Effect of acetabular cup design on metal ion release in two designs of metal-on-metal hip resurfacing

被引:3
作者
Cadossi, Matteo [1 ,2 ]
Tedesco, Giuseppe [1 ,2 ]
Savarino, Lucia [3 ]
Baldini, Nicola [2 ,3 ]
Mazzotti, Antonio [1 ,2 ]
Greco, Michelina [3 ]
Giannini, Sandro [1 ,2 ]
机构
[1] Rizzoli Orthopaed Inst, Dept Orthopaed & Traumatol 1, Bologna, Italy
[2] Univ Bologna, Bologna, Italy
[3] Rizzoli Orthopaed Inst, Complex Orthoped Trauma Pathol Dept, Lab Orthopaed Pathophysiol & Regenerat Med, Bologna, Italy
关键词
metal-on-metal; metal ions; hip resurfacing; acetabular notch; prosthesis; FOLLOW-UP; ARTHROPLASTY; REPLACEMENT; PROSTHESIS; JOINT; FRACTURES; REGISTRY; ALLOYS; CANCER; DIFFER;
D O I
10.1002/jbm.b.33191
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this observational prospective cohort study was to evaluate the serum concentrations of cobalt (Co), chromium (Cr), and nickel (Ni) at a 2-year follow-up in patients operated on with a novel design of hip resurfacing: Romax resurfacing system (RRS). RRS is characterized by the presence of an acetabular notch which theoretically provides a wider range of motion and a reduced incidence of groin pain. The presence of radiolucencies and functional outcome, assessed using the Harris hip score (HHS) and the University of California Activity scale (UCLA), were secondary endpoints. Moreover, these results were compared with those obtained in our previous study from a similar cohort of patients implanted using the Birmingham Hip Resurfacing (BHR) system. At a 2-year follow-up, the serum levels of Co in patients operated on using the RRS were five times higher (p = 0.0002) than those found before surgery (Co, means: 1.04 and 0.20 ng/mL, respectively); similarly, Cr levels were 13 times higher (p<0.0001) at a 2-year follow-up than before surgery (Cr, means: 1.69 and 0.13 ng/mL, respectively). Ni concentrations (0.42 and 0.78 ng/mL) were not significantly different (p = 0.16), even if they increased 86% after surgery. In the RRS patients, an inverse correlation was found between Co and Cr concentrations and length of follow-up (Co: r = -0.64, p = 0.0096; Cr: r = -0.45, p = 0.08). The serum levels of Co and Cr were not significantly different between RRS (Co: 1.04 ng/mL and Cr: 1.69 ng/mL) and BHR (Co: 1.39 ng/mL and Cr: 2.30 ng/mL) patients at 2 years (p = 0.95 and 0.26 for Co and Cr, respectively). Our results showed that RRS patients achieved an excellent clinical outcome with limited metal ion release. (C) 2014 Wiley Periodicals, Inc.
引用
收藏
页码:1595 / 1601
页数:7
相关论文
共 37 条
[21]   Metallosis after contemporary metal-on-metal total hip arthroplasty - Five to nine-year follow-up [J].
Korovessis, P. ;
Petsinis, G. ;
Repanti, M. ;
Repantis, T. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (06) :1183-1191
[22]   Femoral offset: Anatomical concept, definition, assessment, implications for preoperative templating and hip arthroplasty [J].
Lecerf, G. ;
Fessy, M. H. ;
Philippot, R. ;
Massin, P. ;
Giraud, F. ;
Flecher, X. ;
Girard, J. ;
Mertl, P. ;
Marchetti, E. ;
Stindel, E. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2009, 95 (03) :210-219
[23]   Do Large Heads Enhance Stability and Restore Native Anatomy in Primary Total Hip Arthroplasty? [J].
Lombardi, Adolph V., Jr. ;
Skeels, Michael D. ;
Berend, Keith R. ;
Adams, Joanne B. ;
Franchi, Orlando J. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (06) :1547-1553
[24]   Risk of cancer with metal-on-metal hip replacements: population based study [J].
Makela, Keijo T. ;
Visuri, Tuomo ;
Pulkkinen, Pekka ;
Eskelinen, Antti ;
Remes, Ville ;
Virolainen, Petri ;
Junnila, Mika ;
Pukkala, Eero .
BMJ-BRITISH MEDICAL JOURNAL, 2012, 345 :e4646
[25]   Does ion release differ between hip resurfacing and metal-on-metal THA? [J].
Moroni, Antonio ;
Savarino, Lucia ;
Cadossi, Matteo ;
Baldini, Nicola ;
Giannini, Sandro .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2008, 466 (03) :700-707
[26]   Is Patient Selection Important for Hip Resurfacing? [J].
Nunley, Ryan M. ;
Della Valle, Craig J. ;
Barrack, Robert L. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (01) :56-65
[27]   Iliopsoas tendonitis - A complication after total hip arthroplasty [J].
O'Sullivan, Michael ;
Tai, Cheh Chin ;
Richards, Simon ;
Skyrme, Andrew D. ;
Walter, William L. ;
Walter, William K. .
JOURNAL OF ARTHROPLASTY, 2007, 22 (02) :166-170
[28]  
Pailhe R, 2012, INT ORTHOP, V6, P2399
[29]  
Que L, 1998, S COB BAS ALL BIOM A
[30]   Do Ion Levels in Metal-on-metal Hip Resurfacing Differ From Those in Metal-on-metal THA at Long-term Followup? [J].
Savarino, Lucia ;
Cadossi, Matteo ;
Chiarello, Eugenio ;
Baldini, Nicola ;
Giannini, Sandro .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (09) :2964-2971