Calibration transfer between electronic nose systems for rapid In situ measurement of pulp and paper industry emissions

被引:37
作者
Deshmukh, Sharvari [1 ,3 ]
Kamde, Kalyani [1 ]
Jana, Arun [2 ]
Korde, Sanjivani [1 ]
Bandyopadhyay, Rajib [3 ]
Sankar, Ravi [2 ]
Bhattacharyya, Nabarun [2 ]
Pandey, R. A. [1 ]
机构
[1] CSIR, Natl Environm Engn & Res Inst, Nagpur, Maharashtra, India
[2] Ctr Dev Adv Comp, Kolkata, India
[3] Jadavpur Univ, Dept Instrumentat & Elect Engn, Kolkata, India
关键词
Electronic nose; Calibration transfer; Robust regression; Box-Behnken design; Artificial neural network; Reduced sulphur compounds; SENSOR ARRAY; BOX-BEHNKEN;
D O I
10.1016/j.aca.2014.05.054
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Electronic nose systems when deployed in network mesh can effectively provide a low budget and onsite solution for the industrial obnoxious gaseous measurement. For accurate and identical prediction capability by all the electronic nose systems, a reliable calibration transfer model needs to be implemented in order to overcome the inherent sensor array variability. In this work, robust regression (RR) is used for calibration transfer between two electronic nose systems using a Box-Behnken (BB) design. Out of the two electronic nose systems, one was trained using industrial gas samples by four artificial neural network models, for the measurement of obnoxious odours emitted from pulp and paper industries. The emissions constitute mainly of hydrogen sulphide (H2S), methyl mercaptan (MM), dimethyl sulphide (DMS) and dimethyl disulphide (DMDS) in different proportions. A Box-Behnken design consisting of 27 experiment sets based on synthetic gas combinations of H2S, MM, DMS and DMDS, were conducted for calibration transfer between two identical electronic nose systems. Identical sensors on both the systems were mapped and the prediction models developed using ANN were then transferred to the second system using BB-RR methodology. The results showed successful transmission of prediction models developed for one system to other system, with the mean absolute error between the actual and predicted concentration of analytes in mg L-1 after calibration transfer (on second system) being 0.076, 0.1801, 0.0329, 0.427 for DMS, DMDS, MM, H2S respectively. (C) 2014 Elsevier B. V. All rights reserved.
引用
收藏
页码:58 / 67
页数:10
相关论文
共 20 条
  • [1] [Anonymous], SENS ACTUATORS B, DOI DOI 10.1016/j.snb.2009.08.041
  • [2] Application of Box-Behnken design and response surface methodology for modeling of some Turkish coals
    Aslan, N.
    Cebeci, Y.
    [J]. FUEL, 2007, 86 (1-2) : 90 - 97
  • [3] Bhattacharyya N., 2008, IEEE T INSTRUM MEAS, P57
  • [4] Calibration of microhotplate conductometric gas sensors by non-linear multivariate regression methods
    Dable, BK
    Booksh, KS
    Cavicchi, R
    Semancik, S
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2004, 101 (03) : 284 - 294
  • [5] Simultaneous estimation of odor classes and concentrations using an electronic nose with function approximation model ensembles
    Daqi, Gao
    Wei, Chen
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2007, 120 (02) : 584 - 594
  • [6] On field calibration of an electronic nose for benzene estimation in an urban pollution monitoring scenario
    De Vito, S.
    Massera, E.
    Piga, A.
    Martinotto, L.
    Di Francia, G.
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2008, 129 (02) : 750 - 757
  • [7] Quantitative determination of pulp and paper industry emissions and associated odor intensity in methyl mercaptan equivalent using electronic nose
    Deshmukh, Sharvari
    Jana, Arun
    Bhattacharyya, Nabarun
    Bandyopadhyay, Rajib
    Pandey, R. A.
    [J]. ATMOSPHERIC ENVIRONMENT, 2014, 82 : 401 - 409
  • [8] Heiberger R.M., 1992, J. Comput. Graph. Stat., V1, P181, DOI 10.2307/1390715.
  • [9] The emission characteristics and the related malodor intensities of gaseous reduced sulfur compounds (RSC) in a large industrial complex
    Kim, Ki-Hyun
    Jeon, Eui-Chan
    Choi, Ye-Jin
    Koo, Youn-Seo
    [J]. ATMOSPHERIC ENVIRONMENT, 2006, 40 (24) : 4478 - 4490
  • [10] Sharpening Wald-type inference in robust regression for small samples
    Koller, Manuel
    Stahel, Werner A.
    [J]. COMPUTATIONAL STATISTICS & DATA ANALYSIS, 2011, 55 (08) : 2504 - 2515