Battery longevity from cardiac resynchronization therapy defibrillators: differences between manufacturers and discrepancies with published product performance reports

被引:17
作者
Alam, Mian Bilal [1 ]
Munir, Muhammad Bilal [1 ]
Rattan, Rohit [1 ]
Adelstein, Evan [1 ]
Jain, Sandeep [1 ]
Saba, Samir [1 ]
机构
[1] Univ Pittsburgh, Med Ctr, Inst Heart & Vasc, Pittsburgh, PA USA
来源
EUROPACE | 2017年 / 19卷 / 03期
关键词
Cardiac resynchronization therapy; Implantable cardioverter-defibrillator; Battery longevity; Manufacturer; Product performance report; IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS; CHRONIC HEART-FAILURE; REPLACE REGISTRY; LEAD EXTRACTION; INFECTIONS; PREVENTION; PACEMAKER;
D O I
10.1093/europace/euw044
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims Cardiac resynchronization therapy (CRT) is an important treatment for heart failure that requires constant ventricular pacing, placing a high energy burden on CRT defibrillators (CRT-D). Longer battery life reduces the need for device changes and associated complications, thereby affecting patient outcomes and cost of care. We therefore investigated the time to battery depletion of CRT-D from different manufacturers and compared these results with manufacturers' published product performance reports (PPRs). Methods and results All CRT-D recipients at our institution between January 2008 and December 2010 were included in this study cohort. The patients were followed up to the endpoint of battery depletion and were otherwise censored at the time of death, last follow-up, or device removal for any reason other than battery depletion. A total of 621 patients [ 173 Boston Scientific (BSC), 391 Medtronic (MDT), and 57 St. Jude Medical (SJM)] were followed up for a median of 3.7 (IQR 1.6-5.0) years, during which time 253 (41%) devices were replaced for battery depletion. Compared with MDT devices, battery depletion was 85 and 54% less likely to happen with BSC and SJM devices, respectively (P < 0.001 for pairwise comparisons). Product performance reports from all manufacturers significantly overestimated battery longevity by more than 20% 6 years after device implantation. Conclusions Large differences in CRT-D battery longevity exist between manufacturers. Industry-published PPRs significantly overestimate device longevity. These data have important implications to patients, healthcare professionals, hospitals, and third-party payers.
引用
收藏
页码:421 / 424
页数:4
相关论文
共 16 条
[1]   Cardiac resynchronization in chronic heart failure [J].
Abraham, WT ;
Fisher, WG ;
Smith, AL ;
Delurgio, DB ;
Leon, AR ;
Loh, E ;
Kocovic, DZ ;
Packer, M ;
Clavell, AL ;
Hayes, DL ;
Ellestad, M ;
Messenger, J ;
Trupp, RJ ;
Underwood, J ;
Pickering, F ;
Truex, C ;
McAtee, P .
NEW ENGLAND JOURNAL OF MEDICINE, 2002, 346 (24) :1845-1853
[2]   Battery longevity in cardiac resynchronization therapy implantable cardioverter defibrillators [J].
Alam, Mian Bilal ;
Munir, Muhammad Bilal ;
Rattan, Rohit ;
Flanigan, Susan ;
Adelstein, Evan ;
Jain, Sandeep ;
Saba, Samir .
EUROPACE, 2014, 16 (02) :246-251
[3]   Update on Cardiovascular Implantable Electronic Device Infections and Their Management A Scientific Statement From the American Heart Association [J].
Baddour, Larry M. ;
Epstein, Andrew E. ;
Erickson, Christopher C. ;
Knight, Bradley P. ;
Levison, Matthew E. ;
Lockhart, Peter B. ;
Masoudi, Frederick A. ;
Okum, Eric J. ;
Wilson, Walter R. ;
Beerman, Lee B. ;
Bolger, Ann F. ;
Estes, N. A. Mark, III ;
Gewitz, Michael ;
Newburger, Jane W. ;
Schron, Eleanor B. ;
Taubert, Kathryn A. .
CIRCULATION, 2010, 121 (03) :458-477
[4]   Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure [J].
Bristow, MR ;
Saxon, LA ;
Boehmer, J ;
Krueger, S ;
Kass, DA ;
De Marco, T ;
Carson, P ;
DiCarlo, L ;
DeMets, D ;
White, BG ;
DeVries, DW ;
Feldman, AM .
NEW ENGLAND JOURNAL OF MEDICINE, 2004, 350 (21) :2140-2150
[5]   Longevity of biventricular defibrillators: not all devices are created equal [J].
Burri, Haran .
EUROPACE, 2015, 17 (08) :1166-1168
[6]   The effect of cardiac resynchronization on morbidity and mortality in heart failure [J].
Cleland, JGF ;
Daubert, J ;
Erdmann, E ;
Freemantle, N ;
Gras, D ;
Kappenberger, L ;
Tavazzi, L ;
Cleland, JGF ;
Daubert, JC ;
Erdmann, E ;
Gras, D ;
Kappenberger, L ;
Klein, W ;
Tavazzi, L ;
Poole-Wilson, PA ;
Rydén, L ;
Wedel, H ;
Wellens, HJJ ;
Uretsky, B ;
Thygesen, K ;
Böcker, D ;
Marijianowski, MMH ;
Freemantle, N ;
Calvert, MJ ;
Christ, G ;
Fruhwald, F ;
Hofmann, R ;
Krypta, A ;
Leisch, F ;
Pacher, R ;
Rauscha, F ;
Tavernier, R ;
Thomsen, PEB ;
Boesgaard, S ;
Eiskjær, H ;
Esperen, GT ;
Haarbo, J ;
Hagemann, A ;
Korup, E ;
Moller, M ;
Mortensen, P ;
Sogaard, P ;
Vesterlund, T ;
Huikuri, H ;
Niemelä, KI ;
Toivonen, L ;
Bauer, F ;
Cohen-Solal, A ;
Crocq, C ;
Djiane, P .
NEW ENGLAND JOURNAL OF MEDICINE, 2005, 352 (15) :1539-1549
[7]   Percutaneous Pacemaker and Implantable Cardioverter-Defibrillator Lead Extraction in 100 Patients With Intracardiac Vegetations Defined by Transesophageal Echocardiogram [J].
Grammes, Jon A. ;
Schulze, Christopher M. ;
Al-Bataineh, Mohammad ;
Yesenosky, George A. ;
Saari, Christine S. ;
Vrabel, Michelle J. ;
Horrow, Jay ;
Chowdhury, Mashiul ;
Fontaine, John M. ;
Kutalek, Steven P. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (09) :886-894
[8]   Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators - Results of a large prospective study [J].
Klug, Didier ;
Balde, Mamadou ;
Pavin, Dominique ;
Hidden-Lucet, Francoise ;
Clementy, Jacques ;
Sadoul, Nicolas ;
Rey, Jean Luc ;
Lande, Gilles ;
Lazarus, Arnaud ;
Victor, Jacques ;
Barnay, Claude ;
Grandbastien, Bruno ;
Kacet, Salem .
CIRCULATION, 2007, 116 (12) :1349-1355
[9]   Longevity of implantable cardioverter-defibrillators for cardiac resynchronization therapy in current clinical practice: an analysis according to influencing factors, device generation, and manufacturer [J].
Landolina, Maurizio ;
Curnis, Antonio ;
Morani, Giovanni ;
Vado, Antonello ;
Ammendola, Ernesto ;
D'Onofrio, Antonio ;
Stabile, Giuseppe ;
Crosato, Martino ;
Petracci, Barbara ;
Ceriotti, Carlo ;
Bontempi, Luca ;
Morosato, Martina ;
Ballari, Gian Paolo ;
Gasparini, Maurizio .
EUROPACE, 2015, 17 (08) :1251-1258
[10]   The Cardiac Implantable Electronic Device Power Source: Evolution and Revolution [J].
Mond, Harry G. ;
Freitag, Gary .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2014, 37 (12) :1728-1745