Unintended attenuation in the Leksell Gamma Knife® Perfexion™ calibration-phantom adaptor and its effect on dose calibration

被引:16
作者
Bhatnagar, Jagdish P. [1 ]
Novotny, Josef, Jr. [1 ]
Quader, Mubina A. [1 ]
Bednarz, Greg [1 ]
Huq, M. Saiful [1 ]
机构
[1] Univ Pittsburgh, Inst Canc, Dept Radiat Oncol, Pittsburgh, PA 15217 USA
关键词
calibration; dosimeters; dosimetry; phantoms; OUTPUT FACTORS; RADIOSURGERY; BEAMS;
D O I
10.1118/1.3093240
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The calibration of Leksell Gamma Knife Perfexion (LGK PFX) is performed using a spherical polystyrene phantom 160 mm in diameter, which is provided by the manufacturer. This is the same phantom that has been used with LGK models U, B, C, and 4C. The polystyrene phantom is held in irradiation position by an aluminum adaptor, which has stainless steel side-fixation screws. The phantom adaptor partially attenuates the beams from sectors 3 and 7 by 3.2% and 4.6%, respectively. This unintended attenuation introduces a systematic error in dose calibration. The overall effect of phantom-adaptor attenuation on output calibration of the LGK PFX unit is to underestimate output by about 1.0%.
引用
收藏
页码:1208 / 1211
页数:4
相关论文
共 7 条
  • [1] Deconvolution of detector size effect for output factor measurement for narrow Gamma Knife radiosurgery beams
    Bednarz, G
    Huq, MS
    Rosenow, UF
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (20) : 3643 - 3649
  • [2] Monte Carlo calculated output factors of a Leksell Gamma Knife unit
    Cheung, JYC
    Yu, KN
    Ho, RTK
    Yu, CP
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (12) : N247 - N249
  • [3] Lindquist C, 2007, NEUROSURGERY, V61, P130, DOI 10.1227/01.NEU.0000279989.86544.B6
  • [4] Precision dosimetry for narrow photon beams used in radiosurgery -: Determination of Gamma Knife® output factors
    Mack, A
    Scheib, SG
    Major, J
    Gianolini, S
    Pazmandi, G
    Feist, H
    Czempiel, H
    Kreiner, HJ
    [J]. MEDICAL PHYSICS, 2002, 29 (09) : 2080 - 2089
  • [5] NIRANJAN A, STEREOTACT IN PRESS
  • [6] Dosimetric comparison of the Leksell Gamma Knife Perfexion and 4C
    Novotny, Josef, Jr.
    Bhatnagar, Jagdish P.
    Niranjan, Ajay
    Quader, Mubina A.
    Huq, M. Saiful
    Bednarz, Greg
    Flickinger, John C.
    Kondziolka, Douglas
    Lunsford, L. Dade
    [J]. JOURNAL OF NEUROSURGERY, 2008, 109 : 8 - 14
  • [7] Determination of the 4 mm Gamma Knife helmet relative output factor using a variety of detectors
    Tsai, JS
    Rivard, MJ
    Engler, MJ
    Mignano, JE
    Wazer, DE
    Shucart, WA
    [J]. MEDICAL PHYSICS, 2003, 30 (05) : 986 - 992