF THORACIC DISEASE, 2020, 12 (03) : 538 - 549
  • [33] Case report: Transcatheter tricuspid valve-in-valve implantation using novel balloon-expandable aortic valve with 1 year follow-up
    Mussayev, Abdurashid
    Alimbayev, Serik
    Tanaliev, Nursultan
    Kuanyshbek, Aidyn
    Marat, Aripov
    Lesbekov, Timur
    Raissov, Yerkezhan
    Sadykova, Aigerim
    Kamila, Askarovna Kenzhebayeva
    Mukarov, Murat
    Pya, Yuriy
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2023, 10
  • [34] Implantation of a diffractive trifocal intraocular lens: One-year follow-up
    Mojzis, Peter
    Majerova, Katarina
    Hrckova, Lucia
    Pinero, David P.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (08) : 1623 - 1630
  • [35] Paravalvular Regurgitation One Year After Transcatheter Aortic Valve Implantation
    Rajani, Ronak
    Kakad, Mitesh
    Khawaja, Muhammed Z.
    Lee, Lorraine
    James, Rachael
    Saha, Mrinal
    Hildick-Smith, David
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2010, 75 (06) : 868 - 872
  • [36] Follow up Management after Transcatheter Aortic Valve Implantation
    Fortmeier, Vera
    Rudolph, Tanja K.
    DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 2022, 147 (16) : 1047 - +
  • [37] Transcatheter aortic valve-in-valve implantation of a CoreValve in a JenaValve prosthesis: A case report
    Lotfi S.
    Becker M.
    Moza A.
    Autschbach R.
    Marx N.
    Schröder J.
    Journal of Medical Case Reports, 11 (1)
  • [38] Left Ventricular Mass Regression One Year After Transcatheter Aortic Valve Implantation
    Tzikas, Apostolos
    Geleijnse, Marcel L.
    Van Mieghem, Nicolas M.
    Schultz, Carl J.
    Nuis, Rutger-Jan
    van Dalen, Bas M.
    Sarno, Giovanna
    van Domburg, Ron T.
    Serruys, Patrick W.
    de Jaegere, Peter P. T.
    ANNALS OF THORACIC SURGERY, 2011, 91 (03) : 685 - 691
  • [39] Transcatheter aortic valve implantation in very elderly patients: immediate results and medium term follow-up
    Pascual, Isaac
    Munoz-Garcia, Antonio J.
    Lopez-Otero, Diego
    Avanzas, Pablo
    Jimenez-Navarro, Manuel F.
    Cid-Alvarez, Belen
    del Valle, Raquel
    Alonso-Briales, Juan H.
    Ocaranza-Sanchez, Raimundo
    Hernandez, Jose M.
    Trillo-Nouche, Ramiro
    Moris, Cesar
    JOURNAL OF GERIATRIC CARDIOLOGY, 2015, 12 (04) : 340 - 345
  • [40] Transapical transcatheter aortic valve implantation: 1-year outcome in 26 patients
    Ye, Jian
    Cheung, Anson
    Lichtenstein, Samuel V.
    Altwegg, Lukas A.
    Wong, Daniel R.
    Carere, Ronald G.
    Thompson, Christopher R.
    Moss, Robert R.
    Munt, Brad
    Pasupati, Sanjeevan
    Boone, Robert H.
    Masson, Jean-Bernard
    Al Ali, Abdullah
    Webb, John G.
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2009, 137 (01) : 167 - 173