Reduction of radiation delivered to patients undergoing invasive coronary procedures. Effect of a programme for dose reduction based on radiation-protection training

被引:47
|
作者
Georges, Jean-Louis [1 ]
Livarek, Bernard [1 ]
Gibault-Genty, Geraldine [1 ]
Aziza, Jean-Paul [1 ]
Hautecoeur, Jean-Louis [1 ]
Soleille, Henry [1 ]
Messaoudi, Hela [1 ]
Annabi, Nizar [1 ]
Kebsi, Mohammed Al [1 ]
机构
[1] Hop Andre Mignot, Ctr Hosp Versailles, Serv Cardiol, F-78150 Le Chesnay, France
关键词
Radiation protection; Training; Dose area product; Coronary angiography; Percutaneous coronary intervention; INTERVENTIONS; ANGIOPLASTY; DOSAGE;
D O I
10.1016/j.acvd.2009.09.007
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. - Exposure of patients to radiation from invasive cardiac procedures is high and may be deleterious. Aims. - To assess the effectiveness of a dose-reduction programme based on radiation-protection training, according to the recommendations of the Euratom Council, the International Commission on Radiological Protection and the French Society of Cardiology. Methods. - In this single-centre survey, dose-area product (DAP, Gy.cm(2)), fluoroscopy time (minutes) and number of runs were evaluated in 3285 consecutive procedures (2077 coronary angiographies [CAs], 1208 percutaneous coronary interventions [PCIs]), performed one year before (2005) and two years after (2006 to 2007) implementation of a programme for radiation dose-reduction. The programme included a 2-day training course in radiological protection for all medical and paramedical staff and recommendations for routine use of low fluoroscopic and acquisition pulse rates (6.25 and 12.5 i/s, respectively), large field size (23 cm), maximal. collimation and optimal X-ray tube/patient/detector distances. Routine left ventriculography was discouraged. The radial approach was used in > 80% of the procedures. Results. - Compared with 2005, a significant 50% reduction in DAP was observed in 2006 and 2007 during CA (median [interquartile range] 53 Gy.cm(2) [33-84] vs 26 [16-43] and 21 [14-32], respectively; p < 0.0001) and PCI (125 Gy.cm(2) [78-184] vs 49 [31-79] and 44 [27-66], respectively; p < 0.0001). Fluoroscopy time and number of runs did not vary significantly in 2006, and decreased slightly in 2007, likely due to an important reduction in rate of left ventriculographies (from 32 to 4%). Inter-operator variability in DAP was reduced. Conclusion. - Training in radiation protection for interventional cardiologists and use of simple and cost-free dose-reduction techniques were associated with a 50% reduction in radiation exposure to patients undergoing invasive cardiac procedures, without any loss of diagnostic information. (C) 2009 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:821 / 827
页数:7
相关论文
共 50 条
  • [1] Radiation dose reduction during venous access port implantation: the importance of upgrading equipment and radiation-protection training
    Vallot, Delphine
    Cavillon, Ana
    Lusque, Amelie
    Izard, Philippe
    Salvignol, Genevieve
    Delpuech, Bertrand
    Fuzier, Regis
    IRISH JOURNAL OF MEDICAL SCIENCE, 2024, 193 (03) : 1461 - 1466
  • [2] Effective techniques for reduction of radiation dosage to patients undergoing invasive cardiac procedures
    Kuon, E
    Glaser, C
    Dahm, JB
    BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (906): : 406 - 413
  • [3] Effectiveness of a program for reduction of radiation dosage to patients undergoing invasive cardiac procedures
    Georges, J. L.
    Livarek, B.
    Gibault-Genty, G.
    Aziza, J. P.
    Hautecoeur, J. L.
    Soleille, H.
    Messaoudi, H.
    Annabi, N.
    EUROPEAN HEART JOURNAL, 2007, 28 : 861 - 861
  • [4] Radiation Dose Reduction Using a Novel Fluoroscopy System in Patients Undergoing Diagnostic Invasive Coronary Angiography
    Kraemer, Bjoern F.
    Tesche, Christian
    Hapfelmeier, Alexander
    Finsterer, Sebastian
    Ebersberger, Ullrich
    Schluter, Mathis
    Ziegler, Martin
    Martens, Eimo
    Laugwitz, Karl-Ludwig
    Schmidt, Martin
    JOURNAL OF THORACIC IMAGING, 2021, 36 (01) : 52 - 56
  • [5] Reduction of the radiation dose received by interventional cardiologists following training in radiation protection
    Abatzoglou, I.
    Koukourakis, M.
    Konstantinides, S.
    RADIATION PROTECTION DOSIMETRY, 2013, 155 (01) : 119 - 121
  • [6] Reduction of radiation dose to patients undergoing barium enema by dose audit
    Yu, SK
    Cheung, YK
    Chan, TL
    Kung, CM
    Yuen, MK
    BRITISH JOURNAL OF RADIOLOGY, 2001, 74 (878): : 162 - 165
  • [7] An education and training programme for radiological institutes: impact on the reduction of the CT radiation dose
    Schindera, Sebastian T.
    Treier, Reto
    von Allmen, Gabriel
    Nauer, Claude
    Trueb, Philipp R.
    Vock, Peter
    Szucs-Farkas, Zsolt
    EUROPEAN RADIOLOGY, 2011, 21 (10) : 2039 - 2045
  • [8] An education and training programme for radiological institutes: impact on the reduction of the CT radiation dose
    Sebastian T. Schindera
    Reto Treier
    Gabriel von Allmen
    Claude Nauer
    Philipp R. Trueb
    Peter Vock
    Zsolt Szucs-Farkas
    European Radiology, 2011, 21 : 2039 - 2045
  • [9] Variations of radiation dosage delivered to patients undergoing interventional cardiological procedures. A monocentric study 2002-05
    Georges, J. L.
    Livarek, B.
    Gibault-Genty, G.
    Messaoudi, H.
    Aziza, J. P.
    Hautecoeur, J. L.
    Soleille, H.
    ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 2007, 100 (03): : 175 - 181
  • [10] Radiation dose reduction in minimally-invasive intravascular procedures using a magnetic guidance system
    Tanase, D
    Bakker, NH
    van Loon, D
    Goosen, JFL
    Trimp, PJ
    Reekers, JA
    French, PJ
    2ND ANNUAL INTERNATIONAL IEEE-EMBS SPECIAL TOPIC CONFERENCE ON MICROTECHNOLOGIES IN MEDICINE & BIOLOGY, PROCEEDINGS, 2002, : 305 - 308