Single input state polarization sensitive swept source optical coherence tomography based on an all single mode fiber interferometer

被引:38
|
作者
Trasischker, Wolfgang [1 ]
Zotter, Stefan [1 ]
Torzicky, Teresa [1 ]
Baumann, Bernhard [1 ]
Haindl, Richard [1 ]
Pircher, Michael [1 ]
Hitzenberger, Christoph K. [1 ]
机构
[1] Med Univ Vienna, Ctr Med Phys & Biomed Engn, A-1090 Vienna, Austria
来源
BIOMEDICAL OPTICS EXPRESS | 2014年 / 5卷 / 08期
关键词
FLUID-FLOW VELOCITY; IN-VIVO; RESOLVED BIREFRINGENCE; LAYER; REFLECTOMETER; PERFORMANCE; TISSUE; OCT;
D O I
10.1364/BOE.5.002798
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We present a newly developed single mode fiber based swept source polarization sensitive optical coherence tomography system using a single input state at 1040 nm. Two non-polarizing fiber based beam splitters are combined to form a Mach-Zehnder interferometer, while two polarizing beam splitters are used to obtain a polarization sensitive detection. Both types of beam splitters solely feature conventional single mode fibers. Polarization control paddles are used to set and maintain the polarization states in the fibers of the interferometer and detection unit. By use of a special paddle alignment scheme we are able to eliminate any bulk optic wave plates and polarization maintaining fibers in the interferometer and detection paths while preserving the advantages of a single input state system that illuminates the sample with circularly polarized light. To demonstrate the capabilities of our system, we performed retinal measurements on healthy human volunteers. (C)2014 Optical Society of America
引用
收藏
页码:2798 / 2809
页数:12
相关论文
共 50 条
  • [21] Single camera based spectral domain polarization sensitive optical coherence tomography
    Baumann, Bernhard
    Goetzinger, Erich
    Pircher, Michael
    Hitzenberger, Christoph K.
    OPTICS EXPRESS, 2007, 15 (03): : 1054 - 1063
  • [22] Swept source/Fourier domain polarization sensitive optical coherence tomography with a passive polarization delay unit
    Baumann, Bernhard
    Choi, WooJhon
    Potsaid, Benjamin
    Huang, David
    Duker, Jay S.
    Fujimoto, James G.
    OPTICS EXPRESS, 2012, 20 (09):
  • [23] Digital calibration method to enable depth-resolved all-fiber polarization sensitive optical coherence tomography with an arbitrary input polarization state
    Tang, Peijun
    Wang, Ruikang K.
    Chao, Qing
    BIOMEDICAL OPTICS EXPRESS, 2024, 15 (05): : 3329 - 3343
  • [24] Investigation of Anterior Chamber Angle by Swept-Source Polarization Sensitive Optical Coherence Tomography
    Yasuno, Yoshiaki
    Yamanari, Masahiro
    Kawana, Keisuke
    Miura, Masahiro
    Makita, Shuichi
    Oshika, Tetsuro
    OPHTHALMIC TECHNOLOGIES XIX, 2009, 7163
  • [25] Polarization sensitive optical coherence tomography with single input for imaging depth-resolved collagen organizations
    Tang, Peijun
    Kirby, Mitchell A.
    Le, Nhan
    Li, Yuandong
    Zeinstra, Nicole
    Lu, G. Nina
    Murry, Charles E.
    Zheng, Ying
    Wang, Ruikang K.
    LIGHT-SCIENCE & APPLICATIONS, 2021, 10 (01)
  • [26] Polarization sensitive optical coherence tomography with single input for imaging depth-resolved collagen organizations
    Peijun Tang
    Mitchell A. Kirby
    Nhan Le
    Yuandong Li
    Nicole Zeinstra
    G. Nina Lu
    Charles E. Murry
    Ying Zheng
    Ruikang K. Wang
    Light: Science & Applications, 10
  • [27] MEASUREMENT OF BIREFRINGENCE OF THE RETINAL NERVE FIBER LAYER USING SWEPT SOURCE POLARIZATION SENSITIVE OPTICAL COHERENCE TOMOGRAPHY (1 UM)
    Elmaanaoui, Badr
    Dwelle, Jordan
    McElroy, Austin
    Paranjape, Amit
    Liu, Shuang
    Rylander, Henry
    Milner, Thomas
    LASERS IN SURGERY AND MEDICINE, 2010, : 16 - 16
  • [28] Fiber-based polarization-sensitive optical coherence tomography
    Saxer, CE
    de Boer, JF
    Park, BH
    Zhao, YH
    Chen, YP
    Nelson, JS
    COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICAL SCIENCE AND CLINICAL APPLICATIONS IV, 2000, 3915 : 68 - 75
  • [29] Optic axis determination by fibre-based polarization-sensitive swept-source optical coherence tomography
    Lu, Zenghai
    Kasaragod, Deepa K.
    Matcher, Stephen J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (04): : 1105 - 1122
  • [30] Single shot wide field MHz swept source optical coherence tomography angiography
    Niederleithner, Michael
    Britten, Anja
    Matten, Philipp
    Ren, Hugang
    Williams, Rick
    Kubach, Sophie
    Leitgeb, Rainer A.
    Drexler, Wolfgang
    Schmoll, Tilman
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2021, 62 (08)