A portable time-domain LED fluorimeter for nanosecond fluorescence lifetime measurements

被引:21
作者
Wang, Hongtao [1 ,2 ]
Qi, Ying [2 ,3 ]
Mountziaris, T. J. [2 ,3 ]
Salthouse, Christopher D. [1 ,2 ]
机构
[1] Univ Massachusetts, Dept Elect & Comp Engn, Amherst, MA 01003 USA
[2] Univ Massachusetts, Ctr Personalized Hlth Monitoring, Amherst, MA 01003 USA
[3] Univ Massachusetts, Dept Chem Engn, Amherst, MA 01003 USA
基金
美国国家科学基金会;
关键词
INTENSITY VARIATIONS; MODULATION;
D O I
10.1063/1.4873330
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Fluorescence lifetime measurements are becoming increasingly important in chemical and biological research. Time-domain lifetime measurements offer fluorescence multiplexing and improved handling of interferers compared with the frequency-domain technique. In this paper, an all solid-state, filterless, and highly portable light-emitting-diode based time-domain fluorimeter (LED TDF) is reported for the measurement of nanosecond fluorescence lifetimes. LED based excitation provides more wavelengths options compared to laser diode based excitation, but the excitation is less effective due to the uncollimated beam, less optical power, and longer latency in state transition. Pulse triggering and pre-bias techniques were implemented in our LED TDF to improve the peak optical power to over 100 mW. The proposed pulsing circuit achieved an excitation light fall time of less than 2 ns. Electrical resetting technique realized a time-gated photo-detector to remove the interference of the excitation light with fluorescence. These techniques allow the LED fluorimeter to accurately measure the fluorescence lifetime of fluorescein down to concentration of 0.5 mu M. In addition, all filters required in traditional instruments are eliminated for the non-attenuated excitation/emission light power. These achievements make the reported device attractive to biochemical laboratories seeking for highly portable lifetime detection devices for developing sensors based on fluorescence lifetime changes. The device was initially validated by measuring the lifetimes of three commercial fluorophores and comparing them with reported lifetime data. It was subsequently used to characterize a ZnSe quantum dot based DNA sensor. (C) 2014 Author(s).
引用
收藏
页数:12
相关论文
共 34 条
[1]  
[Anonymous], ISS TECHNICAL NOTES
[2]  
[Anonymous], P SOC PHOTO OPT INS
[3]   AN ERROR ANALYSIS OF THE RAPID LIFETIME DETERMINATION METHOD FOR THE EVALUATION OF SINGLE EXPONENTIAL DECAYS [J].
BALLEW, RM ;
DEMAS, JN .
ANALYTICAL CHEMISTRY, 1989, 61 (01) :30-33
[4]   High resolution TCSPC lifetime imaging [J].
Becker, W ;
Bergmann, A ;
Biskup, C ;
Kelbauskas, L ;
Zimmer, T ;
Klöcker, N ;
Benndorf, K .
MULTIPHOTON MICROSCOPY IN THE BIOMEDICAL SCIENCES III, 2003, 4963 :175-184
[5]   Fluorescence Lifetime Measurements and Biological Imaging [J].
Berezin, Mikhail Y. ;
Achilefu, Samuel .
CHEMICAL REVIEWS, 2010, 110 (05) :2641-2684
[6]   Time-resolved measurements of luminescence [J].
Collier, Bradley B. ;
McShane, Michael J. .
JOURNAL OF LUMINESCENCE, 2013, 144 :180-190
[7]   The Concept of λ-Ratiometry in Fluorescence Sensing and Imaging [J].
Demchenko, Alexander P. .
JOURNAL OF FLUORESCENCE, 2010, 20 (05) :1099-1128
[8]   2-D fluorescence lifetime imaging using a time-gated image intensifier [J].
Dowling, K ;
Hyde, SCW ;
Dainty, JC ;
French, PMW ;
Hares, JD .
OPTICS COMMUNICATIONS, 1997, 135 (1-3) :27-31
[9]   Flow cytometry for high-throughput, high-content screening [J].
Edwards, BS ;
Oprea, T ;
Prossnitz, ER ;
Sklar, LA .
CURRENT OPINION IN CHEMICAL BIOLOGY, 2004, 8 (04) :392-398
[10]   Synthesis and properties of biocompatible water-soluble silica-coated CdSe/ZnS semiconductor quantum dots [J].
Gerion, D ;
Pinaud, F ;
Williams, SC ;
Parak, WJ ;
Zanchet, D ;
Weiss, S ;
Alivisatos, AP .
JOURNAL OF PHYSICAL CHEMISTRY B, 2001, 105 (37) :8861-8871