Speckle averaging for optical coherence tomography by vibration of a thin water film

被引:1
作者
Park, BH [1 ]
Pierce, MC [1 ]
Cense, B [1 ]
de Boer, JF [1 ]
机构
[1] Massachusetts Gen Hosp, Wellman Labs Photomed, Boston, MA 02114 USA
来源
COHERENCE DOMAIN OPTICAL METHODS AND OPTICAL COHERENCE TOMOGRAPHY IN BIOMEDICINE VIII | 2004年 / 5316卷
关键词
speckle; reduction; OCT;
D O I
10.1117/12.531400
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A novel method for reducing the appearance of speckle within a single image acquisition in optical coherence tomography is presented. Local variations in the optical path length of beamlets within the beam diameter are introduced by vibrating a thin water film placed in a collimated section of the sample arm beam path. Consequent averaging of small numbers of adjacent depth scans results in a reduced appearance of speckle with no change in the axial point spread function. Reduced speckle images of a layered agar sample and thin reflective surfaces are shown.
引用
收藏
页码:391 / 396
页数:6
相关论文
共 6 条
[1]   Statistics and reduction of speckle in optical coherence tomography [J].
Bashkansky, M ;
Reintjes, J .
OPTICS LETTERS, 2000, 25 (08) :545-547
[2]  
Goodman J., 2000, Statistical Optics
[3]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[4]   Speckle reduction in optical coherence tomography by "path length encoded" angular compounding [J].
Iftimia, N ;
Bouma, BE ;
Tearney, GJ .
JOURNAL OF BIOMEDICAL OPTICS, 2003, 8 (02) :260-263
[5]   Simultaneous intensity, birefringence, and flow measurements with high-speed fiber-based optical coherence tomography [J].
Pierce, MC ;
Park, BH ;
Cense, B ;
de Boer, JF .
OPTICS LETTERS, 2002, 27 (17) :1534-1536
[6]   Array detection for speckle reduction in optical coherence microscopy [J].
Schmitt, JM .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (07) :1427-1439