Assay of creatinine using the peroxidase activity of copper-creatinine complexes

被引:38
作者
Pugia, MJ
Lott, JA
Wallace, JF
Cast, TK
Bierbaum, LD
机构
[1] Ohio State Univ, Ctr Med, Dept Pathol, Columbus, OH 43210 USA
[2] Bayer Corp, Diagnost Business Grp, Elkhart, IN USA
关键词
renal failure; albuminuria; albumin/creatinine ratio; copper binding; test interferences; free-radical scavengers; oxidizing agents; ascorbic acid; 3,3 ',5,5 '-tetramethyl benzidine; peroxidase;
D O I
10.1016/S0009-9120(99)00086-7
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: It was our goal to develop a urine dipstick that could measure creatinine with a peroxidase reaction. The simultaneous measurement of albumin and creatinine permits the estimation of the 24-h albumin excretion, an important value in judging existing or likely development of renal failure. A highly sensitive dye-binding dipstick method for albumin exists, and a suitable dipstick for the assay for urine creatinine is described here. Methods: Copper-creatinine and iron-creatinine complexes have peroxidase activity. With 3,3',5,5'-tetramethylbenzidine (TMB), and diisopropyl benzene dihydroperoxide (DBDH); the peroxidase activity of copper-creatinine and iron-creatinine complexes can be demonstrated. This reaction was used in the assay of urine creatinine either in solution or by a suitably impregnated urine dipstick. Results: Our method based on the peroxidase activity of the copper-creatinine complex has an analytical range for creatinine of 100 mg/L (0.884 mmol/L) to 3000 mg/L (26.52 mmol/L). The creatinine assay is free from most interfering compounds that may be present in urine. Hemoglobin is an interferent, and its effects can be reduced but not eliminated by the addition of 4-hydroxy-2-methyl quinoline. We do not recommend using the dipsticks when visible blood is present or if the dipstick blood test is positive. The copper-creatinine complex oxidizes ascorbic acid; however, we were able to modify the reaction conditions so that ascorbic acid at < 4.4 g/L does not interfere. We found good agreement on fresh urines between the creatinine dipstick results and those by a standard rate-Jaffe cuvet method for creatinine. Discussion: With the simultaneous measurement of creatinine and albumin in urine, the albumin/creatinine ratio can be determined effectively reducing or eliminating the occasional false-negative and false-positive result in those with dilute or concentrated urines, respectively. The dipstick test for these analytes permits the simple identification of individuals with possible albuminuria and could serve well in a point-of-care setting. Copyright (C) 2000 The Canadian Society of Clinical Chemists.
引用
收藏
页码:63 / 70
页数:8
相关论文
共 23 条
[1]   IMPROVED COLOR REAGENT FOR DETERMINATION OF BLOOD-GLUCOSE BY OXIDASE SYSTEM [J].
BARHAM, D ;
TRINDER, P .
ANALYST, 1972, 97 (1151) :142-&
[2]  
*BAYER DIAGN, INSTR OP BAYER OP AN
[3]  
Bright H.B., 1975, ENZYMES, V12B, P421
[4]   MICROALBUMINURIA - IMPLICATIONS FOR MICROVASCULAR AND MACROVASCULAR DISEASE [J].
DECKERT, T ;
KOFOEDENEVOLDSEN, A ;
NORGAARD, K ;
BORCHJOHNSEN, K ;
FELDTRASMUSSEN, B ;
JENSEN, T .
DIABETES CARE, 1992, 15 (09) :1181-1191
[5]   USE OF SINGLE VOIDED URINE SAMPLES TO ESTIMATE QUANTITATIVE PROTEINURIA [J].
GINSBERG, JM ;
CHANG, BS ;
MATARESE, RA ;
GARELLA, S .
NEW ENGLAND JOURNAL OF MEDICINE, 1983, 309 (25) :1543-1546
[6]  
JAFFE RM, 1979, AM J CLIN PATHOL, V72, P468
[7]  
JOSEPHY PD, 1982, J BIOL CHEM, V257, P3669
[8]  
KLEBANOFF SJ, 1989, J BIOL CHEM, V264, P19765
[9]   CREATININE ASSAY BY A REACTION-KINETIC PRINCIPLE [J].
LARSEN, K .
CLINICA CHIMICA ACTA, 1972, 41 (01) :209-&
[10]   A 4-MEMBERED CHELATE COMPLEX OF CU(II) WITH CREATININE [J].
MITEWA, M ;
BONTCHEV, PR ;
KABASSANOV, K .
POLYHEDRON, 1985, 4 (07) :1159-1161