Method for the Estimation of Effective Doses for Dental Intraoral and Panoramic Radiography

被引:0
作者
Shatskiy, I. G. [1 ]
机构
[1] St Petersburg Res Inst Radiat Hyg, St Petersburg, Russia
来源
6TH INTERNATIONAL CONFERENCE ON X-RAY, ELECTROVACUUM AND BIOMEDICAL TECHNIQUE | 2020年 / 2250卷
关键词
REDUCTION;
D O I
10.1063/5.0013321
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The doses of patients for the most common intraoral and panoramic radiographic dental studies were evaluated and compared with the published data. The range of median values of effective doses for intraoral examinations on the units with films detectors was from 3.5 to 8.2 mu Sv and for the units with digital detectors the corresponding range was from 1.2 to 2.5 mu Sv. For panoramic examinations median value of the effective dose was 22.9 mu Sv. The ranges without emissions for the examination of premolars of the upper jaw, molars of both jaws and occlusion have the substantial width. Results compared with published data indicated that the effective doses in St. Petersburg and the regional institutions are lower than the values obtained using standard round collimator and medium and high sensitivity films (class D and F, respectively), as well as lower dose levels in the European Commission Guidelines. However, the effective doses in St. Petersburg and the Leningrad Region are higher than those obtained using a square collimator and high-sensitivity films. The present study suggests the possibility of further optimization. Obtained dose values can be used as reference for local regulations concerning patient dose in dental radiography and preliminary value prior to the establishment of national diagnostic reference level
引用
收藏
页数:6
相关论文
共 8 条
  • [1] ABSORBED DOSES FOR PATIENTS UNDERGOING PANORAMIC RADIOGRAPHY, CEPHALOMETRIC RADIOGRAPHY AND CBCT
    Wrzesien, Malgorzata
    Olszewski, Jerzy
    INTERNATIONAL JOURNAL OF OCCUPATIONAL MEDICINE AND ENVIRONMENTAL HEALTH, 2017, 30 (05) : 705 - 713
  • [2] EFFECTIVE DOSES AND RADIATION RISKS FROM COMMON DENTAL RADIOGRAPHIC, PANORAMIC AND CBCT EXAMINATIONS
    Shatskiy, I
    RADIATION PROTECTION DOSIMETRY, 2021, 195 (3-4) : 296 - 305
  • [3] Influence of lead apron shielding on absorbed doses from panoramic radiography
    Rottke, D.
    Grossekettler, L.
    Sawada, K.
    Poxleitner, P.
    Schulze, D.
    DENTOMAXILLOFACIAL RADIOLOGY, 2013, 42 (10)
  • [4] Absorbed organ and effective doses from digital intra-oral and panoramic radiography applying the ICRP 103 recommendations for effective dose estimations
    Granlund, Christina
    Thilander-Klang, Anne
    Ylhan, Betul
    Lofthag-Hansen, Sara
    Ekestubbe, Annika
    BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1066)
  • [5] Reducing Pediatric Intraoral Radiography Radiation Dose Using Reduced-Power Dental X-Ray Units: A Randomized Trial
    Cully, Jennifer L.
    Somasundaram, Elanchezhian
    Campbell, Richard
    Brady, Samuel L.
    Gosnell, Elizabeth S.
    Specht, Sarah
    Candon, Lisa Atlas
    Strauss, Keith J.
    JOURNAL OF DENTISTRY FOR CHILDREN, 2022, 89 (02) : 95 - 103
  • [6] ASSESSMENT OF ORGAN ABSORBED DOSES AND ESTIMATION OF EFFECTIVE DOSES FROM PEDIATRIC ANTHROPOMORPHIC PHANTOM MEASUREMENTS FOR MULTI-DETECTOR ROW CT WITH AND WITHOUT AUTOMATIC EXPOSURE CONTROL
    Brisse, Herve J.
    Robilliard, Magalie
    Savignoni, Alexia
    Pierrat, Noelle
    Gaboriaud, Genevieve
    De Rycke, Yann
    Neuenschwander, Sylvia
    Aubert, Bernard
    Rosenwald, Jean-Claude
    HEALTH PHYSICS, 2009, 97 (04): : 303 - 314
  • [7] The effect of angular and longitudinal tube current modulations on the estimation of organ and effective doses in x-ray computed tomography
    van Straten, Marcel
    Deak, Paul
    Shrimpton, Paul C.
    Kalender, Willi A.
    MEDICAL PHYSICS, 2009, 36 (11) : 4881 - 4889
  • [8] A Lightweight, Effective Compressibility Estimation Method for Error-bounded Lossy Compression
    Ganguli, Arkaprabha
    Underwood, Robert
    Bessac, Julie
    Krasowska, David
    Calhoun, Jon C.
    Di, Sheng
    Cappello, Franck
    2023 IEEE INTERNATIONAL CONFERENCE ON CLUSTER COMPUTING, CLUSTER, 2023, : 247 - 258