Model Development and Experimental Validation for Analyzing Initial Transients of Irradiation of Tissues During Thermal Therapy Using Short Pulse Lasers

被引:18
作者
Ganguly, Mohit [1 ]
Miller, Stephanie [1 ]
Mitra, Kunal [1 ]
机构
[1] Florida Inst Technol, Dept Biomed Engn, Melbourne, FL 32901 USA
关键词
pulsed lasers; thermal dose; finite element modeling; histology; CONTINUOUS-WAVE; HEAT-TRANSFER; IN-VIVO; ULTRASOUND; SKIN; OPTIMIZATION; TEMPERATURE; EQUATION; FLOW;
D O I
10.1002/lsm.22407
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
Background and Objectives: Short pulse lasers with pulse durations in the range of nanoseconds and shorter are effective in the targeted delivery of heat energy for precise tissue heating and ablation. This photothermal therapy is useful where the removal of cancerous tissue sections is required. The objective of this paper is to use finite element modeling to demonstrate the differences in the thermal response of skin tissue to short-pulse and continuous wave laser irradiation in the initial stages of the irradiation. Models have been developed to validate the temperature distribution and heat affected zone during laser irradiation of excised rat skin samples and live anesthetized mouse tissue. Study Design/Materials and Methods: Excised rat skin samples and live anesthetized mice were subjected to Nd: YAG pulsed laser (1,064 nm, 500 ns) irradiation of varying powers. A thermal camera was used to measure the rise in surface temperature as a result of the laser irradiation. Histological analyses of the heat affected zone created in the tissue samples due to the temperature rise were performed. The thermal interaction of the laser with the tissue was quantified by measuring the thermal dose delivered by the laser. Finite element geometries of three-dimensional tissue sections for continuum and vascular models were developed using COMSOL Multiphysics. Blood flow was incorporated into the vascular model to mimic the presence of discrete blood vessels and contrasted with the continuum model without blood perfusion. Results: The temperature rises predicted by the continuum and the vascular models agreed with the temperature rises observed at the surface of the excised rat tissue samples and live anesthetized mice due to laser irradiation respectively. The vascular model developed was able to predict the cooling produced by the blood vessels in the region where the vessels were present. The temperature rise in the continuum model due to pulsed laser irradiation was higher than that due to continuous wave (CW) laser irradiation in the initial stages of the irradiation. The temperature rise due to pulsed and CW laser irradiation converged as the time of irradiation increased. A similar trend was observed when comparing the thermal dose for pulsed and CW laser irradiation in the vascular model. Conclusion: Finite element models (continuum and vascular) were developed that can be used to predict temperature rise and quantify the thermal dose resulting from laser irradiation of excised rat skin samples and live anesthetized mouse tissue. The vascular model incorporating blood perfusion effects predicted temperature rise better in the live animal tissue. The models developed demonstrated that pulsed lasers caused greater temperature rise and delivered a greater thermal dose than CW lasers of equal average power, especially during the initial transients of irradiation. This analysis will be beneficial for thermal therapy applications where maximum delivery of thermal dose over a short period of time is important. (c) 2015 Wiley Periodicals, Inc.
引用
收藏
页码:711 / 722
页数:12
相关论文
共 38 条
[1]  
Baish JW, 2009, ADV NUMER HEAT TRANS, V3, P29
[2]   Simultaneous spatial and temporal focusing for tissue ablation [J].
Block, Erica ;
Greco, Michael ;
Vitek, Dawn ;
Masihzadeh, Omid ;
Ammar, David A. ;
Kahook, Malik Y. ;
Mandava, Naresh ;
Durfee, Charles ;
Squier, Jeff .
BIOMEDICAL OPTICS EXPRESS, 2013, 4 (06) :831-841
[3]   ESTIMATION OF THE THERMAL EFFECT OF BLOOD-FLOW IN A BRANCHING COUNTERCURRENT NETWORK USING A 3-DIMENSIONAL VASCULAR MODEL [J].
BRINCK, H ;
WERNER, J .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1994, 116 (03) :324-330
[4]   Thermal dosimetry of a focused ultrasound beam in vivo by magnetic resonance imaging [J].
Chung, AH ;
Jolesz, FA ;
Hynynen, K .
MEDICAL PHYSICS, 1999, 26 (09) :2017-2026
[5]  
Darif M, 2008, C P COMSOL C 2008 HA
[6]   Modeling thermotherapy in vocal cords novel laser endoscopic treatment [J].
Fanjul-Velez, Felix ;
Arce-Diego, Jose Luis .
LASERS IN MEDICAL SCIENCE, 2008, 23 (02) :169-177
[7]   Model-based planning and real-time predictive control for laser-induced thermal therapy [J].
Feng, Yusheng ;
Fuentes, David .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2011, 27 (08) :751-761
[8]  
Ganguly M, 2015, HEAT TRANSFER FLUID
[9]   Optical-thermal light-tissue interactions during photoacoustic breast imaging [J].
Gould, Taylor ;
Wang, Quanzeng ;
Pfefer, T. Joshua .
BIOMEDICAL OPTICS EXPRESS, 2014, 5 (03) :832-847
[10]  
Jacques S.L., 1998, Skin Optics