Clinical implementation of pencil beam scanning proton therapy for liver cancer with forced deep expiration breath hold

被引:10
|
作者
Fracchiolla, Francesco [1 ]
Dionisi, Francesco [1 ]
Righetto, Roberto [1 ]
Widesott, Lamberto [1 ]
Giacomelli, Irene [1 ]
Cartechini, Giorgio [2 ]
Farace, Paolo [1 ]
Bertolini, Mattia [1 ]
Amichetti, Maurizio [1 ]
Schwarz, Marco [1 ,3 ]
机构
[1] Azienda Prov & Serv Sanitari APSS, Protontherapy Dept, Via Al Desert 14, I-38122 Trento, Italy
[2] Univ Trento, Trento, Italy
[3] TIFPA Trento Inst Fundamental Phys & Applicat, Trento, Italy
关键词
Proton therapy; Liver treatments; Interplay effect; Robustness analysis;
D O I
10.1016/j.radonc.2020.09.035
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To present our technique for liver cancer treatments with proton therapy in pencil beam scanning mode and to evaluate the impact of uncertainties on plan quality. Materials and Methods: Seventeen patients affected by liver cancer were included in this study. Patients were imaged and treated in forced breath-hold using the Active Breathing Coordinator system and monitored with an optical tracking system. Three simulation CTs were acquired to estimate the anatomical variability between breath-holds and generate an internal target volume (ITV). The treatment plans were optimized with a Single Field Optimization technique aimed at minimizing the use of range shifter. Plan robustness was tested simulating systematic range and setup uncertainties, as well as the interplay effect between breath-holds. The appropriateness of margin was further verified based on the actual positioning data acquired during treatment. Results: The dose distributions of the nominal plans achieved a satisfactory target coverage in 11 out of 17 patients, while in the remaining 6 D95 to the PTV was affected by the constraint on mean liver dose. The constraints for all other organs at risk were always within tolerances. The interplay effect had a limited impact on the dose distributions: the worst case scenario showed a D95 reduction in the ITV < 3.9 GyRBE and no OAR with D-1 > 105% of the prescription dose. The robustness analysis showed that for 13 out of 17 patients the ITV coverage in terms of D95 was better than D95 of the PTV in the nominal plan. For the remaining 4 patients, the maximum difference between ITV D95 and PTV D95 was <= 0.7% even for the largest simulated setup error and it was deemed clinically acceptable. Hot spots in the OARs were always lower than 105% of the prescription dose. Positioning images confirmed that the breath hold technique and the PTV margin were adequate to compensate for inter- and intra-breath-hold variations in liver position. Conclusion: We designed and clinically applied a technique for the treatment of liver cancer with proton pencil beam scanning in forced deep expiration breath-hold. The initial data on plan robustness and patient positioning suggest that the choices in terms of planning technique and treatment margins are able to reach the desired balance between target coverage and organ at risk sparing. (C) 2020 Elsevier B.V. All rights reserved.
引用
收藏
页码:137 / 144
页数:8
相关论文
共 50 条
  • [41] Pencil-beam scanning proton therapy for low-grade glioma patients
    Willmann, Jonas
    Leiser, Dominic
    Weber, Damien Charles
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S822 - S825
  • [42] Endoscopic Resection Followed by Proton Therapy With Pencil Beam Scanning for Skull Base Tumors
    Leeman, Jonathan E.
    Lee, Nancy Y.
    Zhou, Ying
    Neal, Brian
    Sine, Kevin
    Tabar, Viviane
    Cohen, Marc A.
    LARYNGOSCOPE, 2019, 129 (06): : 1313 - 1317
  • [43] The Application of Cloud Computing in Pencil Beam Scanning Proton Therapy Monte Carlo Simulation
    Wang, Z.
    Gao, M.
    MEDICAL PHYSICS, 2014, 41 (06) : 296 - 296
  • [44] Brainstem-Sparing Craniospinal Irradiation Delivered With Pencil Beam Scanning Proton Therapy
    Hill-Kayser, Christine
    Kirk, Maura
    PEDIATRIC BLOOD & CANCER, 2015, 62 (04) : 718 - 720
  • [45] Effects of spot size errors in DynamicARC pencil beam scanning proton therapy planning
    Rana, Suresh
    Rosenfeld, Anatoly B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (23):
  • [46] Log file based Monte Carlo calculations for proton pencil beam scanning therapy
    Winterhalter, Carla
    Meier, Gabriel
    Oxley, David
    Weber, Damien C.
    Lomax, Antony J.
    Safai, Sairos
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (03):
  • [47] Pencil Beam Scanning Proton Therapy for Adolescents and Young Adults with Head and Neck Sarcomas
    Vazquez, Miriam
    Baust, Katja
    Ilundain, Amaia
    Leiser, Dominic
    Bachtiary, Barbara
    Pica, Alessia
    Kliebsch, Ulrike L.
    Calaminus, Gabriele
    Weber, Damien C.
    INTERNATIONAL JOURNAL OF PARTICLE THERAPY, 2023, 10 (02) : 73 - 84
  • [48] Clinical Outcomes of Pencil Beam Scanning Proton Therapy in Locally Advanced Non-Small Cell Lung Cancer: Propensity Score Analysis
    Kim, Nalee
    Noh, Jae Myoung
    Lee, Woojin
    Park, Byoungsuk
    Pyo, Hongryull
    CANCERS, 2021, 13 (14)
  • [49] Clinical outcomes of head and neck adenoid cystic carcinoma patients treated with pencil beam-scanning proton therapy
    Pelak, Maciej J.
    Walser, Marc
    Bachtiary, Barbara
    Hrbacek, Jan
    Lomax, Anthony J.
    Kliebsch, Ulrike L.
    Beer, Jurgen
    Pica, Alessia
    Malyapa, Robert
    Weber, Damien C.
    ORAL ONCOLOGY, 2020, 107
  • [50] Clinical outcomes and quality of life in children and adolescents with primary brain tumors treated with pencil beam scanning proton therapy
    Tran, Sebastien
    Lim, Pei S.
    Bojaxhiu, Beat
    Teske, Carmen
    Baust, Katja
    Zepter, Stefan
    Kliebsch, Ulrike
    Timmermann, Beate
    Calaminus, Gabriele
    Weber, Damien Charles
    PEDIATRIC BLOOD & CANCER, 2020, 67 (12)