A Comparison of the Mechanical Measures Used for Assessing Orthodontic Mini-Implant Stability

被引:9
|
作者
Hosein, Yara K. [1 ]
Dixon, S. Jeffrey [2 ]
Rizkalla, Amin S. [3 ]
Tassi, Ali [4 ]
机构
[1] Western Univ, Bone & Joint Inst, Schulich Sch Med & Dent, Div Grad Orthodont, London, ON, Canada
[2] Western Univ, Bone & Joint Inst, Schulich Sch Med & Dent, Dept Physiol & Pharmacol,Div Grad Orthodont, London, ON, Canada
[3] Western Univ, Schulich Sch Med Dent, Dept Chem & Biochem Engn, Div Grad Orthodont,Bone & Joint Inst,Fac Engn, London, ON, Canada
[4] Western Univ, Div Grad Orthodont, Schulich Sch Med & Dent, London, ON, Canada
关键词
torque; pull-out load; implant loosening; Periotest; mechanical retention; PULL-OUT STRENGTH; INSERTION TORQUE; SURFACE-TREATMENT; DAMPING CAPACITY; TAPER SHAPE; MINISCREWS; LENGTH; THREAD; IMPACT; DIAMETER;
D O I
10.1097/ID.0000000000000514
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose:Mechanical loosening remains a common complication associated with mini-implant failure. The purpose of this study was to compare common mechanical measures of mini-implant stability to determine their association and reliability.Materials and Methods:Ninety self-drilling orthodontic mini-implants from 6 manufacturers were inserted into artificial bone blocks. Insertion torques (ITs) and Periotest values (PVs) were measured. Subsequently, mini-implants underwent pull-out testing for measures of pull-out load (POL) and screw displacement (ScrD). Stability measurements were compared using one-way ANOVA, associations among them were assessed using correlation analyses, and reliability was evaluated using coefficients of variation (COVs).Results:Variations in stability of mini-implants were found, specific to the mechanical measure used for assessment (P < 0.05). The strongest correlations were found between IT and PV (r = -0.68) and between IT and POL (r = 0.66). Overall, PV showed the greatest variability (COV: 11%-100%) compared with IT (11%), POL (4%), and ScrD (19%).Conclusions:IT, PV, and POLs only agreed moderately in their assessment of mini-implant stability, and Periotest showed the least reliability in predicting mini-implant stability. As such, independent and interchangeable use of these stability measures should be avoided.
引用
收藏
页码:225 / 231
页数:7
相关论文
共 50 条
  • [1] Mechanical influence of thread pitch on orthodontic mini-implant stability
    da Cunha, Amanda Carneiro
    Ayres de Freitas, Amanda Osorio
    Marquezan, Mariana
    Nojima, Lincoln Issamu
    BRAZILIAN ORAL RESEARCH, 2015, 29 (01): : 1 - 6
  • [2] How Does Orthodontic Mini-Implant Thread Minidesign Influence the Stability?-Systematic Review with Meta-Analysis
    Jedlinski, Maciej
    Janiszewska-Olszowska, Joanna
    Mazur, Marta
    Grocholewicz, Katarzyna
    Suquia, Pedro Suarez
    Quintanilla, David Suarez
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (18)
  • [3] Effects of gripping volume in the mechanical strengths of orthodontic mini-implant
    Tseng, Yu-Chuan
    Wu, Ju-Hui
    Chen, Hong-Sen
    Chen, Chun-Ming
    Ting, Chun-Chan
    KAOHSIUNG JOURNAL OF MEDICAL SCIENCES, 2017, 33 (11) : 578 - 583
  • [4] Mechanical stability assessment of novel orthodontic mini-implant designs: Part 2
    Hong, Christine
    Peter Truong
    Ha Na Song
    Wu, Benjamin M.
    Moon, Won
    ANGLE ORTHODONTIST, 2011, 81 (06) : 1001 - 1009
  • [5] Influence of bone density, screw size and surgical procedure on orthodontic mini-implant placement - part B: implant stability
    Moehlhenrich, S. C.
    Heussen, N.
    Modabber, A.
    Bock, A.
    Hoelzle, F.
    Wilmes, B.
    Danesh, G.
    Szalma, J.
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2021, 50 (04) : 565 - 572
  • [6] Orthodontic mini-implant stability and the ratio of pilot hole implant diameter
    Uemura, Miwa
    Motoyoshi, Mitsuru
    Yano, Shinya
    Sakaguchi, Masahito
    Igarashi, Yu
    Shimizu, Noriyoshi
    EUROPEAN JOURNAL OF ORTHODONTICS, 2012, 34 (01) : 52 - 56
  • [7] Orthodontic mini-implant stability at different insertion depths Sensitivity of three stability measurement methods
    Nienkemper, Manuel
    Santel, Natascha
    Hoenscheid, Ralf
    Drescher, Dieter
    JOURNAL OF OROFACIAL ORTHOPEDICS-FORTSCHRITTE DER KIEFERORTHOPADIE, 2016, 77 (04): : 296 - 303
  • [8] Experimental & FEM Analysis of Orthodontic Mini-Implant Design on Primary Stability
    Mesic, Elmedin
    Muratovic, Enis
    Redzepagic-Vrazalica, Lejla
    Pervan, Nedim
    Muminovic, Adis J.
    Delic, Muamer
    Glusac, Mirza
    APPLIED SCIENCES-BASEL, 2021, 11 (12):
  • [9] Finite element simulation and statistical investigation of an orthodontic mini-implant's stability in a novel screw design
    Pouyafar, Vahid
    Meshkabadi, Ramin
    Sadr Haghighi, Amir Hooman
    Navid, Ali
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2021, 235 (09) : 1046 - 1057
  • [10] The effect of different orthodontic mini-implant brands and geometry on primary stability (an in vitro study)
    Abdalla, Khanda Latif
    Mahmood, Trefa Mohammed Ali
    HELIYON, 2023, 9 (09)