Test-retest intra-rater reliability of vaginal measurement of pelvic floor muscle strength using Myomed 932

被引:18
|
作者
Sigurdardottir, Thorgerdur [1 ]
Steingrimsdottir, Thora [2 ,3 ]
Arnason, Arni [1 ]
Bo, Kari [4 ]
机构
[1] Univ Iceland, Dept Physiotherapy, IS-101 Reykjavik, Iceland
[2] Univ Iceland, Dept Obstet & Gynecol, IS-101 Reykjavik, Iceland
[3] Reykjavik Primary Hlth Care Serv, Ctr Antenatal Hlth, Reykjavik, Iceland
[4] Norwegian Sch Sport Sci, Dept Sports Med, Oslo, Norway
关键词
Pelvic floor muscle strength; vaginal pressure measurements; intra-tester reliability; test-retest; Myomed; STRESS URINARY-INCONTINENCE; PRESSURE MEASUREMENTS; EXERCISE; WOMEN; CONTRACTION; ULTRASOUND; POSITIONS;
D O I
10.1080/00016340903093567
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Objectives. To investigate test-retest intra-rater reliability of measurements of pelvic floor muscle (PFM) strength using Myomed 932 (R) (Enraf-Nonius, Delft, the Netherlands). Design. Test-retest intra-rater reliability test with a 2-7 day interval. Setting. Physiotherapy outpatient clinic in association with the Department of Physiotherapy, University of Iceland. Population. Twenty healthy female volunteers in Reykjavik metropolitan area, Iceland, with a mean age of 43.8 (SD 9.8) years, mean body mass index (BMI) 26.2 (SD 5.3), and mean parity 2.1 children (SD 1.0). Main outcome measures. PFM strength during two attempts of maximal voluntary contraction (MVC) measured as vaginal squeeze pressure in hectoPascals (hPa). Results. Mean MVC was 34.45 hPa (95% CI: 26.07-42.83) on Day 1 and 34.05 hPa (95% CI: 26.66-41.44) on Day 2. The measurements ranged from 5-74 hPa interindividually on Day 1 and 5-62 hPa on Day 2. Intraclass Correlation Coefficient (ICC) was 0.97; p<0.001 and coefficient of variation (CV%) was 11.09%. There was no significant difference between the first and second tests (p = 0.74). Conclusion. This test-retest intra-rater reliability study for Myomed 932 showed a very good reliability using ICC. The CV%, however, indicated some degree of intra-individual variation. We suggest that this device can be used both in practice and clinical trials for assessing female PFM strength.
引用
收藏
页码:939 / 943
页数:5
相关论文
共 27 条
  • [1] Intra-Rater Reliability and Diagnostic Accuracy of a New Vaginal Dynamometer to Measure Pelvic Floor Muscle Strength in Women With Urinary Incontinence
    Romero-Culleres, Georgia
    Pena-Pitarch, Esteban
    Jane-Feixas, Celia
    Arnau, Anna
    Montesinos, Jesus
    Abenoza-Guardiola, Montserrat
    NEUROUROLOGY AND URODYNAMICS, 2017, 36 (02) : 333 - 337
  • [2] Intra-session test-retest reliability of pelvic floor muscle electromyography during running
    Luginbuehl, H.
    Greter, C.
    Gruenenfelder, D.
    Baeyens, J-P.
    Kuhn, A.
    Radlinger, L.
    INTERNATIONAL UROGYNECOLOGY JOURNAL, 2013, 24 (09) : 1515 - 1522
  • [3] AN INTRA -AND INTER-RATER RELIABILITY STUDY OF THE MEASUREMENT OF PELVIC FLOOR MUSCLE FUNCTION USING MANOMETRY.
    Tennfjord, M.
    Hilde, G.
    Ellstrom-Engh, M.
    Bo, K.
    NEUROUROLOGY AND URODYNAMICS, 2016, 35 : S459 - S460
  • [4] An intra- and interrater reliability and agreement study of vaginal resting pressure, pelvic floor muscle strength, and muscular endurance using a manometer
    Tennfjord, Merete Kolberg
    Engh, Marie Ellstrom
    Bo, Kari
    INTERNATIONAL UROGYNECOLOGY JOURNAL, 2017, 28 (10) : 1507 - 1514
  • [5] Test-Retest Reliability of Pelvic Floor Muscle Contraction Measured by 4D Ultrasound
    Braekken, Ingeborg Hoff
    Majida, Memona
    Engh, Marie Ellstrom
    Bo, Kari
    NEUROUROLOGY AND URODYNAMICS, 2009, 28 (01) : 68 - 73
  • [6] Test-retest reliability of a handheld dynamometer for measurement of isometric cervical muscle strength
    Vannebo, Katrine Tranaas
    Iversen, Vegard Moe
    Fimland, Marius Steiro
    Mork, Paul Jarle
    JOURNAL OF BACK AND MUSCULOSKELETAL REHABILITATION, 2018, 31 (03) : 557 - 565
  • [7] Intra-session test–retest reliability of pelvic floor muscle electromyography during running
    H. Luginbuehl
    C. Greter
    D. Gruenenfelder
    J-P. Baeyens
    A. Kuhn
    L. Radlinger
    International Urogynecology Journal, 2013, 24 : 1515 - 1522
  • [8] Intra and inter-rater reliability study of pelvic floor muscle dynamometric measurements
    Martinho, Natalia M.
    Marques, Joseane
    Silva, Valeria R.
    Silva, Silvia L. A.
    Carvalho, Leonardo C.
    Botelho, Simone
    BRAZILIAN JOURNAL OF PHYSICAL THERAPY, 2015, 19 (02) : 97 - 104
  • [9] Test-retest reliability of the different dynamometric variables used to evaluate pelvic floor musculature during the menstrual cycle
    dos Reis Nagano, Reny C.
    Biasotto-Gonzalez, Daniela A.
    da Costa, Gilmar L.
    Amorim, Karina M.
    Fumagalli, Marco A.
    Amorim, Cesar F.
    Politti, Fabiano
    NEUROUROLOGY AND URODYNAMICS, 2018, 37 (08) : 2606 - 2613
  • [10] Reliability and Diagnostic Accuracy of a New Vaginal Dynamometer to Measure Pelvic Floor Muscle Strength
    Romero-Culleres, Georgia
    Pena-Pitarch, Esteban
    Jane-Feixas, Cecilia
    Vilaseca-Grane, Anna
    Montesinos, Jesus
    Arnau, Anna
    FEMALE PELVIC MEDICINE AND RECONSTRUCTIVE SURGERY, 2020, 26 (08): : 514 - 519