Effects of Ethylene Oxide Resterilization and In-vitro Degradation on Mechanical Properties of Partially Absorbable Composite Hernia Meshes

被引:0
作者
Endogan, T. [1 ,2 ]
Ozyaylali, I. [3 ]
Kulacoglu, H. [4 ]
Serbetci, K. [5 ]
Kiyak, G. [6 ]
Hasirci, N. [1 ,7 ,8 ,9 ]
机构
[1] Middle E Tech Univ, Grad Dept Polymer Sci & Technol, TR-06531 Ankara, Turkey
[2] Middle E Tech Univ, Cent Lab, TR-06531 Ankara, Turkey
[3] Ankara Hernia Ctr, Ankara, Turkey
[4] Diskapi Yildirim Beyazit Teaching & Res Hosp, Dept Surg, Ankara, Turkey
[5] Baskent Univ, Dept Biomed Engn, Ankara, Turkey
[6] Ankara Ataturk Teaching & Res Hosp, Dept Surg, Ankara, Turkey
[7] Middle E Tech Univ, Dept Chem, TR-06531 Ankara, Turkey
[8] Middle E Tech Univ, Grad Dept Biotechnol, TR-06531 Ankara, Turkey
[9] Middle E Tech Univ, TR-06531 Ankara, Turkey
关键词
mesh; hernia; composite mesh; resterilization; degradation; mechanical properties; tensiometry; tensile strength; polypropylene; polyglecaprone; polyglactine; QUALITY-OF-LIFE; POLYPROPYLENE MESHES; REPAIR; STERILIZATION; IMPACT; MODEL;
D O I
暂无
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background and Aim: Prosthetic mesh repair for abdominal wall hernias is widely used because of its technical simplicity and low hernia recurrence rates. The most commonly used material is pure polypropylene mesh, although newer composite materials are recommended by some centers due to their advantages. However, these meshes are more expensive than pure polypropylene meshes. Resterilization of a pure polypropylene mesh has been shown to be quite safe, and many centers prefer slicing a large mesh into smaller pieces, suitable for any hernia type or defect size. Nevertheless there is no data about the safety after resterilization of the composite meshes. The present study was carried out to investigate the effects of resterilization and in vitro degradation in phosphate buffered saline solution on the physical structure and the mechanical properties of partially absorbable lightweight meshes. Methods: Two composite meshes were used in the study: One mesh consists of monofilament polypropylene and monofilament polyglecaprone -a copolymer of glycolide and epsilon (epsilon) - caprolactone -(Ultrapro (R), 28 g/m(2), Ethicon, Hamburg, Germany), and the other one consisted of multifilament polypropylene and multifilament polyglactine (Vypro II (R), 30 g/m(2), Ethicon, Hamburg, Germany). Two large meshes were cut into rectangular specimens sized 50 x 20 mm for mechanical testing and 20 x 20 mm for in vitro degradation experiments. Meshes were divided into control group with no resterilization and gas resterilization. Ethylene oxide gas sterilization was performed at 55 degrees C for 4.5 hours. In vitro degradation in 0.01 M phosphate buffered saline (PBS, pH 7.4) solution at 37 +/- 1 degrees C for 8 weeks was applied to one subgroup in each mesh group. Tensiometric measurements and scanning electron microscopic evaluations were completed for control and resterilization specimens. Results: Regardless of resterilization, when the meshes were exposed to in vitro degradation, all mechanical parameters decreased significantly. Highest reduction in mechanical properties was observed for Ultrapro due to the degradation of absorbable polyglecaprone and polyglactin parts of these meshes. It was observed that resterilization by ethylene oxide did not determine significant difference on the degradation characteristics and almost similar physical structures were observed for resterilized and non-resterilized meshes. For Vypro II meshes, no significant mechanical difference was observed between resterilized and non-resterilized meshes after degradation while resterilized Ultrapro meshes exhibited stronger characteristics than non-resterilized counterparts, after degradation. Conclusion: Resterilization with ethylene oxide did not affect the mechanical properties of partially absorbable composite meshes. No important surface changes were observed in scanning electron microscopy after resterilization.
引用
收藏
页码:840 / 848
页数:9
相关论文
共 32 条
[1]   Sterilization, toxicity, biocompatibility and clinical applications of polylactic acid polyglycolic acid copolymers [J].
Athanasiou, KA ;
Niederauer, GG ;
Agrawal, CM .
BIOMATERIALS, 1996, 17 (02) :93-102
[2]   Time-Dependent Changes in Biomechanical Properties of Four Different Synthetic Materials in a Rabbit Model and the Importance in Respect to Sling Surgery [J].
Atmaca, Ali Fuat ;
Serefoglu, Ege Can ;
Eroglu, Muzaffer ;
Guerdal, Mesut ;
Metin, Ahmet ;
Kayigil, Oender .
UROLOGIA INTERNATIONALIS, 2008, 81 (04) :456-461
[3]   Partially absorbable meshes for hernia. repair offer advantages over nonabsorbable meshes [J].
Bellon, Juan M. ;
Rodriguez, Marta ;
Garcia-Honduvilla, Natalio ;
Pascual, Gernma ;
Bujan, Julia .
AMERICAN JOURNAL OF SURGERY, 2007, 194 (01) :68-74
[4]   Influence of resterilized polypropylen meshes on growth of human fibroblasts -: an experimental in vitro study [J].
Broll, R ;
Bethge, T ;
Windhövel, U ;
Schwandner, O ;
Markert, U ;
Bruch, HP ;
Duchrow, M .
ZENTRALBLATT FUR CHIRURGIE, 2002, 127 (07) :589-593
[5]  
Buben I, 1999, Cent Eur J Public Health, V7, P197
[6]   Use of resterilized polypropylene mesh in inguinal hernia repair:: A prospective, randomized study [J].
Cingi, A ;
Manukyan, MN ;
Güllüoglu, BM ;
Barlas, A ;
Yegen, C ;
Yalin, R ;
Aktan, AÖ .
JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 2005, 201 (06) :834-840
[7]  
Coda A, 2003, Hernia, V7, P29
[8]  
Dahlke H, 1979, Handchirurgie, V11, P3
[9]   Alterations in damage processes in dense cancellous bone following gamma-radiation sterilization [J].
Dux, S. J. ;
Ramsey, D. ;
Chu, E. H. ;
Rimnac, C. M. ;
Hernandez, C. J. .
JOURNAL OF BIOMECHANICS, 2010, 43 (08) :1509-1513
[10]   Vypro II Mesh for Inguinal Hernia Repair A Meta Analysis of Randomized Controlled Trials [J].
Gao, Mingtai ;
Han, Jianxu ;
Tian, Jinhui ;
Yang, Kehu .
ANNALS OF SURGERY, 2010, 251 (05) :838-842