Modeling Elekta VersaHD using the Varian Eclipse treatment planning system for photon beams: A single-institution experience

被引:13
作者
Zhang, You [1 ]
Le, Anh H. [2 ]
Igbal, Zohaib [1 ]
Chiu, Tsuicheng [1 ]
Gu, Xuejun [1 ]
Pugachev, Andrei [4 ]
Reynolds, Robert [1 ]
Park, Yang K. [1 ]
Lin, Mu-Han [1 ]
Stojadinovic, Strahinja [1 ]
Tian, Zhen [3 ]
机构
[1] UT Southwestern Med Ctr, Dallas, TX 75390 USA
[2] Roswell Pk Canc Inst, Buffalo, NY 14263 USA
[3] Emory Univ, Winship Canc Inst, Atlanta, GA 30322 USA
[4] Oregon Hlth & Sci Univ, Portland, OR 97201 USA
关键词
commissioning; Eclipse; Elekta; treatment planning system; varian; VersaHD; DOSIMETRIC CHARACTERISTICS; IMRT; ACCELERATOR;
D O I
10.1002/acm2.12709
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to report a single-institution experience and commissioning data for Elekta VersaHD linear accelerators (LINACs) for photon beams in the Eclipse treatment planning system (TPS). Two VersaHD LINACs equipped with 160-leaf collimators were commissioned. For each energy, the percent-depth-dose (PDD) curves, beam profiles, output factors, leaf transmission factors and dosimetric leaf gaps (DLGs) were acquired in accordance with the AAPM task group reports No. 45 and No. 106 and the vendor-supplied documents. The measured data were imported into Eclipse TPS to build a VersaHD beam model. The model was validated by creating treatment plans spanning over the full-spectrum of treatment sites and techniques used in our clinic. The quality assurance measurements were performed using MatriXX, ionization chamber, and radiochromic film. The DLG values were iteratively adjusted to optimize the agreement between planned and measured doses. Mobius, an independent LINAC logfile-based quality assurance tool, was also commissioned both for routine intensity-modulated radiation therapy (IMRT) QA and as a secondary check for the Eclipse VersaHD model. The Eclipse-generated VersaHD model was in excellent agreement with the measured PDD curves and beam profiles. The measured leaf transmission factors were less than 0.5% for all energies. The model validation study yielded absolute point dose agreement between ionization chamber measurements and Eclipse within +/- 4% for all cases. The comparison between Mobius and Eclipse, and between Mobius and ionization chamber measurements lead to absolute point dose agreement within +/- 5%. The corresponding 3D dose distributions evaluated with 3%global/2mm gamma criteria resulted in larger than 90% passing rates for all plans. The Eclipse TPS can model VersaHD LINACs with clinically acceptable accuracy. The model validation study and comparisons with Mobius demonstrated that the modeling of VersaHD in Eclipse necessitates further improvement to provide dosimetric accuracy on par with Varian LINACs.
引用
收藏
页码:33 / 42
页数:10
相关论文
共 29 条
[1]  
Ajo R., 2016, DOSIMETRIC EVALUATIO
[2]   Intra- and intervariability in beam data commissioning among water phantom scanning systems [J].
Akino, Yuichi ;
Gibbons, John P. ;
Neck, Daniel W. ;
Chu, Connel ;
Das, Indra J. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (04) :251-258
[3]   Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator [J].
Bedford, James L. ;
Thomas, Michael D. R. ;
Smyth, Gregory .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02) :172-185
[4]   RayStation: External beam treatment planning system [J].
Bodensteiner, Dayna .
MEDICAL DOSIMETRY, 2018, 43 (02) :168-176
[5]  
BOYER A, 2001, MED PHYS, V72, P16
[6]   Dosimetric characteristics of Novalis Tx system with high definition multileaf collimator [J].
Chang, Zheng ;
Wang, Zhiheng ;
Wu, Q. Jackie ;
Yan, Hui ;
Bowsher, Jim ;
Zhang, Junan ;
Yin, Fang-Fang .
MEDICAL PHYSICS, 2008, 35 (10) :4460-4463
[7]   Commissioning and dosimetric characteristics of TrueBeam system: Composite data of three TrueBeam machines [J].
Chang, Zheng ;
Wu, Qiuwen ;
Adamson, Justus ;
Ren, Lei ;
Bowsher, James ;
Yan, Hui ;
Thomas, Andrew ;
Yin, Fang-Fang .
MEDICAL PHYSICS, 2012, 39 (11) :6981-7018
[8]   Parallel/Opposed: IMRT QA using treatment log files is superior to conventional measurement-based method [J].
Childress, Nathan ;
Chen, Quan ;
Rong, Yi .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (01) :4-7
[9]   Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM [J].
Das, Indra J. ;
Cheng, Chee-Wai ;
Watts, Ronald J. ;
Ahnesjo, Anders ;
Gibbons, John ;
Li, X. Allen ;
Lowenstein, Jessica ;
Mitra, Raj K. ;
Simon, William E. ;
Zhu, Timothy C. .
MEDICAL PHYSICS, 2008, 35 (09) :4186-4215
[10]   An experimental evaluation of the Agility MLC for motion-compensated VMAT delivery [J].
Davies, G. A. ;
Clowes, P. ;
Bedford, J. L. ;
Evans, P. M. ;
Webb, S. ;
Poludniowski, G. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (13) :4643-4657