Improvement of accuracy in ultrasonic measurement of luminal surface roughness of carotid arterial wall by deconvolution filtering

被引:17
作者
Nagai, Yoshifumi [1 ]
Hasegawa, Hideyuki [1 ,2 ]
Kanai, Hiroshi [1 ,2 ]
机构
[1] Tohoku Univ, Grad Sch Biomed Engn, Sendai, Miyagi 9808579, Japan
[2] Tohoku Univ, Grad Sch Engn, Sendai, Miyagi 9808579, Japan
关键词
INTIMA-MEDIA THICKNESS; NONINVASIVELY IN-VIVO; BLIND DECONVOLUTION; PHASE TRACKING; ATHEROSCLEROSIS; IMAGES; ECHO; VALIDATION; ALGORITHM; DISEASE;
D O I
10.7567/JJAP.53.07KF19
中图分类号
O59 [应用物理学];
学科分类号
摘要
The diagnosis of early-stage atherosclerosis is important for preventing cardiovascular diseases such as a stroke or a heart attack. The main cause of such diseases is atherosclerosis. In early-stage atherosclerosis, the luminal surface of the arterial wall becomes rough because of the detachment of endothelial cells and the degeneration of the internal elastic layer. Therefore, it would be useful if the minute luminal surface roughness of the carotid arterial wall observed in the early stage of atherosclerosis can be measured noninvasively by ultrasonography. In conventional ultrasonography, the axial resolution of a B-mode image depends on the ultrasonic pulse length of 150 mu m (at 10 MHz) because a B-mode image is constructed using the amplitude of the RF echo. However, the thickness of an endothelial cell covering the luminal surface is 10-20 mu m. Therefore, a minute surface roughness cannot be measured from a conventional B-mode image. We have realized the transcutaneous measurement of such a minute surface roughness of about 10 mu m using the phased-tracking method. However, the lateral spatial resolution degrades owing to the point spread function (PSF) because the diameter of an ultrasonic beam is finite. In the present study, we proposed a method of improving the lateral spatial resolution in the estimation of surface roughness using a Wiener filter to suppress the effect of the PSF of the ultrasound system employed. The proposed method was validated using a phantom and then applied to the in vivo measurement of carotid arteries. (C) 2014 The Japan Society of Applied Physics
引用
收藏
页数:9
相关论文
共 42 条
[1]  
ANDREW H, 1997, DIGITAL IMAGE RESTOR
[2]   Accurate ultrasonic measurement of surface profile using phase shift of echo and inverse filtering [J].
Arihara, Chihiro ;
Hasegawa, Hideyuki ;
Kanai, Hiroshi .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 2006, 45 (5B) :4727-4731
[3]   NONINVASIVE DETECTION OF ENDOTHELIAL DYSFUNCTION IN CHILDREN AND ADULTS AT RISK OF ATHEROSCLEROSIS [J].
CELERMAJER, DS ;
SORENSEN, KE ;
GOOCH, VM ;
SPIEGELHALTER, DJ ;
MILLER, OI ;
SULLIVAN, ID ;
LLOYD, JK ;
DEANFIELD, JE .
LANCET, 1992, 340 (8828) :1111-1115
[4]   METHODS FOR ESTIMATION OF SUBSAMPLE TIME DELAYS OF DIGITIZED ECHO SIGNALS [J].
CESPEDES, I ;
HUANG, Y ;
OPHIR, J ;
SPRATT, S .
ULTRASONIC IMAGING, 1995, 17 (02) :142-171
[5]   Evaluation of an ultrasonic echo-tracking method for measurements of arterial wall movements in two dimensions [J].
Cinthio, M ;
Ahlgren, AR ;
Janssen, T ;
Eriksson, A ;
Persson, HW ;
Lindström, K .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2005, 52 (08) :1300-1311
[6]   Minute roughness measurement using phase tracking for arterial wall diagnosis non-invasively in vivo [J].
Cinthio, Magnus ;
Hasegawa, Hideyuki ;
Kanai, Hiroshi .
2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, :997-+
[7]   Initial Phantom Validation of Minute Roughness Measurement Using Phase Tracking for Arterial Wall Diagnosis Non-Invasively In Vivo [J].
Cinthio, Magnus ;
Hasegawa, Hideyuki ;
Kanai, Hiroshi .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2011, 58 (04) :853-857
[8]  
Fienup J. R., 2002, P SOC PHOTO-OPT INS, V4792
[9]   Carotid artery wall motion estimated from B-mode ultrasound using region tracking and block matching [J].
Golemati, S ;
Sassano, A ;
Lever, MJ ;
Bharath, AA ;
Dhanjil, S ;
Nicolaides, AN .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2003, 29 (03) :387-399
[10]   Ultra-wideband radar noise reduction for target classification [J].
Hantscher, S. ;
Reisenzahn, A. ;
Diskus, C. G. .
IET RADAR SONAR AND NAVIGATION, 2008, 2 (04) :315-322