Determining the Orientation of Directional Deep Brain Stimulation Electrodes Using 3D Rotational Fluoroscopy

被引:58
|
作者
Reinacher, P. C. [1 ]
Krueger, M. T. [2 ]
Coenen, V. A. [1 ]
Shah, M. [2 ]
Roelz, R. [2 ]
Jenkner, C. [4 ]
Egger, K. [3 ]
机构
[1] Freiburg Univ, Med Ctr, Dept Stereotact & Funct Neurosurg, Freiburg, Germany
[2] Freiburg Univ, Med Ctr, Dept Neurosurg, Freiburg, Germany
[3] Freiburg Univ, Med Ctr, Dept Neuroradiol, Freiburg, Germany
[4] Freiburg Univ, Clin Trial Unit, Med Ctr, Freiburg, Germany
关键词
RUBRO-THALAMIC TRACT; SUBTHALAMIC NUCLEUS; ASSESSING AGREEMENT; TREMOR; DBS;
D O I
10.3174/ajnr.A5153
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: New deep brain stimulation leads with electrode contacts that are split along their circumference allow steering of the electrical field in a predefined direction. However, imaging-assisted directional stimulation requires detailed knowledge of the exact orientation of the electrode array. The purpose of this study was to evaluate whether this information can be obtained by rotational 3D fluoroscopy. MATERIALS AND METHODS: Two directional leads were inserted into a 3D-printed plaster skull filled with gelatin. The torsion of the lead tip versus the lead at the burr-hole level was investigated. Then, 3 blinded raters evaluated 12 3D fluoroscopies with random lead orientations. They determined the lead orientation considering the x-ray marker only and considering the overlap of the gaps between the contact segments. Intraclass correlation coefficients and an extended version of the Bland-Altman plot were used to determine interrater reliability and agreement of the measurements of the different raters. RESULTS: Electrode torsion of up to 35 degrees could be demonstrated. Evaluation of the lead rotation considering the x-ray marker only revealed limits of agreement of 9.37 degrees and an intraclass correlation coefficient of 0.9975. In addition, taking into account the lines resulting from overlapping of the gaps between the electrode segments, the limits of agreement to the mean were +/- 2.44 degrees and an intraclass correlation coefficient of 0.9998. CONCLUSIONS: In directional deep brain stimulation systems, rotational 3D fluoroscopy combined with the described evaluation method allows for determining the exact orientation of the leads, enabling the full potential of imaging-assisted personalized programming.
引用
收藏
页码:1111 / 1116
页数:6
相关论文
共 50 条
  • [31] Evaluation of the Rotational Stability of Directional Deep Brain Stimulation Leads: A Case Series and Systematic Review
    Cavalloni, Fabian
    Brugger, Florian
    Kagi, Georg
    Naseri, Yashar
    Brogle, Deborah
    Bozinov, Oliver
    Bauer, Ronald
    Hagele-Link, Stefan
    Krueger, Marie Therese
    JOURNAL OF NEUROLOGICAL SURGERY PART A-CENTRAL EUROPEAN NEUROSURGERY, 2024, 85 (03) : 288 - 293
  • [32] A Novel and Simple Method Using Computed Tomography Streak Artifact to Determine the Orientation of Directional Deep Brain Stimulation Leads
    Fayed, Islam
    Syed, Mashaal
    Gingold, Eric
    Alizadeh, Mahdi
    Sharan, Ashwini
    Wu, Chengyuan
    NEUROSURGERY, 2023, 93 (05) : 1036 - 1045
  • [33] Biophysical characterization of local field potential recordings from directional deep brain stimulation electrodes
    Noor, M. Sohail
    McIntyre, Cameron C.
    CLINICAL NEUROPHYSIOLOGY, 2021, 132 (06) : 1321 - 1329
  • [34] Directional electrodes in deep brain stimulation: Results of a survey by the European Association of Neurosurgical Societies (EANS)
    Krauss, P.
    Duarte-Batista, P.
    Hart, M. G.
    Avecillas-Chasin, J. M.
    Bercu, M. M.
    Hvingelby, V.
    Massey, F.
    Ackermans, L.
    Kubben, P. L.
    van der Gaag, N. A.
    Kruger, M. T.
    BRAIN AND SPINE, 2024, 4
  • [35] Investigation of Magnetoelectric Sensor Requirements for Deep Brain Stimulation Electrode Localization and Rotational Orientation Detection
    Yalaz, Mevluet
    Deuschl, Guenther
    Butz, Markus
    Schnitzler, Alfons
    Helmers, Ann-Kristin
    Hoeft, Michael
    SENSORS, 2021, 21 (07)
  • [36] Intraoperative Myelography in Cervical Multilevel Stenosis Using 3D Rotational Fluoroscopy: Assessment of Feasibility and Image Quality
    Westermaier, Thomas
    Koehler, Stefan
    Linsenmann, Thomas
    Kiderlen, Michael
    Pakos, Paul
    Ernestus, Ralf-Ingo
    RADIOLOGY RESEARCH AND PRACTICE, 2015, 2015
  • [37] DETERMINING THE 3D POSITION AND ORIENTATION OF A ROBOT CAMERA USING 2D MONOCULAR VISION
    KITE, DH
    MAGEE, M
    PATTERN RECOGNITION, 1990, 23 (08) : 819 - 831
  • [38] Towards Tracking of Deep Brain Stimulation Electrodes Using an Integrated Magnetometer
    Quirin, Thomas
    Fery, Corentin
    Vogel, Dorian
    Vergne, Celine
    Sarracanie, Mathieu
    Salameh, Najat
    Madec, Morgan
    Hemm, Simone
    Hebrard, Luc
    Pascal, Joris
    SENSORS, 2021, 21 (08)
  • [39] Characterization of the distribution of rotational torque on electrorotation chips with 3D electrodes
    Bahrieh, Garsha
    Ozgur, Ebru
    Koyuncuoglu, Aziz
    Erdem, Murat
    Gunduz, Ufuk
    Kulah, Haluk
    ELECTROPHORESIS, 2015, 36 (15) : 1785 - 1794
  • [40] Determining 3D Kinematics of the Hip Using Video Fluoroscopy: Guidelines for Balancing Radiation Dose and Registration Accuracy
    D'Isidoro, Fabio
    Eschle, Patrik
    Zumbrunn, Thomas
    Sommer, Christian
    Scheidegger, Stephan
    Ferguson, Stephen J.
    JOURNAL OF ARTHROPLASTY, 2017, 32 (10): : 3213 - 3218