An Approach Method to Calibrate the Autocollimator with Small Angle Measurement Range

被引:1
作者
Novyanto, Okasatria [1 ]
Alfiyati, Nurul [1 ]
机构
[1] Indonesian Inst Sci RCM LIPI, Res Ctr Metrol, Mech Metrol Div, Bldg 420, Banten 15314, Indonesia
来源
MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA | 2016年 / 31卷 / 01期
关键词
Dissemination; Traceability; Electronic level; Angle deviation; Rotary encoder; UNCERTAINTY;
D O I
10.1007/s12647-015-0153-1
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new method to calibrate the autocollimator with small angle measurement range has been developed. This method used a system based on rotary encoder for angle measurements with self-calibration function (SelfA47). The SelfA47 has 18,000 graduation lines with angle interval of 0.02A degrees, corresponds to 72 arcsec. Thereby this system cannot be used to calibrate the autocollimator with measurement range < 72 arcsec. On the other hand, SelfA47 can be used to calibrate an electronic level. The deviation of electronic level shaped linear gradient. Therefore a small nominal gradient of electronic level can be approximated by interpolation. Since the gradient of a curve corresponds to the angle quantities, it is feasible to calibrate the autocollimator with nominal angle < 72 arcsec using SelfA47 and an electronic level. In this proposed method, SelfA47 is placed horizontally on the same plane with electronic level and mirror reflector position. The electronic level value that have been corrected by SelfA47 is used as a reference. This reference value was converted to the angle, thus the angle deviation of autocollimator can be determined. The prominent uncertainty of angle measurement using this new technique derived from the uncertainty of electronic level and rotary encoder. Using this method, the expanded uncertainty of 0.36 arcsec is obtained from calibration of autocollimator with measuring range of +/- 20 arcsec and it is consistent statistically with the manufacturer calibration result shown by E (n) value is < 1. In conclusion, this approach method can be applied to calibrate the autocollimator with a resolution smaller than SelfA47 as standard, but it is not recommended for the calibration of highly accurate autocollimator.
引用
收藏
页码:9 / 15
页数:7
相关论文
共 17 条
[1]  
[Anonymous], 2008, JCGM EV MEAS DAT GUI, P100
[2]  
[Anonymous], 2008, JCGM INT VOC METR BA
[3]  
[Anonymous], 2006, 18 IMEKO WORLD C RIO
[4]   Practical interpretation of metrological traceability on gauge block as to VIM 3 [J].
Chow, Lung-Hen ;
Chen, Yi-Ting ;
Chen, Liang-Hsing ;
Peng, Gwo-Sheng .
MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2011, 26 (01) :79-86
[5]  
Hewlett Packard, 1988, 5528A LAS MEAS SYST
[6]   Calibration of high-resolution electronic autocollimators against an angle comparator [J].
Just, A ;
Krause, M ;
Probst, R ;
Wittekopf, R .
METROLOGIA, 2003, 40 (05) :288-294
[7]   Interlaboratory Comparison in Measuring Low Levels of AC Voltage between Italy and Egypt [J].
Mageed, H. A. ;
Ali, R. S. M. ;
Tadros, N. N. ;
Halawa, M. ;
Lanzillotti, M. .
MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2014, 29 (03) :175-181
[8]  
National Association for Proficiency Testing, EX WHYS PROF TEST
[9]   A novel multiple reflections technique to calibrate polygons and evaluation of its uncertainty of measurement [J].
Sanjid, Arif M. ;
Chaudhary, K. P. .
MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2011, 26 (01) :29-35
[10]   Improved direct comparison calibration of small angle blocks [J].
Sanjid, M. Arif .
MEASUREMENT, 2013, 46 (01) :646-653