Effect of contact area on the wear of ultrahigh molecular weight polyethylene in noncyclic pin-on-disk tests

被引:23
作者
Saikko, Vesa [1 ]
机构
[1] Aalto Univ, Sch Engn, Dept Mech Engn, POB 14300, FI-00076 Aalto, Finland
关键词
Biotribology; Prosthetic joint; Noncyclic pin-on-disk; UHMWPE; TOTAL HIP-ARTHROPLASTY; ORTHOPEDIC BIOMATERIALS; ACETABULAR COMPONENTS; SIMULATION; MOTION; DEVICE;
D O I
10.1016/j.triboint.2017.04.020
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The 16-station RandomPOD pin-on-disk device with non-cyclic biaxial motion and load was used to study the effect of contact area on the wear of ultrahigh molecular weight polyethylene as used in total hip prostheses. The counterface was polished CoCr and the lubricant was diluted serum. The wear factor increased strongly and linearly with increasing contact area, which was in agreement with clinical findings for total hip prostheses. The coefficient of friction also increased strongly with increasing contact area. With small contact area the lumpy topography of the worn surface differed from the burnished appearance typical of retrieved acetabular cups. The nominal contact pressure should be kept below 2 MPa in pin-on-disk tests intended to simulate clinical wear mechanisms.
引用
收藏
页码:84 / 87
页数:4
相关论文
共 16 条
[1]   Advances in tribological testing of artificial joint biomaterials using multidirectional pin-on-disk testers [J].
Baykal, D. ;
Siskey, R. S. ;
Haider, H. ;
Saikko, V. ;
Ahlroos, T. ;
Kurtz, S. M. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2014, 31 :117-134
[2]  
Bragdon C R, 1996, Proc Inst Mech Eng H, V210, P157, DOI 10.1243/PIME_PROC_1996_210_408_02
[3]   Mechanical behavior, wear surface morphology, and clinical performance of UHMWPE acetabular components after 10 years of implantation [J].
Edidin, AA ;
Rimnac, CM ;
Goldberg, VM ;
Kurtz, SM .
WEAR, 2001, 250 (PART 1) :152-158
[4]  
Hall R M, 1996, Proc Inst Mech Eng H, V210, P197, DOI 10.1243/PIME_PROC_1996_210_413_02
[5]   Development of a BiotriboPOD testing methodology for the wear evaluation of orthopaedic biomaterials [J].
Hua, Zikai ;
Zhang, Huanhuan ;
Fan, Yongwei ;
Jin, Zhongmin .
RSC ADVANCES, 2014, 4 (38) :19987-19991
[6]   Wear of polyethylene acetabular components in total hip arthroplasty - An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations [J].
Jasty, M ;
Goetz, DD ;
Bragdon, CR ;
Lee, KR ;
Hanson, AE ;
Elder, JR ;
Harris, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (03) :349-358
[7]  
Joyce TJ, 2000, BIO-MED MATER ENG, V10, P241
[8]   How Does Wear Rate Compare in Well-functioning Total Hip and Knee Replacements? A Postmortem Polyethylene Liner Study [J].
Pourzal, Robin ;
Knowlton, Christopher B. ;
Hall, Deborah J. ;
Laurent, Michel P. ;
Urban, Robert M. ;
Wimmer, Markus A. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2016, 474 (08) :1867-1875
[9]   Type of motion and lubricant in wear simulation of polyethylene acetabular cup [J].
Saikko, V ;
Ahlroos, T .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H4) :301-310
[10]   Effect of contact pressure on wear and friction of ultra-high molecular weight polyethylene in multidirectional sliding [J].
Saikko, V. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H7) :723-731