External quality control for embryology laboratories

被引:11
作者
Antonio Castilla, Jose [1 ,2 ,3 ]
Ruiz de Assin, Rafael [1 ]
Carmen Gonzalvo, Maria [1 ]
Clavero, Ana [1 ]
Pablo Ramirez, Juan [2 ,3 ]
Vergara, Francisco [2 ]
Martinez, Luis [1 ]
机构
[1] Hosp Virgen de las Nieves, Human Reprod Unit, Granada 18014, Spain
[2] Sperm Bank CEIFER, Granada, Spain
[3] Asociac Estudio Biol Reprod ASEBIR, Programa Control Calidad Externo Lab Reprod, Madrid, Spain
关键词
embryology; embryo quality; quality control; ASSISTED REPRODUCTIVE TECHNOLOGY; VITRO EMBRYOTOXICITY TESTS; INVITRO FERTILIZATION; PREDICTION MODELS; CULTURE-MEDIA; IVF; BIOASSAY; SCORE; ASSAY; EXPERIENCE;
D O I
10.1016/j.rbmo.2009.09.033
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Participation in external quality control (EQC) programmes is recommended by various scientific societies. Results from an EQC programme for embryology laboratories are presented. This 5-year programme consisted of the annual delivery of (i) materials to test toxicity and (ii) a DVD/CD-ROM with images of zygotes and embryos on days 2 and 3, on the basis of which the participants were asked to judge the embryo quality and to take a clinical decision. A high degree of agreement was considered achieved when over 75% of the laboratories produced similar classifications. With respect to the materials analysed, the specificity was 68% and the sensitivity was 83%. Concerning embryo classification, the proportion of embryos on which a high degree of agreement was achieved increased during this period from 35% to 55%. No improvement was observed in the degree of agreement on the clinical decision to be taken. Day-3 embryos produced a higher degree of agreement (58%) than did day-2 embryos (32%) (P < 0.05). Participation in EQC increased the degree of inter-laboratory agreement on embryo classification, but not the corresponding agreement on clinical decision taking. It is necessary to introduce measures aimed at standardizing decision taking procedures in embryology laboratories. (C) 2009, Reproductive Healthcare Ltd. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:68 / 74
页数:7
相关论文
共 32 条
[1]   External quality control program for semen analysis:: Spanish experience [J].
Alvarez, C ;
Castilla, JA ;
Ramírez, JP ;
Vergara, F ;
Yoldi, A ;
Fernández, A ;
Gaforio, JJ .
JOURNAL OF ASSISTED REPRODUCTION AND GENETICS, 2005, 22 (11-12) :379-387
[2]  
*AM ASS BIOAN, 2007, PROF TEST SERV
[3]  
*AM ASS BIOAN, 2008, PROF TEST SERV
[4]  
*AM SOC REPR MED P, 2006, FERTILITY STERILI S4, V86, P57
[5]   Assisted reproductive technology in Europe, 2003. Results generated from European registers by ESHRE [J].
Andersen, A. Nyboe ;
Goossens, V. ;
Gianaroli, L. ;
Felberbaum, R. ;
de Mouzon, J. ;
Nygren, K. G. .
HUMAN REPRODUCTION, 2007, 22 (06) :1513-1525
[6]   Interobserver agreement and intraobserver reproducibility of embryo quality assessments [J].
Arce, Joan-Carles ;
Ziebe, Soren ;
Lundin, Kersti ;
Janssens, Ronny ;
Helmgaard, Lisbeth ;
Sorensen, Per .
HUMAN REPRODUCTION, 2006, 21 (08) :2141-2148
[7]   A RAPID SPERM MOTILITY BIOASSAY PROCEDURE FOR QUALITY-CONTROL TESTING OF WATER AND CULTURE MEDIA [J].
BAVISTER, BD ;
ANDREWS, JC .
JOURNAL OF IN VITRO FERTILIZATION AND EMBRYO TRANSFER, 1988, 5 (02) :67-75
[8]   Interobserver and intraobserver variation in day 3 embryo grading [J].
Bendus, Allison E. Baxter ;
Mayer, Jacob F. ;
Shipley, Sharon K. ;
Catherino, William H. .
FERTILITY AND STERILITY, 2006, 86 (06) :1608-1615
[9]  
Cai Xia, 2006, Zhonghua Nan Ke Xue, V12, P625
[10]   Optimizing sensitivity of the human sperm motility assay for embryo toxicity testing [J].
Claassens, OE ;
Wehr, JB ;
Harrison, KL .
HUMAN REPRODUCTION, 2000, 15 (07) :1586-1591