Do endoscopy capsules interfere with implantable cardioverter-defibrillators?

被引:20
作者
Bandorski, D. [1 ]
Irnich, W. [2 ]
Brueck, M. [1 ]
Kramer, W. [1 ]
Jakobs, R. [3 ]
机构
[1] Med Klin 1, Klinikum Wetzlar, D-35578 Wetzlar, Germany
[2] Univ Giessen, Univ Klinikum, Giessen, Germany
[3] Med Klin C, Klinikum Ludwigshafen, Ludwigshafen, Germany
关键词
PILLCAM COLON CAPSULE; ELECTROMAGNETIC-INTERFERENCE; CARDIAC-PACEMAKERS; VIDEO-CAPSULE; MULTICENTER; TRIAL;
D O I
10.1055/s-0029-1214610
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background and study aims: Capsule endoscopy is already an established diagnostic tool, and the newly introduced capsule PillCam COLON is now entering clinical studies. Because of the very limited clinical experience in patients with implantable cardioverter-defibrillators (ICDs), it is generally recommended that capsule endoscopy Should not be used in these patients. Therefore, we investigated, in vitro, the possible interference between three different endoscopy capsules (Given Imaging and Olympus) and ICDs. Materials and methods: A total of 45 ICD devices were separately placed in a tank filled with a 0.9% saline solution. With the help of pin jacks, the pacing pulses of ICDs were recorded and inhibiting signals could be coupled in. The capsules were placed in different positions near to the lead and the ICD device, and finally placed on the case of the device for 1 minute each. The function of each device was observed continuously. Results: Even when the capsules were in closest proximity to the ICDs, no interference was observed. Conclusion: The clinical use of the three tested capsule devices is safe in patients with ICDs.
引用
收藏
页码:457 / 461
页数:5
相关论文
共 15 条
[1]   Capsule endoscopy and cardiac pacemakers:: investigation for possible interference [J].
Bandorski, D. ;
Irnich, W. ;
Brueck, M. ;
Beyer, N. ;
Kramer, W. ;
Jakobs, R. .
ENDOSCOPY, 2008, 40 (01) :36-39
[2]   Capsule endoscopy in patients with cardiac pacemakers: Current situation in Germany [J].
Bandorski, D ;
Diehl, KL ;
Jaspersen, D .
ZEITSCHRIFT FUR GASTROENTEROLOGIE, 2005, 43 (08) :715-718
[3]   Electromagnetic interference of analog cellular telephones with pacemakers [J].
Barbaro, V ;
Bartolini, P ;
Donato, A ;
Militello, C .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (10) :1410-1418
[4]  
*CEN CENELEC, 45502212004 EN CENCE
[5]   Electromagnetic interference with implantable cardiac pacemakers by video capsule [J].
Dubner, S ;
Dubner, Y ;
Gallino, S ;
Spallone, L ;
Zagalsky, D ;
Rubio, H ;
Zimmerman, J ;
Goldin, E .
GASTROINTESTINAL ENDOSCOPY, 2005, 61 (02) :250-254
[6]   Electromagnetic interference from wireless video-capsule endoscopy on implantable cardioverter-defibrillators [J].
Dubner, Sergio ;
Dubner, Yael ;
Rubio, Horacio ;
Goldin, Eran .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2007, 30 (04) :472-475
[7]   Evaluation of the PillCam Colon capsule in the detection of colonic pathology: results of the first multicenter, prospective, comparative study [J].
Eliakim, R. ;
Fireman, Z. ;
Gralnek, I. M. ;
Yassin, K. ;
Waterman, M. ;
Kopelman, Y. ;
Lachter, J. ;
Koslowsky, B. ;
Adler, S. N. .
ENDOSCOPY, 2006, 38 (10) :963-970
[8]   The first prospective controlled trial comparing wireless capsule endoscopy with push enteroscopy in chronic gastrointestinal bleeding [J].
Ell, C ;
Remke, S ;
May, A ;
Helou, L ;
Henrich, R ;
Mayer, G .
ENDOSCOPY, 2002, 34 (09) :685-689
[9]   Electromagnetic interference of pacemakers by mobile phones [J].
Irnich, W ;
Batz, L ;
Muller, R ;
Tobisch, R .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (10) :1431-1446
[10]  
Irnich W., 2004, HERZSCHR ELEKTROPHYS, V15, P9