A comparison of imaging methods for use in an array biosensor

被引:44
作者
Golden, JP [1 ]
Ligler, FS [1 ]
机构
[1] USN, Res Lab, Ctr Biomol Sci & Engn, Washington, DC 20375 USA
关键词
CMOS; CCD; florescence; biosensor; array sensor;
D O I
10.1016/S0956-5663(02)00060-X
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
An array biosensor has been developed which uses an actively-cooled, charge-coupled device (CCD) imager. In an effort to save money and space, a complementary metal-oxide semiconductor (CMOS) camera and photodiode were tested as replacements for the cooled CCD imager. Different concentrations of CY5 fluorescent dye in glycerol were imaged using the three different detection systems with the same imaging optics. Signal discrimination above noise was compared for each of the three systems. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:719 / 725
页数:7
相关论文
共 14 条
[1]   Water quality monitoring using an automated portable fiber optic biosensor: RAPTOR [J].
Anderson, GP ;
Rowe-Taitt, CA .
PHOTONIC DETECTION AND INTERVENTION TECHNOLOGIES FOR SAFE FOOD, 2001, 4206 :58-63
[2]  
Duveneck G.L., 1995, [No title captured], Patent No. [WO 9533197, 9533197]
[3]  
Golden J.P., 1998, US patent, Patent No. [5827748, 5,827,748]
[4]   Array biosensor for simultaneous detection of multiple analytes [J].
Ligler, FS ;
Golden, JP ;
Rowe-Taitt, CA ;
Dodson, JP .
ADVANCES IN FLUORESCENCE SENSING TECHNOLOGY V, 2001, 4252 :32-36
[5]  
LIGLER FS, 1992, SYNTHETIC MICROSTRUCTURES IN BIOLOGICAL RESEARCH, P67
[6]   Multiple-analyte fluoroimmunoassay using an integrated optical waveguide sensor [J].
Plowman, TE ;
Durstchi, JD ;
Wang, HK ;
Christensen, DA ;
Herron, JN ;
Reichert, WM .
ANALYTICAL CHEMISTRY, 1999, 71 (19) :4344-4352
[7]   Array biosensor for simultaneous identification of bacterial, viral, and protein analytes [J].
Rowe, CA ;
Tender, LM ;
Feldstein, MJ ;
Golden, JP ;
Scruggs, SB ;
MacCraith, BD ;
Cras, JJ ;
Ligler, FS .
ANALYTICAL CHEMISTRY, 1999, 71 (17) :3846-3852
[8]   An array immunosensor for simultaneous detection of clinical analytes [J].
Rowe, CA ;
Scruggs, SB ;
Feldstein, MJ ;
Golden, JP ;
Ligler, FS .
ANALYTICAL CHEMISTRY, 1999, 71 (02) :433-439
[9]   Simultaneous detection of six biohazardous agents using a planar waveguide array biosensor [J].
Rowe-Taitt, CA ;
Hazzard, JW ;
Hoffman, KE ;
Cras, JJ ;
Golden, JP ;
Ligler, FS .
BIOSENSORS & BIOELECTRONICS, 2000, 15 (11-12) :579-589
[10]  
Silzel JW, 1998, CLIN CHEM, V44, P2036