Effect of acetabular cup design on metal ion release in two designs of metal-on-metal hip resurfacing

被引:3
|
作者
Cadossi, Matteo [1 ,2 ]
Tedesco, Giuseppe [1 ,2 ]
Savarino, Lucia [3 ]
Baldini, Nicola [2 ,3 ]
Mazzotti, Antonio [1 ,2 ]
Greco, Michelina [3 ]
Giannini, Sandro [1 ,2 ]
机构
[1] Rizzoli Orthopaed Inst, Dept Orthopaed & Traumatol 1, Bologna, Italy
[2] Univ Bologna, Bologna, Italy
[3] Rizzoli Orthopaed Inst, Complex Orthoped Trauma Pathol Dept, Lab Orthopaed Pathophysiol & Regenerat Med, Bologna, Italy
关键词
metal-on-metal; metal ions; hip resurfacing; acetabular notch; prosthesis; FOLLOW-UP; ARTHROPLASTY; REPLACEMENT; PROSTHESIS; JOINT; FRACTURES; REGISTRY; ALLOYS; CANCER; DIFFER;
D O I
10.1002/jbm.b.33191
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this observational prospective cohort study was to evaluate the serum concentrations of cobalt (Co), chromium (Cr), and nickel (Ni) at a 2-year follow-up in patients operated on with a novel design of hip resurfacing: Romax resurfacing system (RRS). RRS is characterized by the presence of an acetabular notch which theoretically provides a wider range of motion and a reduced incidence of groin pain. The presence of radiolucencies and functional outcome, assessed using the Harris hip score (HHS) and the University of California Activity scale (UCLA), were secondary endpoints. Moreover, these results were compared with those obtained in our previous study from a similar cohort of patients implanted using the Birmingham Hip Resurfacing (BHR) system. At a 2-year follow-up, the serum levels of Co in patients operated on using the RRS were five times higher (p = 0.0002) than those found before surgery (Co, means: 1.04 and 0.20 ng/mL, respectively); similarly, Cr levels were 13 times higher (p<0.0001) at a 2-year follow-up than before surgery (Cr, means: 1.69 and 0.13 ng/mL, respectively). Ni concentrations (0.42 and 0.78 ng/mL) were not significantly different (p = 0.16), even if they increased 86% after surgery. In the RRS patients, an inverse correlation was found between Co and Cr concentrations and length of follow-up (Co: r = -0.64, p = 0.0096; Cr: r = -0.45, p = 0.08). The serum levels of Co and Cr were not significantly different between RRS (Co: 1.04 ng/mL and Cr: 1.69 ng/mL) and BHR (Co: 1.39 ng/mL and Cr: 2.30 ng/mL) patients at 2 years (p = 0.95 and 0.26 for Co and Cr, respectively). Our results showed that RRS patients achieved an excellent clinical outcome with limited metal ion release. (C) 2014 Wiley Periodicals, Inc.
引用
收藏
页码:1595 / 1601
页数:7
相关论文
共 50 条
  • [21] The development of the Durom™ metal-on-metal hip resurfacing
    Grigoris, P.
    Roberts, P.
    Panousis, K.
    HIP INTERNATIONAL, 2006, 16 (02) : S65 - S72
  • [22] The safety and efficacy of one-stage bilateral metal-on-metal hip resurfacing arthroplasty
    Nam, Denis
    Rebolledo, Brian J.
    Su, Edwin P.
    HIP INTERNATIONAL, 2012, 22 (01) : 100 - 106
  • [23] Metal-on-metal hip resurfacing: a critical review
    Gavin J. Macpherson
    Steffen J. Breusch
    Archives of Orthopaedic and Trauma Surgery, 2011, 131 : 101 - 110
  • [24] Management Guidelines for Metal-on-metal Hip Resurfacing Arthroplasty: A Strategy on Followup
    Nakano, Naoki
    Volpin, Andrea
    Bartlett, Jonathan
    Khanduja, Vikas
    INDIAN JOURNAL OF ORTHOPAEDICS, 2017, 51 (04) : 414 - 420
  • [25] Metal ion levels after metal-on-metal hip resurfacing: A single-center longitudinal study of Chinese patients
    Wan, Yanlin
    You, Jia
    Jiang, Wenxue
    Wang, Pengfei
    Xia, Qun
    ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2024, 58 (04) : 196 - 202
  • [26] Treatment of Pseudotumors After Metal-on-Metal Hip Resurfacing Based on Magnetic Resonance Imaging, Metal Ion Levels and Symptoms
    van der Weegen, Walter
    Sijbesma, Thea
    Hoekstra, Henk J.
    Brakel, Koen
    Pilot, Peter
    Nelissen, Rob G. H. H.
    JOURNAL OF ARTHROPLASTY, 2014, 29 (02) : 416 - 421
  • [27] Migration pattern of a cobalt-chrome monoblock acetabular component after metal-on-metal hip resurfacing
    Roda, Richard
    Whittingham-Jones, Paul M.
    Lee, Joshua
    Ryu, Jae-Jin
    Beaule, Paul E.
    HIP INTERNATIONAL, 2016, 26 (03) : 220 - 225
  • [28] Metal ions levels between metal-on-metal total hip arthroplasty and metal-on-metal hip resurfacing arthroplasty: a meta-analysis
    Zhang, Liang
    Chen, Zhirong
    Li, Lifang
    Li, Peng
    Lu, Zhidong
    Pang, Long
    Song, Kening
    Li, Yaping
    Jin, Qunhua
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2018, 11 (02): : 453 - 462
  • [29] Metal-on-metal hip resurfacing in patients younger than 50 years: a retrospective analysis
    Gaillard, Melissa D.
    Gross, Thomas P.
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2017, 12
  • [30] Imaging of Metal-on-Metal Hip Resurfacing
    Hayter, Catherine L.
    Potter, Hollis G.
    Su, Edwin P.
    ORTHOPEDIC CLINICS OF NORTH AMERICA, 2011, 42 (02) : 195 - +