DC voltage calibrator based on an array of high-temperature superconductor Josephson junctions

被引:0
作者
Klushin, AM [1 ]
Borovitskii, SI [1 ]
Gelikonova, VD [1 ]
Komkov, AV [1 ]
Siegel, M [1 ]
机构
[1] Forschungszentrum Julich GmbH, Inst Schichten & Grenzflachen, D-52428 Julich, Germany
来源
2002 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS, CONFERENCE DIGEST | 2002年
关键词
D O I
10.1109/CPEM.2002.1034887
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We are using a quantum dc voltage generated on an array of high-temperature superconductor (HTS) junctions in the developed voltage calibrator with an output voltage up to 10 V. The described concept and the demonstrated parameters of the main parts of the instrument show, that the measurement uncertainty of the output voltage will be few parts in 10(8).