RF penetration in ultra high field MRI: Challenges in visualizing details within the center of the human brain

被引:15
作者
Robitaille, PML [1 ]
Kangarlu, A [1 ]
Abduljalil, AM [1 ]
机构
[1] Ohio State Univ, MRI Facil, Dept Radiol, Ctr Adv Biomed Imaging, Columbus, OH 43210 USA
关键词
brain; magnetic resonance imaging; techniques; anatomy; tesla; image quality; physics and instrumentation;
D O I
10.1097/00004728-199911000-00006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this work is to discuss radio frequency (RF) penetration and its relevance to imaging the human head and to acquire images containing intricate structures located at the center of the brain with ultra high field MRI (UHFMRI). Method: A simple plane wave analysis of RF penetration was performed based on Maxwell equations as a function of frequency up to 900 MHz. Gradient-recalled images were acquired at 8 T (340 MHz) using an RF resonator operating in quadrature. Typical acquisition parameters were as follows: TR = 750 ms, TE = 17 ms, slice thickness = 2 mm, FOV = 20 x 20 cm, matrix = 1,024 x 1,024. The specific absorption rate was well below 1 W/kg. Results: A simple analytical treatment, for a plane wave up to 900 MHz, reveals a lack of decreasing penetration depth with frequency beyond 200 MHz. Gradient-recalled echo images acquired from the human head displayed good contrast, homogeneity, and resolution. Importantly, excellent structural detail was observed on the resulting MR images, demonstrating that RF penetration is not a problem at 8 T. Images reveal excellent detail including the red nucleus, anterior commissure, fornix, mamillary body, pineal gland, and ependymal lining of the fourth ventricle. Conclusion: Structures located at the center of the human brain can be clearly visualized at 8 T with no detectable loss in signal intensity arising from RF penetration. The ability to examine these structures with UHFMRI will provide a powerful new modality for diagnostic radiology.
引用
收藏
页码:845 / 849
页数:5
相关论文
共 30 条
[1]   Acquisition of human multislice MR images at 8 Tesla [J].
Abduljalil, AM ;
Kangarlu, A ;
Zhang, XL ;
Burgess, RE ;
Robitaille, PML .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1999, 23 (03) :335-340
[2]  
BAERLEIN BA, IN PRESS IEEE T BIOM
[3]  
Barfuss H, 1990, NMR Biomed, V3, P31, DOI 10.1002/nbm.1940030106
[4]   POWER DEPOSITION IN WHOLE-BODY NMR IMAGING [J].
BOTTOMLEY, PA ;
EDELSTEIN, WA .
MEDICAL PHYSICS, 1981, 8 (04) :510-512
[5]  
BURGESS RE, 1998, INT SOC MAGN RES MED, P25
[6]   THE FIELD-DEPENDENCE OF NMR IMAGING .1. LABORATORY ASSESSMENT OF SIGNAL-TO-NOISE RATIO AND POWER DEPOSITION [J].
CHEN, CN ;
SANK, VJ ;
COHEN, SM ;
HOULT, DI .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (05) :722-729
[7]   SAR and B1 field distributions in a heterogeneous human head model within a birdcage coil [J].
Collins, CM ;
Li, SZ ;
Smith, MB .
MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (06) :847-856
[9]   SENSITIVITY OF THE ZEUGMATOGRAPHIC EXPERIMENT INVOLVING HUMAN SAMPLES [J].
HOULT, DI ;
LAUTERBUR, PC .
JOURNAL OF MAGNETIC RESONANCE, 1979, 34 (02) :425-433
[10]  
IBRAHIM TS, 1998, P INT SOC MAGN RES M, P652