摘要
Characteristic ion mobility mass spectrometry data, reduced mobility, and limits of detection (signal-to-noise ratio = 3) were determined for six synthetic drugs and cocaine by ion mobility time-of-flight mass spectrometry (IM-TOF-MS) with electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI). The studied synthetic illicit drugs recently appeared on the recreational drug market as designer drugs and were methylone, 4-MEC (4'-methylethcathinone), 3,4-MDPV (3,4-methylenedioxypyrovalerone), JWH-210 [4-ethylnaphthalen-1-yl-(1-pentylindol-3-yl)methanone], JWH-250 (2-(2-methoxyphenyl)-1-(1-pentyl-1H-indol-3-yl)ethanone], and JWH-203 [1-pentyl-3-(2'-chlorophenylacetyl) indole]. Absolute reduced mobilities in nitrogen were 1.35, 1.28, 1.41, 1.30, 1.18, 0.98, 1.09, and 1.07 cm(2)V(-1) s(-1), for methylone [M-H](+), methylone [M+H](+), 4-MEC [M-H](+), 4-MEC [M+H](+), 3,4-MDPV [M+H](+), JWH-210 [M+H](+), JWH-250 [M+H](+), and JWH-203 [M+H](+), respectively. Selected illicit drugs are easily identified by IM-TOF-MS during a 100s analysis. Relative limits of detection ranged from 4 to 400 nM are demonstrated for these compounds. Such relative Limits of detection correspond to 14 pg to 2 ng absolute limits of detection. Better detection limits are obtained in APCI mode for all the illicit drugs except cocaine. ESI mode was found to be preferable for the IM-TOE-MS detection of cocaine at trace levels. A single sample analysis is completed in an order of magnitude less time than that for conventional liquid chromatography/mass spectrometry approach. The application allows one to consider IM-TOF-MS as a good candidate for a method to determine quickly the recently surfaced designer drugs marketed on the internet as "bath salts," "spice," and "herbal blends".
引用
收藏
页码:185 / 192
页数:8
相关论文
共 42 条
[1]   Letter: A simple ion source set-up for desorption/ionization on silicon with ion mobility spectrometry and ion mobility spectrometry-mass spectrometry [J].
Adamov, Alexey ;
Sysoev, Alexey A. ;
Grigoras, Kestutis ;
Laakia, Jaakko ;
Kotiaho, Tapio .
EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2011, 17 (06) :593-597
[2]   Characterization of a high resolution drift tube ion mobility spectrometer with a multi-ion source platform [J].
Adamov, Alexey ;
Mauriala, Timo ;
Teplov, Victor ;
Laakia, Jaakko ;
Pedersen, Christian Schack ;
Kotiaho, Tapio ;
Sysoev, Alexey A. .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2010, 298 (1-3) :24-29
[3]   Analytical methods to determine cocaine contamination of banknotes from around the world [J].
Armenta, S. ;
de la Guardia, M. .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2008, 27 (04) :344-351
[4]   'Spice' and other herbal blends: harmless incense or cannabinoid designer drugs? [J].
Auwaerter, Volker ;
Dresen, Sebastian ;
Weinmann, Wolfgang ;
Mueller, Michael ;
Puetz, Michael ;
Ferreiros, Nerea .
JOURNAL OF MASS SPECTROMETRY, 2009, 44 (05) :832-837
[5]   Analytical Data for Identification of the Cannabimimetic Phenylacetylindole JWH-203 [J].
Bononi, Monica ;
Belgi, Paolo ;
Tateo, Fernando .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2011, 35 (06) :360-363
[6]   Methylone and mCPP, two new drugs of abuse? [J].
Bossong, MG ;
Van Dijk, JP ;
Niesink, RJM .
ADDICTION BIOLOGY, 2005, 10 (04) :321-323
[7]   Comparison of Cocaine Detections in Corona Discharge Ionization-Ion Mobility Spectrometry and in Atmospheric Pressure Chemical Ionization-Mass Spectrometry [J].
Choi, Sung-Seen ;
Kim, Yun-Ki ;
Kim, Ok-Bae ;
An, Seung Geon ;
Shin, Myung-Won ;
Maeng, Seug-Jin ;
Choi, Gyu Seop .
BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2010, 31 (08) :2383-2385
[8]   The characterization of some 3,4-methylenedioxycathinone (MDCATH) homologs [J].
DalCason, TA .
FORENSIC SCIENCE INTERNATIONAL, 1997, 87 (01) :9-53
[9]  
DeTulleo A., 2000, Int. J. Ion Mobility Spectrom, V3, P38
[10]   Chemical standards in ion mobility spectrometry [J].
Eiceman, GA ;
Nazarov, EG ;
Stone, JA .
ANALYTICA CHIMICA ACTA, 2003, 493 (02) :185-194