PET characteristics of a dedicated breast PET/CT scanner prototype

被引:40
作者
Wu, Yibao [1 ]
Bowen, Spencer L. [1 ]
Yang, Kai [1 ]
Packard, Nathan [1 ]
Fu, Lin [1 ]
Burkett, George, Jr. [2 ]
Qi, Jinyi [1 ]
Boone, John M. [2 ]
Cherry, Simon R. [1 ]
Badawi, Ramsey D. [2 ]
机构
[1] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
[2] Univ Calif Davis, Med Ctr, Dept Radiol, Sacramento, CA 95817 USA
基金
美国国家卫生研究院;
关键词
POSITRON-EMISSION-TOMOGRAPHY; IMAGING-SYSTEM; RESOLUTION; MAMMOGRAPHY; RECONSTRUCTION; PERFORMANCE; FLUORODEOXYGLUCOSE; CONSTRUCTION; SIMULATION; DETECTORS;
D O I
10.1088/0031-9155/54/13/020
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A dedicated breast PET/CT system has been constructed at our institution, with the goal of having increased spatial resolution and sensitivity compared to whole-body systems. The purpose of this work is to describe the design and the performance characteristics of the PET component of this device. Average spatial resolution of a line source in warm background using maximum a posteriori (MAP) reconstruction was 2.5 mm, while the average spatial resolution of a phantom containing point sources using filtered back projection (FBP) was 3.27 mm. A sensitivity profile was computed with a point source translated across the axial field of view (FOV) and a peak sensitivity of 1.64% was measured at the center of the FOV. The average energy resolution determined on a per-crystal basis was 25%. The characteristic dead time for the front-end electronics and data acquisition (DAQ) was determined to be 145 ns and 3.6 mu s, respectively. With no activity outside the FOV, a peak noise-equivalent count rate of 18.6 kcps was achieved at 318 mu Ci (11.766 MBq) in a cylindrical phantom of diameter 75 mm. After the effects of exposing PET detectors to x-ray flux were evaluated and ameliorated, a combined PET/CT scan was performed. The percentage standard deviations of uniformity along axial and transaxial directions were 3.7% and 2.8%, respectively. The impact of the increased reconstructed spatial resolution compared to typical whole-body PET scanners is currently being assessed in a clinical trial.
引用
收藏
页码:4273 / 4287
页数:15
相关论文
共 44 条
[1]   Clear-PEM:: A PET imaging system dedicated to breast cancer diagnostics [J].
Abreu, M. C. ;
Aguiar, D. ;
Albuquerque, E. ;
Almeida, F. G. ;
Almeida, P. ;
Amaral, P. ;
Auffray, E. ;
Bento, P. ;
Bruyndonckx, P. ;
Bugalho, R. ;
Carrico, B. ;
Cordeiro, H. ;
Ferreira, M. ;
Ferreira, N. C. ;
Goncalves, F. ;
Lecoq, P. ;
Leong, C. ;
Lopes, F. ;
Lousa, P. ;
Luyten, J. ;
Martins, M. V. ;
Matela, N. ;
Mendes, P. R. ;
Moura, R. ;
Nobre, J. ;
Oliveira, N. ;
Ortigao, C. ;
Peralta, L. ;
Rego, J. ;
Ribeiro, R. ;
Rodrigues, P. ;
Santos, A. I. ;
Silva, J. C. ;
Silva, M. M. ;
Tavernier, S. ;
Teixeira, I. C. ;
Teixeira, J. P. ;
Trindade, A. ;
Trurnmer, J. ;
Varela, J. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2) :81-84
[2]   Breast imaging with positron emission tomography and fluorine-18 fluorodeoxyglucose:: Use and limitations [J].
Avril, N ;
Rosé, CA ;
Schelling, M ;
Dose, J ;
Kuhn, W ;
Bense, S ;
Weber, W ;
Ziegler, S ;
Graeff, H ;
Schwaiger, M .
JOURNAL OF CLINICAL ONCOLOGY, 2000, 18 (20) :3495-3502
[3]   High-resolution fluorodeoxyglucose positron emission tomography with compression ("positron emission mammography") is highly accurate in depicting primary breast cancer [J].
Berg, Wendie A. ;
Weinberg, Irving N. ;
Narayanan, Deepa ;
Lobrano, Mary E. ;
Ross, Eric ;
Amodei, Laura ;
Tafra, Lorraine ;
Adler, Lee P. ;
Uddo, Joseph ;
Stein, William, III ;
Levine, Edward A. .
BREAST JOURNAL, 2006, 12 (04) :309-323
[4]   Putting 'clear' into nuclear medicine: a decade of PET/CT development [J].
Beyer, Thomas ;
Townsend, David W. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2006, 33 (08) :857-861
[5]   Technique factors and their relationship to radiation dose in pendant geometry breast CT [J].
Boone, JM ;
Kwan, ALC ;
Seibert, JA ;
Shah, N ;
Lindfors, KK ;
Nelson, TR .
MEDICAL PHYSICS, 2005, 32 (12) :3767-3776
[6]   Computed tomography for imaging the breast [J].
Boone, John M. ;
Kwan, Alex L. C. ;
Yang, Kai ;
Burkett, George W. ;
Lindfors, Karen K. ;
Nelson, Thomas R. .
JOURNAL OF MAMMARY GLAND BIOLOGY AND NEOPLASIA, 2006, 11 (02) :103-111
[7]  
Bowen SL, 2006, EUR J NUCL MED MOL I, V33, pS117
[8]   Comparison of 3-D maximum a posteriori and filtered backprojection algorithms for high-resolution animal imaging with microPET [J].
Chatziioannou, A ;
Qi, J ;
Moore, A ;
Annala, A ;
Nguyen, K ;
Leahy, R ;
Cherry, SR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (05) :507-512
[9]   Crystal identification in positron emission tomography using nonrigid registration to a Fourier-based template [J].
Chaudhari, Abhijit J. ;
Joshi, Anand A. ;
Bowen, Spencer L. ;
Leahy, Richard M. ;
Cherry, Simon R. ;
Badawi, Ramsey D. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (18) :5011-5027
[10]   PSPMT/APD hybrid DOI detectors for the PET component of a dedicated breast PET/CT system - A feasibility study [J].
Chaudhari, Abhijit J. ;
Yang, Yongfeng ;
Farrell, Richard ;
Dokhale, Purushottam A. ;
Shah, Kanai S. ;
Cherry, Simon R. ;
Badawi, Ramsey D. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (03) :853-861