Fast LC separation of a myoglobin digest: a case study using monolithic and particulate RP 18 silica capillary columns

被引:8
作者
Rozenbrand, Johan
van Bennekom, Wouter P.
Unger, Klaus K.
de Jong, Gerhardus J.
机构
[1] Univ Utrecht, Fac Sci, Dept Biomed Anal, NL-3584 CA Utrecht, Netherlands
[2] Johannes Gutenberg Univ Mainz, Inst Anorgan Chem & Analyt Chem, D-55099 Mainz, Germany
关键词
proteomics; myoglobin digest; peptides; monolithic capillary silica columns;
D O I
10.1007/s00216-006-0520-4
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A method was developed for the fast separation of a myoglobin digest using a monolithic RP 18 silica capillary column of 100 mu m I.D. The results were compared with those obtained with a particulate RP 18 silica capillary column of 100 mu m I.D. at a flow-rate between 0.6 and 1.2 mu l/min. The digest was analyzed at the monolithic column at a flow-rate up to 2.8 mu l/min. This high flow-rate could not be applied to the particulate column due to the high back-pressure. When the starting composition of the gradient was changed from 0 to 20% and a gradient steepness of 16%/min was used, the analysis time was less than 4 min. A positive Mascot identification score of 115 was achieved for the MS-MS data. When a lower gradient steepness was employed, the chromatographic resolution and the peak capacity did not increase for most compounds. The intraday repeatability for the retention time of the monolithic column was better than 1.5% at 2.8 mu l/min and even less than 0.5% using a flow-rate of 0.6 or 1.0 mu l/min. For the particulate column, it was between 0.5 and 1.4% for a flow-rate of 0.6 mu l/min, probably due to the high column back-pressure. The interday reproducibility for the retention time of the monolithic column was less than 0.9% using a flow-rate of 1.0 mu l/min.
引用
收藏
页码:1055 / 1061
页数:7
相关论文
共 24 条
[1]   Liquid chromatographic high-throughput analysis of ketamine and its metabolites in human plasma using a monolithic silica column and solid phase extraction [J].
Aboul-Enein, HY ;
Hefnawy, MM .
TALANTA, 2005, 65 (01) :67-73
[3]   Applications of monolithic silica capillary columns in proteomics [J].
Barroso, B ;
Lubda, D ;
Bischoff, R .
JOURNAL OF PROTEOME RESEARCH, 2003, 2 (06) :633-642
[4]   High-throughput liquid chromatography-tandem mass spectrometry determination of bupropion and its metabolites in human, mouse and rat plasma using a monolithic column [J].
Borges, V ;
Yang, E ;
Dunn, J ;
Henion, J .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 804 (02) :277-287
[5]   Rapid analysis of benzodiazepines in whole blood by high-performance liquid chromatography: use of a monolithic column [J].
Bugey, A ;
Staub, C .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2004, 35 (03) :555-562
[6]   Rapid protein identification using a microscale electrospray LC/MS system on an ion trap mass spectrometer [J].
Davis, MT ;
Lee, TD .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1998, 9 (03) :194-201
[7]   Practical aspects of fast reversed-phase high-performance liquid chromatography using 3 μm particle packed columns and monolithic columns in pharmaceutical development and production working under current good manufacturing practice [J].
Gerber, F ;
Krummen, M ;
Potgeter, H ;
Roth, A ;
Siffrin, C ;
Spoendlin, C .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1036 (02) :127-133
[8]   A robust micro-electrospray ionization technique for high-throughput liquid chromatography/mass spectrometry proteomics using a sanded metal needle as an emitter [J].
Guzzetta, AW ;
Thakur, RA ;
Mylchreest, IC .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2002, 16 (21) :2067-2072
[9]   EVALUATION OF ADVANCED SILICA PACKINGS FOR THE SEPARATION OF BIO-POLYMERS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY .3. RETENTION AND SELECTIVITY OF PROTEINS AND PEPTIDES IN GRADIENT ELUTION ON NONPOROUS MONODISPERSE 1.5-MU-M REVERSED-PHASE SILICAS [J].
JILGE, G ;
JANZEN, R ;
GIESCHE, H ;
UNGER, KK ;
KINKEL, JN ;
HEARN, MTW .
JOURNAL OF CHROMATOGRAPHY, 1987, 397 :71-80
[10]  
Meiring HD, 2002, J SEP SCI, V25, P557, DOI 10.1002/1615-9314(20020601)25:9<557::AID-JSSC557>3.0.CO