Changing pulsatility by delaying the rotational speed phasing of a rotary left ventricular assist device

被引:12
作者
Date, Kazuma [1 ,2 ]
Nishimura, Takashi [3 ]
Arakawa, Mamoru [1 ]
Takewa, Yoshiaki [1 ]
Kishimoto, Satoru [1 ]
Umeki, Akihide [2 ]
Ando, Masahiko [2 ]
Mizuno, Toshihide [1 ]
Tsukiya, Tomonori [1 ]
Ono, Minoru [2 ]
Tatsumi, Eisuke [1 ]
机构
[1] Natl Cerebral & Cardiovasc Ctr Res Inst, Dept Artificial Organs, 5-7-1 Fujishiro Dai, Suita, Osaka 5658565, Japan
[2] Univ Tokyo, Dept Cardiothorac Surg, Tokyo, Japan
[3] Tokyo Metropolitan Geriatr Hosp, Dept Cardiac Surg, Itabashi Ku, 35-2 Sakaecho, Tokyo 1730015, Japan
关键词
Continuous-flow LVAD; Synchronization with cardiac beat; Rotary pump; LEFT-HEART BYPASS; BEAT SYNCHRONIZING SYSTEM; FLOW; SUPPORT; PERFUSION; PRESSURE; RECOVERY; MODEL;
D O I
10.1007/s10047-016-0920-y
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Continuous-flow left ventricular assist devices (LVADs) have improved the prognosis of end-stage heart failure. However, continuous-flow LVADs diminish pulsatility, which possibly result in bleeding, aortic insufficiency, and other adverse effects. We previously developed a novel control system for a continuous-flow LVAD (EVAHEART(A (R)); Sun Medical), and demonstrated that we could create sufficient pulsatility by increasing its rotational speed (RS) in the systolic phase (Pulsatile Mode) in the normal heart model. Here, we aimed to evaluate differences between systolic assist with advanced and delayed loads by shifting the timing of increased RS. We implanted EVAHEART in six goats (55.3 +/- 4.3 kg) with normal hearts. We reduced their heart rates to < 60 bpm using propranolol and controlled the heart rates at 80 and 120 bpm using ventricular pacing. We shifted the timing of increasing RS from -60 to +60 ms in the systolic phase. We found significant increases in all the following parameters when assessments of delayed timing (+60 ms) were compared with assessments of advanced timing (-60 ms): pulse pressure, mean dP/dt max of aortic pressure, and energy-equivalent pulse pressure. During continuous-flow LVAD support, pulsatility can be controlled using a rotary pump. In particular, pulsatility can be shifted by delaying increased RS.
引用
收藏
页码:18 / 25
页数:8
相关论文
共 50 条
  • [31] The left ventricular assist device as a patient monitoring system
    Moscato, Francesco
    Gross, Christoph
    Maw, Martin
    Schloglhofer, Thomas
    Granegger, Marcus
    Zimpfer, Daniel
    Schima, Heinrich
    [J]. ANNALS OF CARDIOTHORACIC SURGERY, 2021, 10 (02) : 221 - 232
  • [32] Medical Management of Patients With a Left Ventricular Assist Device for the Non-Left Ventricular Assist Device Specialist
    DeVore, Adam D.
    Patel, Priyesh A.
    Patel, Chetan B.
    [J]. JACC-HEART FAILURE, 2017, 5 (09) : 621 - 631
  • [33] Anthracycline Treatment and Ventricular Remodeling in Left Ventricular Assist Device Patients
    Segura, Ana Maria
    Radovancevic, Rajko
    Demirozu, Zumrat T.
    Frazier, O. H.
    Buja, L. Maximilian
    [J]. TEXAS HEART INSTITUTE JOURNAL, 2014, 42 (02): : 124 - 130
  • [34] Postoperative Pulmonary Artery Pulsatility Index Improves Prediction of Right Ventricular Failure After Left Ventricular Assist Device Implantation
    Wei, Johnny
    Franke, Jack
    Kee, Abigail
    Dukes, Rachel
    Leonardo, Vincent
    Flynn, Brigid C.
    [J]. JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2024, 38 (01) : 214 - 220
  • [35] Enhancement of Arterial Pressure Pulsatility by Controlling Continuous-Flow Left Ventricular Assist Device Flow Rate in Mock Circulatory System
    Bozkurt, Selim
    van de Vosse, Frans N.
    Rutten, Marcel C. M.
    [J]. JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2016, 36 (03) : 308 - 315
  • [36] Assessment of myocardial viability and left ventricular function in patients supported by a left ventricular assist device
    Gupta, Deepak K.
    Skali, Hicham
    Rivero, Jose
    Campbell, Patricia
    Griffin, Leslie
    Smith, Colleen
    Foster, Courtney
    Claggett, Brian
    Glynn, Robert J.
    Couper, Gregory
    Givertz, Michael M.
    Mehra, Mandeep R.
    Di Carli, Marcelo
    Solomon, Scott D.
    Pfeffer, Marc A.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2014, 33 (04) : 372 - 381
  • [37] Echo-guided left ventricular assist device speed optimisation for exercise maximisation
    Stapor, Maciej
    Pilat, Adam
    Gackowski, Andrzej
    Misiuda, Agnieszka
    Gorkiewicz-Kot, Izabela
    Kaleta, Michal
    Kleczynski, Pawel
    Zmudka, Krzysztof
    Legutko, Jacek
    Kapelak, Boguslaw
    Wierzbicki, Karol
    [J]. HEART, 2022, 108 (13) : 1055 - 1062
  • [38] The diagnosis of left ventricular assist device thrombosis
    Gerds, H. Z. R.
    Brugemann, J.
    Rienstra, M.
    Erasmus, M. E.
    [J]. NETHERLANDS HEART JOURNAL, 2015, 23 (7-8) : 389 - 391
  • [39] Left Ventricular Assist Device Management and Complications
    Birati, Edo Y.
    Rame, J. Eduardo
    [J]. CRITICAL CARE CLINICS, 2014, 30 (03) : 607 - +
  • [40] Percutaneous Decommissioning of Left Ventricular Assist Device
    Soon, Jia-Lin
    Tan, Ju-Le
    Lim, Choon-Pin
    Tan, Teing-Ee
    Tan, Swee-Yaw
    Kerk, Ka-Lee
    Sim, Kheng-Leng David
    Sivathasan, Cumaraswamy
    [J]. HEART LUNG AND CIRCULATION, 2018, 27 (07) : 853 - 855