Detector Performance Characterization for High Sensitivity Single-Photon Imaging

被引:3
|
作者
Walker, Katherine L. [1 ]
Cherry, Simon R. [1 ]
Mitchell, Gregory S. [1 ]
机构
[1] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
关键词
Gamma-ray detectors; medical imaging; nuclear imaging; spect; SMALL ANIMAL SPECT; GAMMA-CAMERA; MANGANESE; CADMIUM; READOUT; MATRIX;
D O I
10.1109/TNS.2014.2299494
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We are developing detectors for an un-collimated scanner for very high sensitivity single-photon imaging. The scanner consists of two large, thin, closely spaced, pixelated scintillation detectors of either thallium-doped sodium iodide (NaI(Tl)), sodium-doped cesium iodide (CsI(Na)), or bismuth germanate (BGO). The scintillator arrays are interchangeable to optimize performance for a wide range of isotope energies and imaging subjects. For each scintillator, flood histograms were obtained utilizing test gamma sources: Cd-109, Am-241, Co-57, Ba-133, Cs-137, Na-22, and beta source Si-32. Using a 3 mm thick 21 x 41 NaI(Tl) pixelated array (2 mm x 2 mm pixels on a 2.2 mm pitch), read out by two adjacent Hamamatasu H8500 multi-channel photomultiplier tubes (MCPMTs), the energy resolution for Co-57 was 12.2% over the center of the MCPMTs and 45.3% for pixels over the gap, with flood histogram peak-to-valley ratios >6.1 for the center of the MCPMTs and >4.2 for the gap between the two MCPMTs. The NaI(Tl) detector consistently produced peak-to-valley ratios of >1.6 in the center of the MCPMTs for all radioisotopes studied. CsI(Na) and BGO detectors had flood histograms of similar quality to NaI(Tl) at intermediate to high-energies, allowing for optimal imaging across a wide energy range. The measured sensitivity for point sources placed 10 mm from the detector approached 35% for low-energy isotopes and all scintillator materials, which results in an expected system sensitivity (two opposing detectors) of similar to 70% for these isotopes. The 5 mm thick BGO arrays had the greatest efficiency for high-energy isotopes due to the increased stopping power of this material. The measured spatial resolution for Tc-99m for the CsI(Na) detector (1.6 mm pitch) was 7.5 mm FWHM at the aluminum entrance window (1.5 mm from scintillator face), decreasing to 15 mm at a distance of 5 mm from the scintillator face. The NaI(Tl) detector showed similar spatial resolution. This spatial resolution is sufficient for imaging well-localized radiotracer distributions in thin objects, and the very high sensitivity can be used for fast dynamic imaging, or imaging using very low injected doses.
引用
收藏
页码:1118 / 1125
页数:8
相关论文
共 50 条
  • [31] A single-photon detector without cooler
    Zhou, Bing
    Huang, Fu-yu
    Wu, Dong-sheng
    Liu, Jie
    ADVANCED SENSOR SYSTEMS AND APPLICATIONS V, 2012, 8561
  • [32] Thermoelectric Nanowire Single-Photon Detector
    Kuzanyan, Astghik A.
    Kuzanyan, Armen S.
    PHOTON COUNTING APPLICATIONS IV; AND QUANTUM OPTICS AND QUANTUM INFORMATION TRANSFER AND PROCESSING, 2013, 8773
  • [33] Photoconductance quantization in a single-photon detector
    Kosaka, H
    Rao, DS
    Robinson, HD
    Bandaru, P
    Sakamoto, T
    Yablonovitch, E
    PHYSICAL REVIEW B, 2002, 65 (20): : 1 - 4
  • [34] Single-photon performance
    James Baxter
    Nature Photonics, 2012, 6 (1) : 4 - 5
  • [35] Nonlocal Superconducting Single-Photon Detector
    Paolucci, Federico
    PHYSICAL REVIEW APPLIED, 2023, 20 (01)
  • [36] Ultrafast superconducting single-photon detector
    Goltsman, G.
    Korneev, A.
    Divochiy, A.
    Minaeva, O.
    Tarkhov, M.
    Kaurova, N.
    Seleznev, V.
    Voronov, B.
    Okunev, O.
    Antipov, A.
    Smirnov, K.
    Vachtomin, Yu.
    Milostnaya, I.
    Chulkova, G.
    JOURNAL OF MODERN OPTICS, 2009, 56 (15) : 1670 - 1680
  • [37] Single-photon detector design features
    Zaitsev, Sergey V.
    Kurochkin, Vladimir L.
    Kurochkin, Yury. V.
    INTERNATIONAL CONFERENCE ON MICRO- AND NANO-ELECTRONICS 2016, 2016, 10224
  • [38] Single-photon detector flashes faster
    Gaughan, R
    PHOTONICS SPECTRA, 2001, 35 (10) : 34 - 34
  • [39] Silicon-Based High-Sensitivity Near-Infrared Single-Photon dTOF Detector
    Wang Shuaikang
    Liu Danlu
    Chen Qianyu
    Han Dong
    Wang Jiayuan
    Xu Yue
    Cao Ping
    ACTA OPTICA SINICA, 2023, 43 (20)
  • [40] Photon-pair characterization using one single-photon detector by time multiplexing
    Zheng, Yulin
    He, Jiakun
    Zhou, Cheng
    Geng, Wei
    Kong, Yunchuan
    OPTICAL ENGINEERING, 2019, 58 (10)