A Comparison of the Mechanical Measures Used for Assessing Orthodontic Mini-Implant Stability

被引:9
|
作者
Hosein, Yara K. [1 ]
Dixon, S. Jeffrey [2 ]
Rizkalla, Amin S. [3 ]
Tassi, Ali [4 ]
机构
[1] Western Univ, Bone & Joint Inst, Schulich Sch Med & Dent, Div Grad Orthodont, London, ON, Canada
[2] Western Univ, Bone & Joint Inst, Schulich Sch Med & Dent, Dept Physiol & Pharmacol,Div Grad Orthodont, London, ON, Canada
[3] Western Univ, Schulich Sch Med Dent, Dept Chem & Biochem Engn, Div Grad Orthodont,Bone & Joint Inst,Fac Engn, London, ON, Canada
[4] Western Univ, Div Grad Orthodont, Schulich Sch Med & Dent, London, ON, Canada
关键词
torque; pull-out load; implant loosening; Periotest; mechanical retention; PULL-OUT STRENGTH; INSERTION TORQUE; SURFACE-TREATMENT; DAMPING CAPACITY; TAPER SHAPE; MINISCREWS; LENGTH; THREAD; IMPACT; DIAMETER;
D O I
10.1097/ID.0000000000000514
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose:Mechanical loosening remains a common complication associated with mini-implant failure. The purpose of this study was to compare common mechanical measures of mini-implant stability to determine their association and reliability.Materials and Methods:Ninety self-drilling orthodontic mini-implants from 6 manufacturers were inserted into artificial bone blocks. Insertion torques (ITs) and Periotest values (PVs) were measured. Subsequently, mini-implants underwent pull-out testing for measures of pull-out load (POL) and screw displacement (ScrD). Stability measurements were compared using one-way ANOVA, associations among them were assessed using correlation analyses, and reliability was evaluated using coefficients of variation (COVs).Results:Variations in stability of mini-implants were found, specific to the mechanical measure used for assessment (P < 0.05). The strongest correlations were found between IT and PV (r = -0.68) and between IT and POL (r = 0.66). Overall, PV showed the greatest variability (COV: 11%-100%) compared with IT (11%), POL (4%), and ScrD (19%).Conclusions:IT, PV, and POLs only agreed moderately in their assessment of mini-implant stability, and Periotest showed the least reliability in predicting mini-implant stability. As such, independent and interchangeable use of these stability measures should be avoided.
引用
收藏
页码:225 / 231
页数:7
相关论文
共 50 条
  • [41] Three-dimensional digital imaging analysis of the palatal bone thickness for orthodontic mini-implant insertion - determination of the safe zone and angulation
    Schubert, Hanna
    Matta, Ragai
    Seidel, Anna
    Adler, Werner
    Wichmann, Manfred
    Kesting, Marco
    Lutz, Rainer
    BMC ORAL HEALTH, 2024, 24 (01):
  • [42] Impact of implant design on primary stability of orthodontic mini-implants
    Wilmes, Benedict
    Ottenstreuer, Stephanie
    Su, Yu-Yu
    Drescher, Dieter
    JOURNAL OF OROFACIAL ORTHOPEDICS-FORTSCHRITTE DER KIEFERORTHOPADIE, 2008, 69 (01): : 42 - 50
  • [43] Comparison of primary stability of used and unused self-tapping and self-drilling orthodontic mini-implants
    Baser, Baris
    Ozel, Mehmet Birol
    ADVANCES IN CLINICAL AND EXPERIMENTAL MEDICINE, 2024, 33 (05): : 483 - 489
  • [44] Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies
    Walter, Andre
    Winsauer, Heinz
    Marce-Nogue, Jordi
    Mojal, Sergi
    Puigdollers, Andreu
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2013, 18 (05): : E804 - E810
  • [45] Influence of bone density, screw size and surgical procedure on orthodontic mini-implant placement - part A: temperature development
    Moehlhenrich, S. C.
    Heussen, N.
    Modabber, A.
    Kniha, K.
    Hoelzle, F.
    Wilmes, B.
    Danesh, G.
    Szalma, J.
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2021, 50 (04) : 555 - 564
  • [46] Influence of different implant materials on the primary stability of orthodontic mini-implants
    Pan, Chin-Yun
    Chou, Szu-Ting
    Tseng, Yu-Chuan
    Yang, Yi-Hsin
    Wu, Chao-Yi
    Lan, Ting-Hsun
    Liu, Pao-Hsin
    Chang, Hong-Po
    KAOHSIUNG JOURNAL OF MEDICAL SCIENCES, 2012, 28 (12) : 673 - 678
  • [47] Suitability of virtual plaster models superimposed with the lateral cephalogram for guided paramedian orthodontic mini-implant placement with regard to the bone support
    Moehlhenrich, Stephan Christian
    Brandt, Maximilian
    Kniha, Kristian
    Bock, Anna
    Prescher, Andreas
    Hoelzle, Frank
    Modabber, Ali
    Danesh, Golamreza
    JOURNAL OF OROFACIAL ORTHOPEDICS-FORTSCHRITTE DER KIEFERORTHOPADIE, 2020, 81 (05): : 340 - 349
  • [48] Two-dimensional radiographic and clinical references of the tooth crown for orthodontic mini-implant insertion: A guide-free technique
    Barros, Sergio Estelita
    Janson, Guilherme
    Chiqueto, Kelly
    Garib, Daniela Gamba
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2010, 110 (05): : E8 - E16
  • [49] Influence of Chemically-Modified Implant Surfaces on the Stability of Orthodontic Mini-Implants
    Bratu, Dana Cristina
    Popa, George
    Petrescu, Horatiu
    Karancsi, Olimpiu Ladislau
    Bratu, Emanuel Adrian
    REVISTA DE CHIMIE, 2014, 65 (10): : 1222 - 1225
  • [50] Automatic Segmentation of the Jaws Used in Guided Insertion of Orthodontic Mini Implants to Improve Their Stability and Precision
    David, Andra Patricia
    Brad, Silviu
    Rusu, Laura-Cristina
    David, Ovidiu Tiberiu
    Samoila, Christian
    Leretter, Marius Traian
    MEDICINA-LITHUANIA, 2024, 60 (10):