Advanced near-infrared monitor for stable real-time measurement and control of Pichia pastoris bioprocesses

被引:24
作者
Goldfeld, Marina [1 ]
Christensen, Jens [1 ]
Pollard, David [1 ]
Gibson, Elizabeth R. [2 ]
Olesberg, Jonathon T. [2 ]
Koerperick, Edwin J. [2 ]
Lanz, Kaylee [2 ]
Small, Gary W. [2 ]
Arnold, Mark A. [2 ]
Evans, Christine E. [2 ]
机构
[1] Merck & Co Inc, BioProcess Dev, Kenilworth, NJ 07033 USA
[2] ASL Analyt Inc, BioVenture Ctr, Coralville, IA 52241 USA
关键词
near-infrared; Pichia pastoris; spectroscopy; mAb; fermentation; METHANOL CONCENTRATION; CELL-DENSITY; DIELECTRIC-SPECTROSCOPY; MASS-SPECTROMETRY; PILOT-SCALE; FERMENTATION; BATCH; CULTIVATIONS; OPTIMIZATION; EXPRESSION;
D O I
10.1002/btpr.1890
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Near-infrared spectroscopy is considered to be one of the most promising spectroscopic techniques for upstream bioprocess monitoring and control. Traditionally the nature of near-infrared spectroscopy has demanded multivariate calibration models to relate spectral variance to analyte concentrations. The resulting analytical measurements have proven unreliable for the measurement of metabolic substrates for bioprocess batches performed outside the calibration process. This paper presents results of an innovative near-infrared spectroscopic monitor designed to follow the concentrations of glycerol and methanol, as well as biomass, in real time and continuously during the production of a monoclonal antibody by a Pichia pastoris high cell density process. A solid state instrumental design overcomes the ruggedness limitations of conventional interferometer-based spectrometers. Accurate monitoring of glycerol, methanol, and biomass is demonstrated over 274 days postcalibration. In addition, the first example of feedback control to maintain constant methanol concentrations, as low as 1 g/L, is presented. Postcalibration measurements over a 9-month period illustrate a level of reliability and robustness that promises its adoption for online bioprocess monitoring throughout product development, from early laboratory research and development to pilot and manufacturing scale operation. (c) 2014 American Institute of Chemical Engineers Biotechnol. Prog., 30:749-759, 2014
引用
收藏
页码:749 / 759
页数:11
相关论文
共 32 条
[1]   On-line monitoring of responses to nutrient feed additions by multi-frequency permittivity measurements in fed-batch cultivations of CHO cells [J].
Ansorge, Sven ;
Esteban, Geoffrey ;
Schmid, Georg .
CYTOTECHNOLOGY, 2010, 62 (02) :121-132
[2]   A Review of Advanced Small-Scale Parallel Bioreactor Technology for Accelerated Process Development: Current State and Future Need [J].
Bareither, Rachel ;
Pollard, David .
BIOTECHNOLOGY PROGRESS, 2011, 27 (01) :2-14
[3]   Future aspects of bioprocess monitoring [J].
Becker, Thomas ;
Hitzmann, Bernd ;
Muffler, K. ;
Poertner, Ralf ;
Reardon, Kenneth F. ;
Stahl, Frank ;
Ulber, Roland .
WHITE BIOTECHNOLOGY, 2007, 105 :249-293
[4]   Improved production of monoclonal antibodies through oxygen-limited cultivation of glycoengineered yeast [J].
Berdichevsky, Marina ;
d'Anjou, Marc ;
Mallem, Muralidhar R. ;
Shaikh, Seemab S. ;
Potgieter, Thomas I. .
JOURNAL OF BIOTECHNOLOGY, 2011, 155 (02) :217-224
[5]  
Bilitewski U, 2005, COMP ANAL C, V44, P539, DOI 10.1016/S0166-526X(05)44011-8
[6]   Application of Near-Infrared Spectroscopy for Monitoring and Control of Cell Culture and Fermentation [J].
Cervera, Albert E. ;
Petersen, Nanna ;
Lantz, Anna Eliasson ;
Larsen, Anders ;
Gernaey, Krist V. .
BIOTECHNOLOGY PROGRESS, 2009, 25 (06) :1561-1581
[7]   On-line near-infrared spectrometer to monitor urea removal in real time during hemodialysis [J].
Cho, David S. ;
Olesberg, Jonathon T. ;
Flanigan, Michael J. ;
Arnold, Mark A. .
APPLIED SPECTROSCOPY, 2008, 62 (08) :866-872
[8]   Monitoring a high cell density recombinant Pichia pastoris fed-batch bioprocess using transmission and reflectance near infrared spectroscopy [J].
Crowley, J ;
Arnold, SA ;
Wood, N ;
Harvey, LM ;
McNeil, B .
ENZYME AND MICROBIAL TECHNOLOGY, 2005, 36 (5-6) :621-628
[9]   Expression of heterologous proteins in Pichia pastoris:: a useful experimental tool in protein engineering and production [J].
Daly, R ;
Hearn, MTW .
JOURNAL OF MOLECULAR RECOGNITION, 2005, 18 (02) :119-138
[10]   A novel approach for using dielectric spectroscopy to predict viable cell volume (VCV) in early process development [J].
Downey, Brandon J. ;
Graham, Lisa J. ;
Breit, Jeffrey F. ;
Glutting, Nathaniel K. .
BIOTECHNOLOGY PROGRESS, 2014, 30 (02) :479-487