Enhanced a-Si/CsI-based flat panel X-ray detector for mammography

被引:11
作者
Shaw, J [1 ]
Albagli, D [1 ]
Wei, CY [1 ]
Granfors, P [1 ]
机构
[1] GE Global Res Ctr, Niskayuna, NY 12309 USA
来源
MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2 | 2004年 / 5368卷
关键词
D O I
10.1117/12.539141
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The GE Senographe 2000D, the first full field digital mammography system based on amorphous silicon (a-Si) flat panel arrays and a cesium iodide (CsI) scintillator, has been in clinical use for over five years. One of the major advantages of this technology platform over competing platforms is the inherent flexibility of the design. Specifically, it is possible to optimize the x-ray conversion layer (scintillator) independently of the light conversion layer (panel) and vice versa. This is illustrated by a new detector utilizing the same amorphous silicon (a-Si) flat panel design, but an optimized scintillator layer, which provides up to 15% higher DQE than the existing detector. By utilizing the existing flat panel with an optimized scintillator layer, it is possible to significantly boost performance without changes to the panel design. Future enhancements to both the panel and scintillator will raise the DQE at zero frequency to more than 80%. The a-Si/CsI platform is especially well suited to advanced applications utilizing very low doses.
引用
收藏
页码:370 / 378
页数:9
相关论文
共 9 条
[1]   Performance of advanced a-Si/CsI-based flat panel x-ray detectors for mammography [J].
Albagli, D ;
Hudspeth, H ;
Possin, G ;
Lee, J ;
Granfors, P ;
Giambattista, B .
MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 :553-563
[2]  
GRANFORS PR, 2002, P SOC PHOTO-OPT INS, V3977, P2
[3]   Signal-to-noise ratio and spatial resolution in x-ray electronic imagers: Is the MTF a relevant parameter? [J].
Moy, JP .
MEDICAL PHYSICS, 2000, 27 (01) :86-93
[4]   Microcalcification detectability for four mammographic detectors: Flat-panel, CCD, CR, and screen/film [J].
Rong, XJJ ;
Shaw, CS ;
Johnston, DA ;
Lemacks, MR ;
Liu, XM ;
Whitman, GJ ;
Dryden, MJ ;
Stephens, TW ;
Thompson, SK ;
Krugh, KT ;
Lai, CJ .
MEDICAL PHYSICS, 2002, 29 (09) :2052-2061
[5]  
SMITH A, 2002, RADIOLOGY MANAGEMENT, V25
[6]   The x-ray sensitivity of amorphous selenium for mammography [J].
Stone, MF ;
Zhao, W ;
Jacak, BV ;
O'Connor, P ;
Yu, B ;
Rehak, P .
MEDICAL PHYSICS, 2002, 29 (03) :319-324
[7]   CALCULATION OF MODULATION TRANSFER-FUNCTIONS OF X-RAY FLUORESCENT SCREENS [J].
SWANK, RK .
APPLIED OPTICS, 1973, 12 (08) :1865-1870
[8]   Full breast digital mammography with an amorphous silicon-based flat panel detector: Physical characteristics of a clinical prototype [J].
Vedantham, S ;
Karellas, A ;
Suryanarayanan, S ;
Albagli, D ;
Han, S ;
Tkaczyk, EJ ;
Landberg, CE ;
Opsahl-Ong, B ;
Granfors, PR ;
Levis, I ;
D'Orsi, CJ ;
Hendrick, RE .
MEDICAL PHYSICS, 2000, 27 (03) :558-567
[9]   Characterization of a full field digital mammography detector based on direct x-ray conversion in selenium [J].
Yorker, JG ;
Jeromin, LS ;
Lee, DLY ;
Palecki, EF ;
Golden, KP ;
Jing, ZX .
MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 :21-29