GEANT4;
Monte Carlo;
Gamma spectroscopy;
Compton suppression;
SYSTEM;
D O I:
10.1007/s10967-014-3029-x
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
GEANT4 Monte Carlo simulations have been successfully utilised to characterise a Compton suppressed broad-energy HPGe detector. The detector setup has been fully recreated in the simulation, which has been optimised to consistently reproduce the detector response. The peak efficiencies for both the primary BEGe detector and NaI(Tl) guard detectors agree with the simulated values for multiple test sources within 3 %. Compton suppression has also been simulated, with good agreement seen between the simulated and actual CSF values (< 10 %) for multiple radionuclides. A secondary reference source was also simulated, which contained up to 30 radionuclides in a different geometry to that of the previous source. This showed excellent agreement with experimental data in both unsuppressed and suppressed modes of operation.